DQR ID | Subject | Data Streams Affected |
---|---|---|
D000210.3 | SGP/EBBR - EBBR E22 Moved, Data Missing/Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D000210.4 | SGP/EBBR - EBBR E22 Net Radiation Incorrect | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D000229.2 | SGP/EBBR - EBBR E8 Left Air Temperature (thermocouple) Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D000301.10 | SGP/EBBR - EBBR E8 Wind Direction Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D000301.11 | SGP/EBBR - EBBR E22 Soil Moistures Negative | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D000302.2 | SGP/EBBR - EBBR E22 Soil Moistures Negative | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D010213.8 | SGP/EBBR/E22 - EBBR E22 Wind Speed Iced | sgp30ebbrE22.b1, sgp15ebbrE22.b1, sgp5ebbrE22.b1 |
D010307.1 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.10 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.100 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.101 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.102 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.103 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.104 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.105 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.11 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.12 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.13 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.14 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.15 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.16 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.17 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.18 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.19 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.20 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.21 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.22 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.23 | SGP/EBBR/E8 - E8 EBBR Low Battery Condition | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.24 | SGP/EBBR/E8 - E8 EBBR Wind Speed Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.25 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.26 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.27 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.28 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.29 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.3 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.30 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.31 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.32 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.33 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.34 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.35 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.36 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.37 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.38 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.39 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.4 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.40 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.41 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.42 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.43 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.44 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.45 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.46 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.47 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.48 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.49 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.5 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.50 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.51 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.52 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.53 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.54 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.55 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.56 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.57 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.58 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.59 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.6 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.60 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.61 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.62 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.63 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.64 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.65 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.66 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.67 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.68 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.69 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.7 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.70 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.71 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.73 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.74 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.75 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.76 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.77 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.78 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.79 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.8 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.80 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.81 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.82 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.83 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.85 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.86 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.87 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.88 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.89 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.9 | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.90 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.91 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.92 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.93 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.94 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.95 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.96 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.97 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.98 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010307.99 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.100 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.101 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.102 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.103 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.105 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.106 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.107 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.108 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.109 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.110 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.111 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.112 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.113 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.114 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.115 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.116 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.117 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.118 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.119 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.12 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.120 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.121 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.122 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.123 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.124 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.125 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.126 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.127 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.128 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.129 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.13 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.130 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.131 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.132 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.133 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.134 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.135 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.136 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.137 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.138 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.139 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.14 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.140 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.141 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.142 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.143 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.144 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.145 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.146 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.15 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.16 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.17 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.18 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.19 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.20 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.21 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.22 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.23 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.24 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.25 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.26 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.27 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.28 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.29 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.30 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.31 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.32 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.33 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.34 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.35 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.36 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.37 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.38 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.39 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.40 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.41 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.42 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.43 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.44 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.45 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.46 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.47 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.48 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.49 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.50 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.51 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.52 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.53 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.55 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.56 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.57 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.58 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.59 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.60 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.61 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.62 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.63 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.64 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.65 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.66 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.67 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.68 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.69 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.70 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.71 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.72 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.73 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.74 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.75 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.76 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.77 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.78 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.79 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.80 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.81 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.82 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.83 | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.84 | SGP/EBBR/E8 - E8 EBBR Data Incorrect During System Replacement | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.85 | SGP/EBBR/E8 - E8 EBBR Data Not Based On 30 Minute Average | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.87 | SGP/EBBR/E8 - E8 EBBR Wind Direction Frozen | sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1 |
D010316.88 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.89 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.90 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.91 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.92 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.93 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.94 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.95 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.96 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D010316.97 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.98 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010316.99 | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D010427.11 | SGP EBBR E22 - E22 EBBR Data Incorrect During Logger Replacement | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
D010427.13 | SGP EBBR E22 - E22 EBBR Soil Moisture Lower Than Measurement Capability | sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
D010427.14 | SGP EBBR E22 - E22 EBBR Soil Moisture #3 Lower Than Can Be Measured | sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
D010427.15 | SGP EBBR E22 - E22 EBBR Net Radiation Incorrect | sgp15ebbrE22.b1, sgp5ebbrE22.b1, sgp30ebbrE22.b1 |
D011127.24 | SGP EBBR E8 - E8 EBBR Data Incorrect During 6 Month Checks | sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1 |
D011127.32 | SGP EBBR E22 - E22 EBBR Data Incorrect During 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D011206.4 | SGP EBBR E22 - E22 EBBR 30 Minute Data Incorrect After Missing Data | sgp30ebbrE22.b1 |
D011219.2 | SGP EBBR E8 - E8 EBBR Soil Heat Flow #1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D011219.3 | SGP EBBR E8 - E8 EBBR Soil Heat Flow #1 Incorrect | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D011219.4 | SGP EBBR E8 - E8 EBBR AEM Frozen In One Position | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D011219.5 | SGP EBBR E8 - E8 EBBR Wind Speed Frozen | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D011220.1 | SGP EBBR E8 - E8 EBBR Battery Voltage Low, Data Incorrect | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.00, sgp15ebbrE8.b1 |
D011220.2 | SGP EBBR E8 - E8 EBBR Battery Voltage Low, Data Incorrect | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.00 |
D020628.13 | SGP/EBBR/E8 - T/RH Check | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D020628.19 | SGP/EBBR/E22 - T/RH Check | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D020628.5 | SGP/EBBR/E22 - 3-month Check | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D020628.9 | SGP/EBBR/E8 - 3-month Check | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D020716.17 | SGP/EBBR/E22 - 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D020716.29 | SGP/EBBR/E8 - Low Battery Voltage Causes Incorrect Measurements | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.00, sgp5ebbrE8.b1 |
D020716.3 | SGP/EBBR/E22 - Annual Calibration Changeout | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D020716.30 | SGP/EBBR/E8 - AEM Hung | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D020716.9 | SGP/EBBR/E8 - 6 Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D020717.34 | SGP/EBBR/E22 - All Measurements Zero | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.00, sgp5ebbrE22.b1 |
D020717.35 | SGP/EBBR/E22 - AEM Hung or High Home Signals | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D021113.1 | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR Header | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE18.b1, sgp30ebbrE19.b1, sgp30ebbrE2.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
D021113.4 | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR Header | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
D030110.26 | SGP/EBBR/E22 - Data Incorrect during 6 month checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030110.31 | SGP/EBBR/E8 - Data Incorrect during 6 month checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030113.10 | SGP/EBBR/E8 - AEM Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030113.11 | SGP/EBBR/E8 - Wind Speed Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030113.12 | SGP/EBBR/E8 - Water in broken Net Rad Dome | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030113.9 | SGP/EBBR/E8 - Battery Voltage Low | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030117.16 | SGP/EBBR/E22 - AEM Hung between home positions | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030117.17 | SGP/EBBR/E22 - Water inside Broken Net Rad Dome | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030711.7 | SGP/EBBR/E22 - home30 too large | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030712.14 | SGP/EBBR/E8 - 6 Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030712.22 | SGP/EBBR/E22 - 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030712.4 | SGP/EBBR/E8 - AEM Hung Inbetween Home Positions | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030712.5 | SGP/EBBR/E8 - Battery Voltage Low | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D030729.1 | SGP/EBBR/E8 - Pressure Sensor Measurements Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgpebbrE8.01 |
D030729.4 | SGP/EBBR/E22 - AEM Hung Inbetween Home Positions | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D030916.2 | SGP/EBBR/E22 - AEM Home Signals Too High | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D030919.6 | SGP/EBBR/E8 - Water in Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D031031.8 | SGP/EBBR/E22 - Water in Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D031204.20 | SGP/EBBR/E22 - 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D031204.7 | SGP/EBBR/E8 - 6 Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D031222.12 | SGP/EBBR/E8 - Net Radiometer Iced Over During Storm | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D031222.21 | SGP/EBBR/E22 - Wind Speed Sensor Frozen By Ice Storm | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D031222.22 | SGP/EBBR/E22 - Wind Direction Sensor Frozen By Ice Storm | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D040303.2 | SGP/EBBR/E22 - Wind Speed Sensor Bearings Worn | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D040304.21 | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D040304.22 | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D040304.6 | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D040304.7 | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D040318.17 | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped By Icing | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D040331.2 | SGP/EBBR/E8 - AEM Replacement | sgp15ebbrE8.b1, sgp30baebbrE8.c1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D040706.2 | SGP/EBBR/E22 - Net Radiation Incorrect | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D040723.3 | SGP/EBBR/E22 - Water In Net Radiometer | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D040915.5 | SGP/EBBR/E8 - Net Radiation Incorrect, Fluxes Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D041222.17 | SGP/EBBR/E22 - 6 Month Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D041222.9 | SGP/EBBR/E8 - 6 Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050206.15 | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050206.7 | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050207.11 | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050207.3 | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice Storm | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050215.12 | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice Storm | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050215.13 | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050215.21 | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050215.22 | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050323.2 | SGP/EBBR/E22 - Net Radiation Too Low - Top Dome Broken | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D050413.3 | SGP/EBBR/E8 - Battery Voltage Problem | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D050719.14 | SGP/EBBR/E22 - Improved EBBR CR10 Program | sgp30ebbrE22.b1 |
D050719.7 | SGP/EBBR/E8 - Improved EBBR CR10 Program | sgp30ebbrE8.b1 |
D051106.13 | SGP/EBBR/E22 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D051106.21 | SGP/EBBR/E8 - Average Soil Heat Flux Incorrect | sgp30ebbrE8.b1 |
D051106.29 | SGP/EBBR/E22 - Average Soil Heat Flux Incorrect | sgp30ebbrE22.b1 |
D051106.4 | SGP/EBBR/E8 - Soil Heat Flow #2 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D051106.5 | SGP/EBBR/E8 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D051112.5 | SGP/EBBR/E8 - metadata corrections | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D051112.8 | SGP/EBBR/E22 - metadata corrections | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D060111.4 | SGP/EBBR/E8 - AEM Not Exchanging | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060112.1 | SGP/EBBR/E22 - Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D060428.4 | SGP/EBBR/E22 - Fluxes Incorrect When Soil Heat Flow #3 Out of Range | sgp30ebbrE22.b1 |
D060501.25 | SGP/EBBR/E22 - Wind Speed Sensor Frozen Stopped | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D060501.6 | SGP/EBBR/E8 - Flux and Gradient Measurements Suspect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060501.8 | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060501.9 | SGP/EBBR/E8 - Fluxes Incorrect; Left Humidity Near Zero | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060508.2 | SGP/EBBR/E8 - Sensible and Latent Heat Flux Suspect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060818.8 | SGP/EBBR/E8 - Sensible and Latent Heat Fluxes and Net Radiation Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060921.8 | SGP/EBBR/E22 - Net Radiation and Sensible and Latent Heat Fluxes Incorrect | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D061106.5 | SGP/EBBR/E22 - Soil Heat Flow #4 Incorrect - Sensible and Latent Heat Fluxes Sometimes Affected | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D061220.13 | SGP/EBBR/E22 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE22.b1 |
D061220.4 | SGP/EBBR/E8 - Wind Direction Dependency of Surface Fluxes and General Data Quality Information | sgp30ebbrE8.b1 |
D070223.2 | SGP/EBBR/E8 - Failed Temp/RH sensor | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D070426.1 | SGP/EBBR/E22 - Wind Direction Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D070815.6 | SGP/EBBR/E22 - H and LE Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D080520.1 | SGP/EBBR/E8 - Wind Direction Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D090313.15 | SGP/EBBR/E8 - Wind Speed Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D090313.6 | SGP/EBBR/E22 - Wind Speed Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D090817.1 | SGP/EBBR/E8 - Insufficient voltage | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D110105.3 | SGP/EBBR/E8 - RH Maximum too high | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D931115.3 | SGP/EBBR/E8 - Drop in atmospheric pressure | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D940105.10 | SGP/EBBR/E22 - Air Temperature Sensor malfunction | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D940105.7 | SGP/EBBR/E22 - Incorrect averaging interval at E22 | sgp30ebbrE22.b1 |
D940106.1 | SGP/EBBR/E8 - Invalid data due to AEM fuse blow out at E8 | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D940211.2 | SGP/EBBR/E8 - Invalid wind data frozen sensor | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D940301.3 | SGP/EBBR/E8 - Invalid wind speed frozen sensor | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D940830.10 | SGP/EBBR/E8 - Battery Voltage Drop | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D940912.1 | SGP/EBBR/E8 - Data Corruption and Loss | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D941021.10 | SGP/EBBR/E22 - Net Radiometer Hemisphere Changes | sgp5ebbrE22.b1, sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
D941114.2 | SGP/EBBR/E8 - Water in Net Radiometer | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D941212.3 | SGP/EBBR/E8 - Frozen Wind Instruments | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D950126.3 | SGP/EBBR/E8 - Frozen Wind Speed Instrument | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D950828.2 | AEM Home 30 Values Low | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D950922.2 | SGP/EBBR/E8 - Home 30 Values Correct | sgp30ebbrE8.b1 |
D960104.3 | SGP/EBBR/E22 - Wind Speed Cups Frozen | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D960408.2 | SGP/EBBR/E22 - Reprocess: Soil Heat Flow Sensor 4 Incorrect Data | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D960508.3 | SGP/EBBR/E8 - 3-month check | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D960508.9 | SGP/EBBR/E22 - 3-month check | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D961122.1 | SGP/EBBR/E8 - Annual Calibration Changeout | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D961122.14 | SGP/EBBR/E8 - Incorrect Right Humidity Probe Temperature | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
D961122.15 | SGP/EBBR/E8 - Reprocess: Soil Heat Flow Sensor 2 Incorrect Data | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D961122.2 | SGP/EBBR/E22 - Annual Calibration Changeout | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D961122.21 | SGP/EBBR/E8 - EBBR 3 Month Checks - 10 EFs | sgp15ebbrE8.b1, sgp5ebbrE8.b1, sgp30ebbrE8.b1 |
D961122.27 | SGP/EBBR/E22 - EBBR 3 Month Checks - 10 EFs | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D961122.6 | SGP/EBBR/E8 - Incorrect Soil Moisture #1 | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D961122.7 | SGP/EBBR/E8 - Incorrect Soil Moisture #2, Soil Temperature #2 | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D970616.1 | SGP/EBBR/E8 - Incorrect Soil Temperature #s 1 and 2 | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D970703.1 | SGP/EBBR/E8 - Incorrect Soil Temperature #s 1 and 2 | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D970707.1 | SGP/EBBR/E22 - Reprocess: E22 Soil Heat Flow Sensor # 5 Incorrect Data | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D970716.2 | SGP/EBBR/E8 - AEM Inoperative | sgp30ebbrE8.b1 |
D970730.1 | SGP/EBBR/E22 - Right Aspirator Unit Stolen | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D971017.1 | SGP/EBBR/E8 - Reprocess: E8 EBBR Program Compile Error, Effect On Net Radiation | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D971017.2 | E8 EBBR Lightning Strike | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D971017.7 | SGP/EBBR/E22 - Reprocess: E22 EBBR Soil Heat Flow Probe #5 Incorrect Data | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D971017.8 | SGP/EBBR/E8 - Reprocess: E8 EBBR Lightning Strike / sensor Replacements | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D971021.2 | SGP/EBBR/E8 - EBBR Biannual Checks | sgp15ebbrE8.b1, sgp5ebbrE8.b1, sgp30ebbrE8.b1 |
D971021.8 | SGP/EBBR/E22 - EBBR Biannual Checks | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D971111.1 | SGP/EBBR/E8 - Reprocess: SHF2 Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D980117.1 | SGP/EBBR/E22 - Reprocess: E22 Soil Moisture Values Off End of Calibration at High Soil Moistures | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D980117.16 | SGP/EBBR/E8- EBBR Frozen Wind Speed | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D980117.7 | SGP/EBBR/E8 - Battery Low - Data Affected | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
D980117.9 | SGP/EBBR/E8 - Reprocess: E8 Soil Heat Flow #2 Failure - Replaced | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
D980410.4 | SPG/EBBR/E8 - Wind Speed Sensor and AEM Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980413.5 | SGP/EBBR/E22 - Low Battery - Resulting In Incorrect Data | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D980416.17 | SGP/EBBR/E22 - 3-Month Check | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
D980416.36 | SGP/EBBR/E8 - Wind Speed Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980416.37 | SGP/EBBR/E8 - Wind Speed Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980416.38 | SGP/EBBR/E8 - Wind Speed Frozen | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980416.9 | SGP/EBBR/E8 - 3-Month Check | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980526.2 | EBBR E8 Net Radiation Data Quality | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980529.15 | SGP/EBBR/E22 - Six Month Checks | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
D980529.7 | SGP/EBBR/E8 - Six Month Checks | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980603.1 | EBBR E8 Adjustment of Negative Net Radiation Values | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980603.2 | EBBR E8 Adjustment of Negative Net Radiation Values | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D980603.3 | SGP/EBBR/E8 - Wind Direction Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D990127.13 | SGP/EBBR/E8 - Wind Direction and Speed Data Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D990127.2 | SGP/EBBR/E22 - Battery Voltage Low, Data Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
D990127.30 | SGP/EBBR/E22 - Wind Direction Data Incorrect | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
D990128.8 | SGP/EBBR/E8 - Wind Speed Data Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D990202.3 | SGP/EBBR - EBBR E8 Wind Speed Iced, Data Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp5ebbrE8.00 |
D990208.1 | SGP/EBBR/E8 - Wind Direction Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D990208.2 | SGP/EBBR/E8 - Wind Speed Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp5ebbrE8.00 |
D990316.1 | SGP/EBBR/E22 - Right Aspirator Fan Replaced | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
D990415.2 | SGP/EBBR/E8 - Wind Speed Frozen | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/1999 | 1415 | 11/02/1999 | 2030 |
Subject: | SGP/EBBR - EBBR E22 Moved, Data Missing/Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During, and as a consequence of, the relocation of E22 EBBR, some data is missing or incorrect during the period above. The data was sneakernetted on 3 Nov 99, so some of the data may have been recovered. Some of the data during the period is incorrect as a consequence of damage done by cattle during the period; inspection of the data is required to determine incorrect data. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/1999 | 1430 | 11/03/1999 | 1930 |
Subject: | SGP/EBBR - EBBR E22 Net Radiation Incorrect |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The upper dome of the net radiometer was broken, probably by migrating birds, and water had drained into the bottom dome. Since much of the data is missing during the period, it is not clear exactly when the top dome was broken during the period. All net radiation data and sensible and latent heat fluxes data are incorrect for this period. The affected data fields are: 5 minute: q 15 minute: mv_q 30 minute: q, e, h |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/1999 | 1600 | 02/01/2000 | 1900 |
Subject: | SGP/EBBR - EBBR E8 Left Air Temperature (thermocouple) Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The left air temperature (thermocouple) frequently went offscale, for an unknown reason, thereby affecting many quantities. The affected data are: 5 minute: tair_bot, tair_top 15 minute: tair_l 30 minute: tair_bot, tair_top, bowen, h, e |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/22/1999 | 1300 | 09/28/1999 | 1830 |
Subject: | SGP/EBBR - EBBR E8 Wind Direction Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | A malfunctioning wind direction sensor caused the wind direction measurement to be incorrect for the period. The affected data are: 5 minute: res_ws, wind_d 15 minute: mv_wind_d 30 minute: res_ws, wind_d |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/1999 | 0300 | 08/04/1999 | 1730 |
Subject: | SGP/EBBR - EBBR E22 Soil Moistures Negative |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Because of the lack of rain, soil moistures were negative during much of the period. All five soil moisture sensors had negative outputs by early on Aug 1. The negative soil moistures affected sensible and latent heat fluxes slightly. The affected data are: 15 minute: r_sm1, r_sm2, r_sm3, r_sm4, r_sm5 30 minute: sm1, sm2, sm3, sm4, sm5, c_shf1, c_shf2, c_shf3, c_shf4, c_shf5, cs1, cs2, cs3, cs4, cs5, ces1, ces2, ces3, ces4, ces5, g1, g2, g3, g4, g5, ave_shf, e, h |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/15/1999 | 0800 | 10/30/1999 | 0300 |
Subject: | SGP/EBBR - EBBR E22 Soil Moistures Negative |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Because of the lack of rain, all soil moistures were negative during much of the period. The negative soil moistures caused soil heat storage and thus average soil heat flux to be smaller than they should have been. The sensible and latent heat fluxes were therefore reduced slightly (approximately 0-5%). The affected data are: 15 minute: r_sm1, r_sm2, r_sm3, r_sm4, r_sm5 30 minute: sm1, sm2, sm3, sm4, sm5, c_shf1, c_shf2, c_shf3, c_shf4, c_shf5, cs1, cs2, cs3, cs4, cs5, ces1, ces2, ces3, ces4, ces5, g1, g2, g3, g4, g5, ave_shf, e, h |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2000 | 1315 | 12/27/2000 | 2015 |
Subject: | SGP/EBBR/E22 - EBBR E22 Wind Speed Iced |
DataStreams: | sgp30ebbrE22.b1, sgp15ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The anemometer was frozen stopped by freezing rain. Indicated data is incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/03/2000 | 1030 | 02/15/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | When the AEM (automatic exchange mechanism) is stuck in one postion, half hour latent and sensible heat fluxes and the Bowen ratio are incorrect. Some of the raw EBBR measurements are also incorrect. I do not normally submit a DQR for this situation, as the qc checks for AEM home signals normally indicate the problem. However, the frequency of the problem at this site seems to warrant DQRs. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/20/2000 | 1530 | 03/28/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 2300 | 06/26/2000 | 2335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2000 | 0000 | 06/27/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2000 | 1500 | 06/27/2000 | 1505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/2000 | 1100 | 06/29/2000 | 1635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/29/2000 | 1830 | 06/30/2000 | 2135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/30/2000 | 2230 | 07/01/2000 | 1205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/28/2000 | 2000 | 04/09/2000 | 1435 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/09/2000 | 1600 | 04/11/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/11/2000 | 1800 | 04/25/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2000 | 2330 | 05/13/2000 | 0235 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2000 | 0500 | 05/14/2000 | 0205 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 0330 | 05/14/2000 | 1035 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 1200 | 06/06/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for much of this period. Please look at the home signal qc to determine whether the data is indeed incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/14/2000 | 0930 | 06/20/2000 | 1635 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/22/2000 | 0900 | 06/23/2000 | 1705 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 0330 | 06/25/2000 | 1935 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 1330 | 06/26/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 1530 | 07/31/2000 | 1630 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect for this period. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/2000 | 0700 | 12/05/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Low Battery Condition |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | All indicated data fields should be considered to be incorrect when the battery voltage is less than 10.4 volts. Check the battery voltage to determine times of incorrect data. The battery was replaced on 5 Dec. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2000 | 0800 | 11/20/2000 | 1750 |
Subject: | SGP/EBBR/E8 - E8 EBBR Wind Speed Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The indicated data fields are incorrect for the stated period because of failure of the anemometer electronics. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/27/2000 | 0030 | 03/27/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/07/2000 | 1600 | 04/08/2000 | 1705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 0530 | 04/10/2000 | 1405 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 1800 | 04/10/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 2030 | 04/10/2000 | 2035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/17/2000 | 0130 | 02/21/2000 | 0035 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 2200 | 04/10/2000 | 2235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2000 | 2330 | 04/11/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/13/2000 | 0130 | 04/13/2000 | 0135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2000 | 0000 | 04/16/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2000 | 0100 | 04/16/2000 | 0235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2000 | 0430 | 04/16/2000 | 0635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2000 | 0830 | 04/16/2000 | 2105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/2000 | 1400 | 04/17/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/20/2000 | 0230 | 04/21/2000 | 0035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/2000 | 0100 | 04/22/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/2000 | 1000 | 02/23/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/2000 | 0600 | 04/22/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2000 | 0030 | 04/24/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2000 | 0630 | 04/25/2000 | 0635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2000 | 1400 | 04/25/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2000 | 2000 | 04/25/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/26/2000 | 0430 | 04/26/2000 | 0505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2000 | 1530 | 05/09/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2000 | 1700 | 05/09/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/10/2000 | 0230 | 05/10/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/12/2000 | 1530 | 05/12/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/24/2000 | 1530 | 02/25/2000 | 0405 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2000 | 0300 | 05/13/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2000 | 1600 | 05/13/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 0230 | 05/14/2000 | 0305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 1100 | 05/14/2000 | 1305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2000 | 1530 | 05/14/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/18/2000 | 1430 | 05/18/2000 | 1435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/2000 | 1700 | 05/19/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/23/2000 | 1800 | 05/23/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/07/2000 | 0430 | 06/07/2000 | 0505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/07/2000 | 1230 | 06/07/2000 | 1235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/26/2000 | 0200 | 02/29/2000 | 1835 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/08/2000 | 1030 | 06/08/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/09/2000 | 0000 | 06/09/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/09/2000 | 0100 | 06/09/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 0000 | 06/10/2000 | 0205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 0330 | 06/10/2000 | 0335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 0500 | 06/10/2000 | 0605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 1230 | 06/10/2000 | 1635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 1730 | 06/10/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2000 | 1900 | 06/10/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/12/2000 | 1500 | 06/12/2000 | 1505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/29/2000 | 2000 | 03/19/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/13/2000 | 1000 | 06/13/2000 | 1235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/13/2000 | 1330 | 06/13/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/13/2000 | 1700 | 06/13/2000 | 1705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 0900 | 06/16/2000 | 0905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1030 | 06/16/2000 | 1135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1230 | 06/16/2000 | 1235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1400 | 06/16/2000 | 1435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1500 | 06/16/2000 | 1505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1530 | 06/16/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/19/2000 | 2330 | 03/20/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 1630 | 06/16/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2000 | 2200 | 06/16/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/17/2000 | 2030 | 06/17/2000 | 2035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 0100 | 06/21/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 0430 | 06/21/2000 | 0505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 0730 | 06/21/2000 | 0805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 0900 | 06/21/2000 | 0905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2000 | 2230 | 06/21/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/24/2000 | 0000 | 06/24/2000 | 0005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/20/2000 | 0230 | 03/20/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The raw temperature and humidity data is incorrect and thus Bowen ratio, and half hour sensible and latent heat flux are incorrect. Normally I do not submit DQRs for this condition, but the frequency of the problem was high for this site. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/24/2000 | 0100 | 06/24/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/24/2000 | 0300 | 06/24/2000 | 0335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 1000 | 06/25/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 1730 | 06/25/2000 | 1735 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 1900 | 06/25/2000 | 1935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 2030 | 06/25/2000 | 2035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 2200 | 06/25/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/25/2000 | 2300 | 06/25/2000 | 2335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 0000 | 06/26/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2000 | 1200 | 06/26/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 0500 | 08/25/2000 | 0501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 1800 | 08/25/2000 | 1801 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 1930 | 08/25/2000 | 1931 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 2100 | 08/25/2000 | 2101 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/20/2000 | 2230 | 08/20/2000 | 2231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/11/2000 | 0230 | 09/11/2000 | 0231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/11/2000 | 1030 | 09/11/2000 | 1131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/11/2000 | 1700 | 09/11/2000 | 1831 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/2000 | 0600 | 09/12/2000 | 1031 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/2000 | 0330 | 09/12/2000 | 0431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/12/2000 | 1730 | 09/12/2000 | 1901 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/13/2000 | 0200 | 09/13/2000 | 0631 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/13/2000 | 1730 | 09/13/2000 | 1901 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2000 | 0130 | 09/14/2000 | 0431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2000 | 1800 | 09/14/2000 | 2201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/2000 | 0005 | 09/15/2000 | 0201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/2000 | 1700 | 09/16/2000 | 0301 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/16/2000 | 1800 | 09/17/2000 | 0231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/17/2000 | 1900 | 09/18/2000 | 0301 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2000 | 1400 | 07/01/2000 | 1835 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/18/2000 | 1430 | 09/18/2000 | 1531 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/18/2000 | 1800 | 09/19/2000 | 0401 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2000 | 1100 | 09/19/2000 | 1131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2000 | 1400 | 09/20/2000 | 0201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/20/2000 | 1230 | 09/20/2000 | 2131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/20/2000 | 2230 | 09/21/2000 | 0431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/22/2000 | 1730 | 09/23/2000 | 0301 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/25/2000 | 2330 | 09/26/2000 | 0031 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2000 | 1800 | 09/27/2000 | 1801 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2000 | 2030 | 09/28/2000 | 0031 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2000 | 1930 | 07/01/2000 | 2105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2000 | 1730 | 09/29/2000 | 0131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2000 | 0230 | 09/29/2000 | 0401 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2000 | 1730 | 09/30/2000 | 0431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/30/2000 | 1800 | 10/01/2000 | 0131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/2000 | 1830 | 10/02/2000 | 0501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/02/2000 | 1900 | 10/03/2000 | 0231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/03/2000 | 1900 | 10/04/2000 | 0501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2000 | 2030 | 10/09/2000 | 2031 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2000 | 2000 | 10/14/2000 | 2301 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/2000 | 1730 | 10/16/2000 | 2001 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2000 | 2200 | 07/01/2000 | 2335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/2000 | 2330 | 10/17/2000 | 0201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2000 | 0300 | 10/19/2000 | 0701 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/2000 | 0345 | 10/22/2000 | 1601 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/2000 | 1445 | 10/25/2000 | 1501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/26/2000 | 2015 | 10/26/2000 | 2201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/08/2000 | 2330 | 12/09/2000 | 0201 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 0900 | 12/10/2000 | 2001 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 0400 | 07/02/2000 | 0405 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 0530 | 07/02/2000 | 0805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 1000 | 07/02/2000 | 1135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 1300 | 07/02/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 1500 | 07/02/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2000 | 1700 | 07/04/2000 | 0035 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/2000 | 0430 | 07/04/2000 | 0835 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/2000 | 0930 | 07/04/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/2000 | 1500 | 07/04/2000 | 1505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/2000 | 1600 | 07/04/2000 | 1605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2000 | 2100 | 07/05/2000 | 2105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2000 | 2200 | 07/05/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2000 | 2300 | 07/06/2000 | 0135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 0600 | 07/06/2000 | 0605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 1030 | 07/06/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 1400 | 07/06/2000 | 2005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 2100 | 07/06/2000 | 2105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2000 | 2300 | 07/07/2000 | 0205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/07/2000 | 1600 | 07/07/2000 | 1805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/07/2000 | 1900 | 07/07/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/07/2000 | 2200 | 07/07/2000 | 2235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2000 | 0700 | 07/14/2000 | 0805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2000 | 1200 | 07/14/2000 | 1305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2000 | 2230 | 07/14/2000 | 2235 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2000 | 0100 | 07/15/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2000 | 0300 | 07/15/2000 | 1305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2000 | 1400 | 07/15/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/15/2000 | 2000 | 07/16/2000 | 0135 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 0330 | 07/16/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 0630 | 07/16/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 0830 | 07/16/2000 | 0835 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1000 | 07/16/2000 | 1205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1330 | 07/16/2000 | 1335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1530 | 07/16/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1630 | 07/16/2000 | 1635 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 1900 | 07/16/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 2000 | 07/16/2000 | 2005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2000 | 2100 | 07/16/2000 | 2205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/2000 | 0430 | 07/17/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0100 | 07/18/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0200 | 07/18/2000 | 0205 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0330 | 07/18/2000 | 0435 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0530 | 07/18/2000 | 0605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 0700 | 07/18/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/18/2000 | 1930 | 07/18/2000 | 1935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/19/2000 | 2330 | 07/20/2000 | 0935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2000 | 1900 | 07/20/2000 | 1905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2000 | 2300 | 07/20/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 0000 | 07/21/2000 | 0335 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 0430 | 07/21/2000 | 0505 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 0600 | 07/21/2000 | 0605 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 0830 | 07/21/2000 | 0905 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 1000 | 07/21/2000 | 1005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 1100 | 07/21/2000 | 1535 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 1700 | 07/21/2000 | 1705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2000 | 1800 | 07/21/2000 | 1935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2000 | 0630 | 07/22/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2000 | 1100 | 07/22/2000 | 1305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2000 | 2300 | 07/22/2000 | 2305 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 0100 | 07/23/2000 | 0105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 0200 | 07/23/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 0800 | 07/23/2000 | 0805 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 0900 | 07/23/2000 | 2005 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2000 | 2330 | 07/24/2000 | 0705 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2000 | 1300 | 07/24/2000 | 1405 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/28/2000 | 0930 | 07/28/2000 | 0935 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/28/2000 | 1100 | 07/28/2000 | 1105 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/29/2000 | 0130 | 07/29/2000 | 0131 |
Subject: | SGP/EBBR/E8 - E8 EBBR Right Air T (and thus fluxes) Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The right air temperature was incorrect, causing several measurements calculated using it, including sensible and latent heat fluxes, to be incorrect also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/31/2000 | 1630 | 07/31/2000 | 1656 |
Subject: | SGP/EBBR/E8 - E8 EBBR Data Incorrect During System Replacement |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The data was incorrect while preparations were made to replace it with a recalibrated unit. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/31/2000 | 2030 | 07/31/2000 | 2031 |
Subject: | SGP/EBBR/E8 - E8 EBBR Data Not Based On 30 Minute Average |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The 30 minute value was based on only 15 minutes of data. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 2330 | 12/11/2000 | 0731 |
Subject: | SGP/EBBR/E8 - E8 EBBR Wind Direction Frozen |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor was frozen into reading one direction during the indicated time period. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2000 | 0130 | 08/05/2000 | 0131 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/12/2000 | 0400 | 08/12/2000 | 0501 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/13/2000 | 0230 | 08/13/2000 | 0231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2000 | 0530 | 08/16/2000 | 0731 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2000 | 1730 | 08/16/2000 | 1801 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2000 | 2100 | 08/16/2000 | 2231 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2000 | 0530 | 08/17/2000 | 0531 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2000 | 0700 | 08/17/2000 | 0801 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2000 | 1430 | 08/17/2000 | 1431 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2000 | 1800 | 08/17/2000 | 1901 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/24/2000 | 1830 | 08/24/2000 | 1931 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2000 | 0100 | 08/25/2000 | 0331 |
Subject: | SGP/EBBR/E8 - E8 EBBR SHF1 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow probe #1 was offscale, large negative, or large positive during the indicated period. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2000 | 1900 | 04/05/2000 | 1901 |
04/05/2000 | 2000 | 04/05/2000 | 2001 |
Subject: | SGP EBBR E22 - E22 EBBR Data Incorrect During Logger Replacement |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
Description: | The data was incorrect during replacement of the logger. |
Measurements: | sgp5ebbrE22.00:
sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/21/2000 | 1900 | 04/27/2000 | 1031 |
05/10/2000 | 2000 | 06/11/2000 | 0531 |
07/08/2000 | 1800 | 10/16/2000 | 0231 |
Subject: | SGP EBBR E22 - E22 EBBR Soil Moisture Lower Than Measurement Capability |
DataStreams: | sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The indicated measurements were incorrect because soil moisture dropped below (sometimes even negative) the ability of the probes to measure it. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/2000 | 0235 | 10/22/2000 | 0316 |
Subject: | SGP EBBR E22 - E22 EBBR Soil Moisture #3 Lower Than Can Be Measured |
DataStreams: | sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The soil moisture was too low to be measured correctly by probe #3. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/16/2000 | 0130 | 10/18/2000 | 1841 |
Subject: | SGP EBBR E22 - E22 EBBR Net Radiation Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp5ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Water inside the domes of the net radiometer caused measurements from it to be incorrect. Calculated latent and sensible heat fluxes were therefore incorrect also. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/2001 | 1815 | 11/06/2001 | 1931 |
04/24/2001 | 1700 | 04/24/2001 | 1746 |
Subject: | SGP EBBR E8 - E8 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/18/2001 | 1715 | 04/18/2001 | 1821 |
10/10/2001 | 1800 | 10/10/2001 | 1857 |
Subject: | SGP EBBR E22 - E22 EBBR Data Incorrect During 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/2001 | 2000 | 03/07/2001 | 2001 |
Subject: | SGP EBBR E22 - E22 EBBR 30 Minute Data Incorrect After Missing Data |
DataStreams: | sgp30ebbrE22.b1 |
Description: | Complete soil temperature information is not available when the data is missing for the previous half hour. Since soil temperature difference with time cannot be computed properly, the selected calculated quantities are incorrect. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2000 | 2015 | 03/31/2001 | 2356 |
Subject: | SGP EBBR E8 - E8 EBBR Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil Heat Flow #1 was frequently large negative or large positive and was incorrect. The indicated measurements were also incorrect as a result. Correct sensible and latent heat fluxes can be determined by using the other four sets of soil measurements to determine the energy balance. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 12/14/2001 | 2356 |
Subject: | SGP EBBR E8 - E8 EBBR Soil Heat Flow #1 Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Soil heat flow #1 was usually large negative or positive, and therefore incorrect. Sensible and latent heat fluxes can be recalculated by using the four good sets of soil measurements for the energy balance. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2001 | 2145 | 02/13/2001 | 1830 |
Subject: | SGP EBBR E8 - E8 EBBR AEM Frozen In One Position |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The AEM was frozen in one position by an ice storm for 5 days. The indicated measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/08/2001 | 1615 | 02/11/2001 | 1745 |
Subject: | SGP EBBR E8 - E8 EBBR Wind Speed Frozen |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The wind speed sensor was frozen stopped by an ice storm. The indicated measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/13/2001 | 0600 | 03/31/2001 | 2356 |
Subject: | SGP EBBR E8 - E8 EBBR Battery Voltage Low, Data Incorrect |
DataStreams: | sgp5ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.00, sgp15ebbrE8.b1 |
Description: | The battery voltage was frequently less than 10.5 volts. When the battery voltage is 10.4 V or less, all of the data is suspect. Examine the battery voltage measurement in the 15 minute data to determine whether the measurements can be used. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 12/14/2001 | 2356 |
Subject: | SGP EBBR E8 - E8 EBBR Battery Voltage Low, Data Incorrect |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.00 |
Description: | The battery voltage was frequently less than 10.5 volts. When the voltage is 10.4 V or less, all of the measurements are suspect and should not be used (they are likely incorrect). |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/02/1994 | 1640 | 11/02/1994 | 1900 |
Subject: | SGP/EBBR/E8 - T/RH Check |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1640, 1645, 1650, 1655, 1700, 1705, 1710, 1715, 1720, 1725, 1730, 1735, 1740, 1745, 1750, 1755, 1800, 1805, 1810, 1815, 1820, 1825, 1830, 1835 15 minute: 1645, 1700, 1715, 1730, 1745, 1800, 1815, 1830, 1845 30 minute: 1700, 1730, 1800, 1830, 1900 (EDITOR'S NOTE: The following analysis referrs to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/08/1994 | 1650 | 11/08/1994 | 1800 |
Subject: | SGP/EBBR/E22 - T/RH Check |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During the month of November 1994, site operations personnel conducted field comparisons of the temperature and relative humidity sensors in the EBBRs and the HMI31 portable T/RH meter (with the removable probe). As is reflected in the home signal output, the automatic exchange mechanism was disabled in order to conduct the checks. Measurements were taken every 5 minutes, for a half hour each in the aspirated radiation shields (the HMI31 was inserted into the EBBR aspirated radiation shield to provide the same aspiration as the EBBR sensors received). Measurements of aspirator flow were also recorded. This information was provided to the mentor. The mentor then compared the field checks with the data recorded by the EBBRs. The periods of data and data fields that are incorrect because of the checks are listed below. Times are those at which the data are reported. Data Fields 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h 5 minute: 1650, 1655, 1700, 1705, 1710, 1715, 1720, 1725, 1730, 1735 15 minute: 1700, 1715, 1730, 1745 30 minute: 1700, 1730, 1800 (EDITOR'S NOTE: The following analysis refers to all EBBRs installed in Nov 1994) Results of the T/RH comparisons: The combined accuracy of the HMI31 and EBBR temperature sensors (thermocouples or PRTDs) is +/- 0.4 degrees C. The combined accuracy of the HMI31 and EBBR RH sensors is +/- 4% for RH less than 80%, and +/- 6% for RH greater than 80%. The combined accuracy for each pair of EBBR sensors of the same type is essentially the same as the quantities stated above. All of the pairs of EBBR sensors show differences that fall within the stated combined accuracies. Although this sounds wonderful, a few of the differences are close to the combined accuracy, which, because of the small differences in temperature being measured (normally smaller than the magnitude of the combined accuracy) could lead to some incorrect estimation of sensible and latent heat flux. However, unless there would be substantial differences of calibration slope of two sensors in a pair, the switching of height of the sensors every 15 minutes by the automatic exchange mechanism will remove the offset between the sensors. The comparisons between the EBBR sensors and the HMI31 T/RH meter showed much larger differences. The range of differences were: -2.4 to 0.0 degrees for the thermocouples, -2.1 to 0.1 degrees for the RH probe PRTDs, and -6.6 to 6.7 percent for the RH sensors. Six Thermocouples indicated differences from the HMI31 greater than the combined accuracy. Seven PRTDs indicated differences from the HMI31 greater than the combined accuracy. Four RH sensors indicated differences from the HMI31 greater than the combined accuracy. The temperature differences (for both thermocouples and PRTDs) were nearly all of one sign; the HMI31 indicated the greater temperature, which may suggest a bias in the HMI31 temperature measurement, even though a dozen differences of +/- 0.2 degrees were measured. Four differences of zero and only one positive difference (out of 40) were measured. Thirteen of the 20 RH differences were positive (HMI31 indicating the lower RH), with the differences near the level of the combined accuracies being approximately 50% positive and 50% negative. It does not appear that the HMI31 has a RH bias. Plots of EBBR sensor/HMI31 differences as a function of time from the installation of the individual sensors (a range of 10 to 30 months in the field) shows no obvious pattern of sensor drift with time. This is encouraging. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/1995 | 1630 | 10/11/1995 | 1930 |
Subject: | SGP/EBBR/E22 - 3-month Check |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/02/1995 | 1930 | 10/02/1995 | 2230 |
Subject: | SGP/EBBR/E8 - 3-month Check |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/15/2002 | 1810 | 05/15/2002 | 1846 |
Subject: | SGP/EBBR/E22 - 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/15/2001 | 0000 | 01/15/2002 | 2001 |
Subject: | SGP/EBBR/E8 - Low Battery Voltage Causes Incorrect Measurements |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.00, sgp5ebbrE8.b1 |
Description: | All measurements should be considered to be suspect when the battery voltage (see the 15 minute data) is less than 10.5 volts. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/07/1996 | 1901 | 11/07/1996 | 1930 |
11/06/1996 | 1300 | 11/06/1996 | 2300 |
Subject: | SGP/EBBR/E22 - Annual Calibration Changeout |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | On 6 November 1996 the EBBR changeout for annual calibration occurred at E22, Cordell, OK. SEBS unit 7 was removed and unit 3 was installed. Between 1300 and 2300 GMT no data or incorrect data was collected; most of the data is missing. Data for half hour 1930 GMT on 7 November is incorrect; 3 month calibration checks were being performed. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/15/2002 | 2015 | 01/28/2002 | 1801 |
02/26/2002 | 0230 | 02/26/2002 | 1831 |
02/27/2002 | 0315 | 03/26/2002 | 1731 |
04/06/2002 | 0000 | 04/09/2002 | 1731 |
01/30/2002 | 0445 | 02/12/2002 | 1801 |
Subject: | SGP/EBBR/E8 - AEM Hung |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The indicated data is incorrect while the automatic exchange mechanism was hung in one position. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/21/2002 | 1730 | 05/21/2002 | 1801 |
Subject: | SGP/EBBR/E8 - 6 Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2002 | 1845 | 02/07/2002 | 1901 |
Subject: | SGP/EBBR/E22 - All Measurements Zero |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.00, sgp5ebbrE22.b1 |
Description: | The measurements are often all zero the first period after a CR10 restart, as occurred here. |
Measurements: | sgp5ebbrE22.00:
sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2002 | 0515 | 01/15/2002 | 0031 |
02/08/2002 | 0845 | 02/08/2002 | 2031 |
01/09/2002 | 1545 | 01/12/2002 | 0001 |
02/08/2002 | 2145 | 02/14/2002 | 2001 |
02/14/2002 | 2015 | 02/16/2002 | 1531 |
02/16/2002 | 1945 | 02/17/2002 | 0001 |
02/17/2002 | 0415 | 02/20/2002 | 0531 |
02/20/2002 | 0815 | 02/20/2002 | 1831 |
02/28/2002 | 1030 | 03/06/2002 | 1901 |
04/21/2002 | 2000 | 05/29/2002 | 1831 |
12/17/2001 | 1230 | 01/09/2002 | 1531 |
Subject: | SGP/EBBR/E22 - AEM Hung or High Home Signals |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The automatic exchange mechanism hung some times and continued to switch at other times, although with high home signal values. If home signal values are not in the proper ranges, the gradient calculation is not performed properly by the CR10 program. Affected flux measurements cannot be recalculated. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 06/18/2003 | 2359 |
Subject: | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR |
DataStreams: | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE18.b1, sgp30ebbrE19.b1, sgp30ebbrE2.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
Description: | The units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in the EBBR 30 minute netcdf header have been incorrect from the beginning of processing the data into netcdf files. The correct units are MJ/m^3/C. This does not affect the data quality. |
Measurements: | sgp30ebbrE2.b1:
sgp30ebbrE26.b1:
sgp30ebbrE9.b1:
sgp30ebbrE22.b1:
sgp30ebbrE13.b1:
sgp30ebbrE7.b1:
sgp30ebbrE18.b1:
sgp30ebbrE12.b1:
sgp30ebbrE4.b1:
sgp30ebbrE15.b1:
sgp30ebbrE19.b1:
sgp30ebbrE20.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/1993 | 1200 | 06/18/2003 | 1159 |
Subject: | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR |
DataStreams: | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
Description: | The units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in the EBBR 30 minute netcdf header have been incorrect from the beginning of processing the data into netcdf files. The correct units are MJ/m^3/C. This does not affect the data quality. (NOTE: The actual begin date of some of these data streams is as early as 7/4/93. This DQR applies to all 30m EBBR data.) |
Measurements: | sgp30ebbrE26.b1:
sgp30ebbrE9.b1:
sgp30ebbrE7.b1:
sgp30ebbrE4.b1:
sgp30ebbrE12.b1:
sgp30ebbrE15.b1:
sgp30ebbrE22.b1:
sgp30ebbrE13.b1:
sgp30ebbrE8.b1:
sgp30ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2002 | 1900 | 11/13/2002 | 1930 |
Subject: | SGP/EBBR/E22 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/05/2002 | 1930 | 11/05/2002 | 2000 |
Subject: | SGP/EBBR/E8 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/2002 | 1230 | 11/05/2002 | 2000 |
Subject: | SGP/EBBR/E8 - AEM Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | When the AEM (Automatic Exchange Mechanism) is hung inbetween a home position, the gradients of temperature and relative humidity, and thus sensible and latent heat fluxes, are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/2002 | 0230 | 10/25/2002 | 1300 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind speed sensor was frozen stopped by ice. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/12/2002 | 0000 | 08/12/2002 | 1730 |
Subject: | SGP/EBBR/E8 - Water in broken Net Rad Dome |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer had water in the bottom, resulting in incorrect measurements of net radation. Calculated Sensible (h) and Latent Heat (e) Flux are therefore incorrect also. |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/28/2002 | 0945 | 10/28/2002 | 1530 |
10/28/2002 | 1630 | 10/28/2002 | 1700 |
10/28/2002 | 2130 | 11/04/2002 | 0200 |
Subject: | SGP/EBBR/E8 - Battery Voltage Low |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | When Battery Voltage drops below 10.5 volts, most measurements are suspect and some are often incorrect. The user will need to inspect the data carefully to determine the usability of the data. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/04/2002 | 1715 | 09/04/2002 | 1730 |
10/13/2002 | 1300 | 10/15/2002 | 1930 |
10/15/2002 | 2145 | 10/15/2002 | 1630 |
10/13/2002 | 1230 | 10/13/2002 | 1234 |
Subject: | SGP/EBBR/E22 - AEM Hung between home positions |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | When the AEM (Automatic Exchange Mechanism) is hung inbetween home positions, the gradients of temperature and relative humidity, and thus sensible and latent heat fluxes, are incorrect. There is no work-around to this problem. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/29/2002 | 0830 | 08/07/2002 | 1800 |
Subject: | SGP/EBBR/E22 - Water inside Broken Net Rad Dome |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The net radiation measurement is incorrect. Measurements calculated using net radiation are also incorrect. No work-around is possible for this situation. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2003 | 0000 | 06/25/2003 | 1700 |
Subject: | SGP/EBBR/E22 - home30 too large |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The home30 measurements sometimes exceeded 34.9999, resulting in the qc flag being set. However, this does not create any data quality issues. The qc flag is set just to provide an alert that the automatic exchange mechanism may have a problem; in this case data was not affected. An electronics repair solved the problem. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/2003 | 1750 | 05/06/2003 | 1825 |
Subject: | SGP/EBBR/E8 - 6 Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2003 | 1720 | 04/30/2003 | 1800 |
Subject: | SGP/EBBR/E22 - 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/16/2003 | 1245 | 01/28/2003 | 1915 |
02/06/2003 | 1400 | 02/11/2003 | 1930 |
02/23/2003 | 0945 | 03/11/2003 | 1800 |
03/19/2003 | 1900 | 03/25/2003 | 1800 |
Subject: | SGP/EBBR/E8 - AEM Hung Inbetween Home Positions |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The AEM was hung inbetween home positions, causing many measurements to be incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/16/2003 | 0000 | 03/22/2003 | 1445 |
Subject: | SGP/EBBR/E8 - Battery Voltage Low |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Many of the measurements are incorrect much of the time during this period, when the battery voltage was 10.5 volts or less. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/12/2003 | 1800 | 09/02/2003 | 1800 |
Subject: | SGP/EBBR/E8 - Pressure Sensor Measurements Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgpebbrE8.01 |
Description: | The EBBR barometer did not agree with the SMOS at all times. However this problem only rarely affected the sensible and latent heat flux values. Replacement of the barometer on 29 july 2003, 12 aug 2003, and 26 aug 2003 caused some obvious disruption in data flow. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgpebbrE8.01:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2003 | 0330 | 07/23/2003 | 1830 |
07/24/2003 | 1500 | 08/06/2003 | 1730 |
Subject: | SGP/EBBR/E22 - AEM Hung Inbetween Home Positions |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The AEM was hung inbetween home positions. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2003 | 1745 | 09/17/2003 | 1900 |
Subject: | SGP/EBBR/E22 - AEM Home Signals Too High |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The AEM home signals were too large at times during the indicated period; it has been determined by the vendor that this is usually caused by electrical noise in the vicinity. When the 30 minute data shows a home 30 signal of 35 or more, the bowen ratio, and sensible and latent heat flux measurements are incorrect Removing the jumper between 4L and AG on the CR10 and adding a jumper to Multiplexer 2, Set 6, between L1 and L2 fixed the problem. The vendor intends to make this modification to all EBBR systems as they return for re-calibration. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/27/2003 | 0230 | 09/09/2003 | 1730 |
Subject: | SGP/EBBR/E8 - Water in Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2003 | 1800 | 10/01/2003 | 1630 |
Subject: | SGP/EBBR/E22 - Water in Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/2003 | 1900 | 10/15/2003 | 1930 |
Subject: | SGP/EBBR/E22 - 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/07/2003 | 1700 | 10/07/2003 | 1735 |
Subject: | SGP/EBBR/E8 - 6 Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2003 | 1400 | 12/14/2003 | 2200 |
Subject: | SGP/EBBR/E8 - Net Radiometer Iced Over During Storm |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer was covered with ice from a storm. Net radiation values are incorrect, and are sometimes even negative during the daytime. Sensible and latent heat fluxes are therefore incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2003 | 1810 | 12/13/2003 | 2100 |
Subject: | SGP/EBBR/E22 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2003 | 2345 | 12/14/2003 | 1520 |
Subject: | SGP/EBBR/E22 - Wind Direction Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2003 | 0000 | 03/17/2004 | 1835 |
Subject: | SGP/EBBR/E22 - Wind Speed Sensor Bearings Worn |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Wind speed was too low and/or often at or near zero due to worn bearings in the wind speed sensor. Sensor was replaced 3/17/04. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/24/2004 | 2345 | 01/25/2004 | 0135 |
02/01/2004 | 1920 | 02/02/2004 | 1740 |
02/03/2004 | 0350 | 02/03/2004 | 1440 |
02/05/2004 | 0925 | 02/05/2004 | 1615 |
Subject: | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/05/2004 | 1120 | 02/05/2004 | 1500 |
Subject: | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/31/2004 | 1910 | 02/04/2004 | 1750 |
02/05/2004 | 1020 | 02/06/2004 | 0820 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/25/2004 | 2300 | 01/26/2004 | 2135 |
01/27/2004 | 0425 | 01/27/2004 | 0630 |
01/27/2004 | 1130 | 01/27/2004 | 1740 |
Subject: | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/26/2004 | 0710 | 02/26/2004 | 1410 |
Subject: | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped By Icing |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2004 | 0000 | 04/06/2004 | 1755 |
Subject: | SGP/EBBR/E8 - AEM Replacement |
DataStreams: | sgp15ebbrE8.b1, sgp30baebbrE8.c1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The AEM at E8 kept hanging in between home positions. This DQR covers only the period during which the AEM was being replaced. DQRs generally are not issued for AEMs that are not working as qc flags indicate this clearly. |
Measurements: | sgp15ebbrE8.b1:
sgp30baebbrE8.c1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/22/2004 | 0400 | 06/23/2004 | 1650 |
Subject: | SGP/EBBR/E22 - Net Radiation Incorrect |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Rain filled the bottom of the net radiometer with water, resulting in very high net radiations. Net radiation is incorrect during this period, also causing sensible and latent heat fluxes to be incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2004 | 1330 | 07/07/2004 | 1900 |
Subject: | SGP/EBBR/E22 - Water In Net Radiometer |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Water entered the Net Radiometer through a cracked upper dome, causing the net radiation values to be too large. As a result, latent and sensible heat flux values are also incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/30/2004 | 1200 | 09/07/2004 | 1730 |
Subject: | SGP/EBBR/E8 - Net Radiation Incorrect, Fluxes Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer dome was broken and water entered the botom dome, causing net radiation measurements, and thus sensible and latent heat flux measurements, to be incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/27/2004 | 1700 | 10/27/2004 | 1730 |
Subject: | SGP/EBBR/E22 - 6 Month Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2004 | 1730 | 10/19/2004 | 1800 |
Subject: | SGP/EBBR/E8 - 6 Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0845 | 01/07/2005 | 1555 |
Subject: | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2005 | 0810 | 01/04/2005 | 2100 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2005 | 0855 | 01/06/2005 | 1940 |
Subject: | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2005 | 0850 | 01/04/2005 | 2100 |
Subject: | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2005 | 0720 | 01/12/2005 | 1605 |
02/07/2005 | 0200 | 02/08/2005 | 1930 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen Stopped By Ice Storm |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2005 | 0500 | 02/08/2005 | 0330 |
Subject: | SGP/EBBR/E8 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2005 | 0915 | 01/11/2005 | 2100 |
Subject: | SGP/EBBR/E22 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2005 | 1730 | 01/11/2005 | 1825 |
Subject: | SGP/EBBR/E22 - Wind Direction Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind direction sensor was frozen stopped by ice from a storm. The wind direction, sigma_wd, and residual wind speed measurements are incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/06/2005 | 1400 | 03/16/2005 | 1900 |
Subject: | SGP/EBBR/E22 - Net Radiation Too Low - Top Dome Broken |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Net radiation values were too low because of a broken top dome. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/21/2005 | 0230 | 04/26/2005 | 2215 |
Subject: | SGP/EBBR/E8 - Battery Voltage Problem |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Battery voltage was too low to give correct data. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/30/2005 | 1930 |
Subject: | SGP/EBBR/E22 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE22.b1 |
Description: | Effective 20050330.1930, the EBBR.E20 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050330, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE20.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE20.a1). |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/22/2005 | 1730 |
Subject: | SGP/EBBR/E8 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE8.b1 |
Description: | Effective 20050322.1730, the EBBR.E8 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050322, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE8.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE8.a1). |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/27/2005 | 2000 | 08/31/2005 | 1730 |
Subject: | SGP/EBBR/E22 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The net radiometer top dome was broken and water got inside, making the net radiation, and sensible and latent heat flux measurements incorrect. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/2005 | 1800 | 03/22/2005 | 1730 |
Subject: | SGP/EBBR/E8 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE8.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/16/2005 | 1830 | 03/30/2005 | 1930 |
Subject: | SGP/EBBR/E22 - Average Soil Heat Flux Incorrect |
DataStreams: | sgp30ebbrE22.b1 |
Description: | A typo in the version 9 EBBR program caused average soil heat flow to be zero, resulting in both sensible and latent heat fluxes being approximately 20 watts per meter squared too large during the middle of the day. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/2005 | 1930 | 06/28/2005 | 1930 |
Subject: | SGP/EBBR/E8 - Soil Heat Flow #2 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Heat Flow #2 spiked upwards briefly at 1930 GMT. |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/04/2005 | 2000 | 08/09/2005 | 2030 |
Subject: | SGP/EBBR/E8 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The top dome of the net radiometer was cracked and water had gotten inside it, so net radiation, and sensible and latent heat fluxes were incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E8 - metadata corrections |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E8 data collected by ARM back to the installation of the instrument in December 1992. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E22 - metadata corrections |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E22 data collected by ARM back to the installation of the instrument in April 1993. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil type 4) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/26/2005 | 1900 | 01/10/2006 | 1630 |
Subject: | SGP/EBBR/E8 - AEM Not Exchanging |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The AEM hung in one position frequently and sometimes for many days during this period While the AEM was not switching, the sensible and latent heat fluxes and gradient measurements were incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/10/2006 | 0730 | 01/18/2006 | 2000 |
Subject: | SGP/EBBR/E22 - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/12/2006 | 1900 | 05/24/2006 | 1700 |
Subject: | SGP/EBBR/E22 - Fluxes Incorrect When Soil Heat Flow #3 Out of Range |
DataStreams: | sgp30ebbrE22.b1 |
Description: | When soil heat flow #3 exceeded the range limit in the EBBR program, it was replaced with 9999es. This caused the ave_shf and the sensible and latent heat fluxes to be incorrect. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/21/2006 | 1130 | 01/21/2006 | 1700 |
01/22/2006 | 0100 | 01/22/2006 | 0730 |
Subject: | SGP/EBBR/E22 - Wind Speed Sensor Frozen Stopped |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind speed sensor was frozen to a stopped condition. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/2005 | 0000 | 01/24/2006 | 2330 |
Subject: | SGP/EBBR/E8 - Flux and Gradient Measurements Suspect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The flux and gradient measurements were often incorrect when the AEM did not function properly. The user needs to inspect the data (particularly the home signals) to determine when the AEM was and was not working. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/21/2006 | 0230 | 03/21/2006 | 1600 |
Subject: | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Freezing rain froze the wind speed sensor to a stopped condition. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2006 | 2030 | 04/04/2006 | 1500 |
Subject: | SGP/EBBR/E8 - Fluxes Incorrect; Left Humidity Near Zero |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The left humidity was intermittently incorrect, therefore the sensible and latent heat fluxes were also. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/2006 | 1030 | 05/31/2006 | 0030 |
Subject: | SGP/EBBR/E8 - Sensible and Latent Heat Flux Suspect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The left humidity sensor produced large spikes frequently. This resulted in incorrect measurements of vapor pressure gradient and thus calculations of sensible and latent heat flux. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/23/2006 | 0130 | 06/27/2006 | 1730 |
Subject: | SGP/EBBR/E8 - Sensible and Latent Heat Fluxes and Net Radiation Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer top dome was broken and water had gotten into it. Sensible and latent heat fluxes and net radiation were therefore incorrect. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/21/2006 | 0630 | 08/30/2006 | 1800 |
Subject: | SGP/EBBR/E22 - Net Radiation and Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | Rain entered the net radiometer, causing net radiation, and thus sensible and latent heat fluxes, to be too large. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/16/2006 | 0000 | 11/08/2006 | 1940 |
Subject: | SGP/EBBR/E22 - Soil Heat Flow #4 Incorrect - Sensible and Latent Heat Fluxes Sometimes |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil heat flow sensor #4 dropped to low values or offscale. Sensible and latent heat fluxes were sometimes affected. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/04/1993 | 1200 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E22 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE22.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-49, 139-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/12/1993 | 1630 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E8 - Wind Direction Dependency of Surface Fluxes and General Data Quality |
DataStreams: | sgp30ebbrE8.b1 |
Description: | Beginning in FY2006, Data Quality Reports are not written for missing data or for situations when qc flags clearly show that the data is incorrect (this is true for most of the conditions listed below). DQRs are written for periods when data is incorrect, when the situation is not represented by qc flags in the data, and it's not obvious that the data should have been flagged as incorrect. The EBBR data is only useful for particular wind directions at each Extended Facility. Please use the following resources to help in interpreting the EBBR data. 1) qc flags in the EBBR data. 2) Instrument Mentor Monthly Summaries (IMMS) at http://www.db.arm.gov/IMMS/ 3) SGP EF Surface Conditions Observations (EFSCO) at http://198.124.96.210:591/sfc_cond1/default.htm 4) Surface fluxes are valid only for the following wind directions (in degrees): 0-224, 314-360 Note that there are wind directions at most sites for which the fetch is insufficient and therefore EBBR data is invalid. Appropriate fetch was determined from a 1/40 measurement height to fetch ratio, resulting in a required minimum fetch of 120 m. Particular attention should be paid to the limited valid wind directions for the Central Facility EBBR. The surface vegetation type for the EBBRs is grazed or ungrazed grassland (refer to the EFSCO for state). 5) Some conditions that commonly cause the EBBR primary measurements to be incorrect or missing follow: a) Sensible and latent heat fluxes are not accurate during times when the Automatic Exchange Mechanism (AEM) is not functioning properly. The AEM switches the gradient measuring instrumentation between the top and bottom positions every 15 minutes; this reduces the effects of instrument offsets. Sometimes the AEM does not reach it's full extent of travel, resulting in the home signal being zero. b) Sensible and latent heat fluxes are sometimes incorrect when surface soil heat flow is out of range, as seen in the average soil heat flux (ave_shf). c) Very light winds may be seen at nights for brief periods. Wind direction flops around a lot during low wind conditions (< 1.5 m/s) and is often unreliable during those periods. d) Missing data periods occur at times; this is usually a site data system collection/communication problem; it can also occur during replacement of the system with a newly calibrated unit. Missing data is sometimes filled in later from manual or automatic re-collection of the data. |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2007 | 1630 | 03/06/2007 | 2100 |
Subject: | SGP/EBBR/E8 - Failed Temp/RH sensor |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The right Temperature-Humidity sensor failed. The measured temperature gradients were incorrect and thus the sensible and latent heat fluxes were incorrect for this extended period. The sensor was replaced March 6. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2007 | 0400 | 04/26/2007 | 1630 |
Subject: | SGP/EBBR/E22 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Wind direction was flat-lining at 135 degrees and occasionally changed to 0.4 degrees. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2007 | 0630 | 08/01/2007 | 2200 |
Subject: | SGP/EBBR/E22 - H and LE Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil Heat Flow 1 bounced around and went offscale at times, resulting in the EBBR program setting sensible and latent heat fluxes to 9999es. The problem ended during a PM visit, possibly because of routine checks and tightening of leads in terminal strips. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/2008 | 0330 | 05/28/2008 | 1745 |
Subject: | SGP/EBBR/E8 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction flatlined at times at 135 degrees (an indication of failed electronics) and at other times trended correctly with the co-located SMOS but differed from it by 15 to 100 degrees. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2009 | 1230 | 01/03/2009 | 1500 |
Subject: | SGP/EBBR/E8 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind speed sensor was frozen stopped. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/26/2009 | 1845 | 01/28/2009 | 1810 |
Subject: | SGP/EBBR/E22 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The wind speed sensor was frozen stopped by freezing rain. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/10/2009 | 0930 | 08/19/2009 | 1800 |
Subject: | SGP/EBBR/E8 - Insufficient voltage |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Low DC supply voltage caused many measurements to be incorrect during this period. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/22/2007 | 1830 | 11/10/2009 | 1430 |
Subject: | SGP/EBBR/E8 - RH Maximum too high |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The absolute RH measurements were too high for RH > 100%. Use of first (instead of second order) coefficients was no improvement. The calculated sensible and latent heat fluxes were unaffected. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/22/1993 | 1630 | 11/03/1993 | 1730 |
Subject: | SGP/EBBR/E8 - Drop in atmospheric pressure |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Beginning September 22, 1993, the atmospheric pressure sensor at the E8 EBBR began showing drops of as much as 5 kPa for about half a day. These could be easily seen in comparisons with the SMOS pressure sensor. The drops were diurnal and occurred during daytime, at the same time as other quantities such as temperature, net radiation, and battery voltage were at the highest levels of the day. It is unclear what exactly caused this situation. Replacement of the Campbell multiplexer to which the pressure sensor was connected caused the problem to go away. The same multiplexer was then used as a replacement for the other multiplexer in the E8 EBBR and has not shown any odd behavior on the channel that previously served the pressure sensor. My conclusion is that there was poor electronic contact of the pressure sensor to the multiplexer. Affected pressure data can only be easily seen by comparison with SMOS pressure data. No EBBR data other than pressure is affected by the pressure problem; calculations of specific heat at constant pressure (which includes the term 1/(p-vp)) times other terms (including p-.378*vp) result in a very small error in the resulting Bowen Ratio. For example; assume a common vapor pressure (vp) of 2.0 kPa, a common atmospheric pressure of 97.0 kPa, and a corrupted atmospheric pressure of 92.0 kPa. Uncorrupted Corrupted (97.0-.378*2.0)/(97.0-2.0) = 1.0131 (92.0-.378*2.0)/(92.0-2.0) = 1.0138 1.0138/1.0131 = 1.00069 or 0.069% increase in the Bowen Ratio as a result of the corruption Therefore, no correction to the quantities calculated using pressure is warranted. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/1993 | 0000 | 12/21/1993 | 1800 |
Subject: | SGP/EBBR/E22 - Air Temperature Sensor malfunction |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | The right air temperature probe (thermocouple) malfunctioned during the periods stated above at Cordell, OK (E22). It was replaced on December 21, 1993 and good data began again with the 1830 GMT half hour average. For the periods of interest, 5 and 30 minute values of tair_top and tair_bot are incorrect when the right probe was in the specified position; 15 minute values of tair_r are incorrect; and 30 minute values of bowen, e, and h are incorrect. None of these values are recoverable or recalculatable through the use of other quantities since both air temperature probes are needed to determine bowen ratio. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/09/1993 | 1700 | 11/22/1993 | 1700 |
Subject: | SGP/EBBR/E22 - Incorrect averaging interval at E22 |
DataStreams: | sgp30ebbrE22.b1 |
Description: | On November 9, 1993 version 4 of the EBBR program installed at Cordell, OK (E22) was mistakenly modified in one area during the process of modifying a calibration in another part of the program. The mistaken modification was a change in the execution interval from 30 seconds to 10 seconds. This modification remained in the program until November 22, 1993 when it was discovered and changed back to a 30 second execution interval. The modification produced no ill effects as the execution interval only changes sampling and averaging periods. A 10 second execution interval results in three times as many samples being taken, which, when averaged, should actually produce more accurate results. Therefore, data quality was not affected. This PIF was written only to document the fact that the mistaken modification was made, and when it was made. |
Measurements: | sgp30ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/1993 | 1100 | 12/29/1993 | 1700 |
Subject: | SGP/EBBR/E8 - Invalid data due to AEM fuse blow out at E8 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On December 17, 1993, before 1100 GMT, the fuse in the AEM (Automatic Exchange Mechanism) of Coldwater, KS (E8) EBBR blew out. The AEM remained nonfunctional until the fuse was replaced just before 1700 GMT on December 29, 1993. During the 1000 GMT half hour on December 17 the AEM was hung up; this is clear from the home signal. These situations resulted in the following invalid data: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, home 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, mv_home, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h, home_15, home_30 Bowen, h, and e can not be recalculated from the 15 minute and/or 5 minute data since we cannot determine the position of the sensors on the AEM from the home signal; home signal is near zero when there is no power to the AEM. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/1994 | 1100 | 01/12/1994 | 1630 |
Subject: | SGP/EBBR/E8 - Invalid wind data frozen sensor |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On January 12, 1994 the EBBR wind speed sensor at E8, Coldwater, KS was frozen to a stationary by ice. Invalid data for the time period 1100 to 1630 GMT is: 30 minute: wind_s, res_ws 15 minute: wind_s 5 minute: wind_s, res_ws. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/1994 | 2330 | 02/23/1994 | 1800 |
Subject: | SGP/EBBR/E8 - Invalid wind speed frozen sensor |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | On February 21, 1994 the EBBR wind speed sensor at E8, Coldwater, KS was frozen to a stationary condition by ice. Invalid data for the time period indicated is: 30 minute: wind_s, res_ws 15 minute: wind_s 5 minute: wind_s, res_ws. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/19/1994 | 1900 | 08/24/1994 | 1600 |
Subject: | SGP/EBBR/E8 - Battery Voltage Drop |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Sometime after 0000 GMT on August 19, 1994 the cable from the solar panel to the EBBR battery was broken, presumably the result of severe weather that was in the area. The result was that the battery could not be charged by the solar panel, leading to a gradual reduction in battery voltage. By 1730 GMT on the 19th, the battery voltage had dropped to 10.8 volts and at about 1905 GMT had dropped too low to power the AEM. Some thirty minute data after 1900 GMT is therefore incorrect. By 2000 GMT the voltage had dropped to 10.7 volts and was too low to properly power the barometric pressure sensor (it's output dropped off from then on). At 1530 GMT on August 24, 1994 the solar panel connection to the battery was reestablished and normal data acquisition resumed at 1600 GMT. The last data file to be produced was apparently 0000 GMT (15 minute file) on 20 August. In an apparently related incident, soil temperature #3 skyrocketed in output beginning at about 1815 GMT; the cause for this is unknown, but is probably a result of the inability of the CR10 data logger to correctly measure the sensor resistance. By 1900 GMT this situation had caused elevated average soil heat flux, which in turn affected calculated e and h. Data fields that are incorrect for the time period listed, follow: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/01/1994 | 1700 | 09/01/1994 | 2230 |
09/03/1994 | 1530 | 09/07/1994 | 1700 |
Subject: | SGP/EBBR/E8 - Data Corruption and Loss |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Data loss was experienced during the periods indicated for unknown reasons. My understanding is that ingestion of the data failed for these periods because the record length was shorter than it is supposed to be. Site Operations personnel visited the site on September 7 and powered down and then up, the EBBR, essentially causing a reload of the CR10 program. All output locations had been -9999 or 0000 when they had arrived. The battery condition was normal. Failure of a portion of the storage module may be indicated; data sneakernetted from the storage module shows an alternating sequence of two days, September 2 and 3 and then a stretch of garbage characters. No other days of data are found. The portion of the storage module memory that contained the program was not corrupted, as the CR10 program compiled and ran properly when it was reloaded from the storage module. The data after the period above is reasonable. The last part of the storage module memory may be bad and perhaps in trying to access it remotely multiple times the situation above resulted. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/13/1994 | 1600 | 09/27/1994 | 1600 |
Subject: | SGP/EBBR/E22 - Net Radiometer Hemisphere Changes |
DataStreams: | sgp5ebbrE22.b1, sgp30ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | The upper hemisphere of the net radiometer at the site was punctured allowing water to collect in the lower (and sometimes upper) section of the sensor. The punctures are most likely the result of migratory birds (a problem that we have experienced almost every year here at ANL also). The punctures (and water in some net radiometers) did not result in obvious invalidation of the data. Whether there is water in the lower hemisphere or not, a small puncture of the upper hemisphere would not have a major effect on the net radiation measurement. Significant amounts of water in the lower hemisphere should have a major effect on nighttime measurements particularly, although that effect is hard to detect in the data. The date when the puncture occurred is approximate - but is certainly no more than two weeks before the hemisphere was replaced. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1994 | 0300 | 11/02/1994 | 1715 |
Subject: | SGP/EBBR/E8 - Water in Net Radiometer |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | At some time previous to October 21, 1994, the top hemisphere of the net radiometer of the EBBR at E8 was pierced, probably by migrating birds. A DQR was issued by me a couple of weeks ago indicating other extended facilities at which this has happened. Net radiometer data was correct until October 21 at sometime between 0230 and 0300 GMT, when rain began to enter the radiometer; the effect can be easily seen in the data. Collection of 3/4 inch of water in the bottom hemisphere was not the only consequence of the situation. Unlike other net radiometers that have had this experience, this one produced incorrect data during daylight hours as well as during nightime hours until the water was removed, the sensor dried, and the hemispheres were replaced on November 2, 1994. During the period listed above, the following data are incorrect and cannot be recalculated: 5 minute: q 15 minute: mv_q 30 minute: q, h, e Science Team members or others using the data should recognize that the flux data is not correct for this fairly lengthy time period. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/06/1994 | 1200 | 12/09/1994 | 1530 |
Subject: | SGP/EBBR/E8 - Frozen Wind Instruments |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On December 6, 1994 a frontal system deposited freezing rain in central Kansas that resulted in the EBBR wind instruments becoming frozen in place at two extended facilities (E4 and E8) for a few days. The direction and speed instruments were frozen for slightly differents lengths of time. The times and fields to be indicated as incorrect are shown below (all 5, 15, and 30 minute data measured during the periods, inclusive, is incorrect): Wind Speed: 12/06/94 1200 GMT - 12/09/94 1530 GMT 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws Wind Direction: 12/06/94 1730 GMT - 12/07/94 1830 GMT 5 minute: wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: wind_d, sigma_wd |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/31/1994 | 1330 | 01/01/1995 | 1745 |
Subject: | SGP/EBBR/E8 - Frozen Wind Speed Instrument |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On December 31, 1994 EBBR wind speed instrument froze in place from icing. Wind speeds increased sufficiently on January 1, 1995 to break the anemometer free. The wind direction instrument was not affected. The times and fields that are incorrect are shown below. 5 minute: wind_s, res_ws 12/31 1335 GMT - 1/1 1745 GMT 15 minute: wind_s 12/31 1345 GMT - 1/1 1745 GMT 30 minute: wind_s, res_ws 12/31 1400 GMT - 1/1 1800 GMT |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/1995 | 0000 | 08/28/1995 | 2359 |
Subject: | AEM Home 30 Values Low |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | DQR No: Platform: EBBR Subject: E8 EBBR AEM Home 30 Values Low Date Submitted: August 28, 1995 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other _____________________________ For questions or problems, please contact the ARM Experiment Center at 509-375-6898 or via email at dqr@arm.gov. Platform/Measurement: EBBR/AEM What level data: (raw,a0,a1,b1,c1 etc): a1 (5 and 30 minute) a0 (15 minute) What location was the data collected at: E8, Coldwater, KS Period of time in question Begin Date 08/18/95 to present Data should be labeled: ___ questionable ___ All data fields affected ___ incorrect ___ Only some data fields affected ___ wrong calibration _X_ others Discussion of Problem: At least by August 18, 1995, the home 30 value intermittently dropped below 15, to typically a small negative number; this causes a data quality message to be output in the site operations log as the stated acceptable range for home 30 is 15 to 34.999999. However, no data is incorrect as a result of this condition. No replacement AEM is available, as the changeout kit AEMs are somewhat different from the initial units. I have requested site operations to diagnose the cause of the problem on the next PM visit and hopefully correct the problem. I will submit another DQR or addendum when successful replacement of the AEM is accomplished. Other observations/measurements impacted by this problem: Suggested Corrections of the Problem: (e.g. change calibration factor and recompute, flag data with this comment, etc.) Flag data with this comment. Data Processing Notes Date ------------------------------------------------------------------- END |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/1995 | 0630 | 09/05/1995 | 1800 |
Subject: | SGP/EBBR/E8 - Home 30 Values Correct |
DataStreams: | sgp30ebbrE8.b1 |
Description: | Beginning 0630 GMT on August 18, 1995, the home 30 value intermittently dropped below 15, to typically a small negative number; this causes a data quality message to be output in the site operations log as the stated acceptable range for home 30 is 15 to 34.999999. However, no data is incorrect as a result of this condition. On September 15 at 1800 GMT the home 30 signal returned to normal and has remained normal since then. This occurred without the intervention of site operations personnel. |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/18/1995 | 1705 | 12/19/1995 | 1450 |
Subject: | SGP/EBBR/E22 - Wind Speed Cups Frozen |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | Freezing rain, snow, and below freezing conditions during 18-19 January 1995 caused the wind speed sensor cups to freeze in a stopped state at E22, Cordell, OK on 18 January 1995. When temperatures rose to above freezing on 19 January, the ice thawed and the cups turned freely again. Measurements affected by the stalled wind speed sensor include: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/10/1996 | 0230 | 02/10/1996 | 1300 |
02/10/1996 | 1400 | 02/10/1996 | 1600 |
02/11/1996 | 0100 | 02/11/1996 | 0900 |
02/11/1996 | 1030 | 02/11/1996 | 1400 |
02/11/1996 | 1730 | 02/12/1996 | 0100 |
02/12/1996 | 1200 | 02/12/1996 | 1230 |
02/13/1996 | 0130 | 02/13/1996 | 0230 |
02/13/1996 | 1600 | 02/13/1996 | 1700 |
02/14/1996 | 0130 | 02/14/1996 | 0230 |
02/23/1996 | 0330 | 02/23/1996 | 0530 |
02/23/1996 | 1200 | 02/24/1996 | 0200 |
02/28/1996 | 2330 | 02/29/1996 | 0000 |
03/03/1996 | 1500 | 03/03/1996 | 1830 |
03/04/1996 | 0100 | 03/04/1996 | 0230 |
03/04/1996 | 0400 | 03/04/1996 | 1530 |
03/05/1996 | 1200 | 03/05/1996 | 1430 |
03/05/1996 | 1630 | 03/05/1996 | 1830 |
03/06/1996 | 0200 | 03/07/1996 | 0430 |
03/07/1996 | 2300 | 03/07/1996 | 2330 |
03/08/1996 | 0330 | 03/08/1996 | 1730 |
03/10/1996 | 0400 | 03/10/1996 | 1130 |
03/12/1996 | 0400 | 03/12/1996 | 1130 |
03/12/1996 | 1430 | 03/12/1996 | 1600 |
03/13/1996 | 0830 | 03/13/1996 | 1130 |
03/16/1996 | 0400 | 03/16/1996 | 0830 |
03/16/1996 | 1030 | 03/16/1996 | 1200 |
03/16/1996 | 1500 | 03/16/1996 | 1700 |
03/17/1996 | 0600 | 03/17/1996 | 0630 |
03/17/1996 | 0730 | 03/17/1996 | 2030 |
03/18/1996 | 0300 | 03/18/1996 | 1730 |
03/20/1996 | 1700 | 03/20/1996 | 1830 |
03/21/1996 | 0200 | 03/21/1996 | 0230 |
03/22/1996 | 0330 | 03/22/1996 | 0400 |
03/22/1996 | 1600 | 03/22/1996 | 1700 |
03/23/1996 | 0400 | 03/23/1996 | 1930 |
03/27/1996 | 1800 | 03/27/1996 | 2000 |
01/31/1996 | 1430 | 01/31/1996 | 1500 |
02/12/1996 | 1600 | 02/12/1996 | 1830 |
03/03/1996 | 0130 | 03/03/1996 | 0200 |
03/07/1996 | 2030 | 03/07/1996 | 2130 |
03/19/1996 | 0130 | 03/20/1996 | 1430 |
03/24/1996 | 2200 | 03/25/1996 | 1500 |
02/13/1996 | 1400 | 02/13/1996 | 1430 |
03/15/1996 | 0400 | 03/15/1996 | 2200 |
Subject: | SGP/EBBR/E22 - Reprocess: Soil Heat Flow Sensor 4 Incorrect Data |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil heat flow sensor #4 periodically output incorrect values during the period indicated. The output often indicated large negative heat flow. The sensor was replaced on March 19, 1996, which corrected the problem. The data affected (and thus incorrect) includes: 15 minute: mv_hft4 30 minute: shf4, c_shf4, g4, ave_shf, e, h. The dates and times (GMT) of incorrect values are listed below. Only half hour data times are listed. 01/31/96 1500 (shf4 only) 02/10/96 0230-1300, 1400-1600 02/11/96 0100-0900, 1030-1400, 1730-2330 02/12/96 0000-0100, 1200-1230, 1600-1830 02/13/96 0130-0230, 1430, 1600-1700 02/14/96 0130-0230 02/23/96 0330-0530, 1200-2330 02/24/96 0000-0200 02/29/96 0000 03/03/96 0130-0200, 1500-1830 03/04/96 0100-0230, 0400-1530 03/05/96 1200-1430, 1630-1830 03/06/96 0200-2330 03/07/96 0000-0430, 2030-2130, 2330 03/08/96 0330-1730 03/10/96 0400-1130 03/12/96 0400-1130, 1430-1600 03/13/96 0830-1130 03/15/96 0400-2200 03/16/96 0400-0830, 1030-1200, 1500-1700 03/17/96 0630, 0730-2030 03/18/96 0300-1730 03/19/96 0130-2330 03/20/96 0000-1430, 1700-1830 03/21/96 0230 03/22/96 0400, 1600-1700 03/23/96 0400-1930 03/24/96 2200-2330 03/25/96 0000-1500 03/27/96 1800-2000 The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1, 2, 3, and 5 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/1996 | 1502 | 04/17/1996 | 1553 |
Subject: | SGP/EBBR/E8 - 3-month check |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | During EBBR 3-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed above; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/24/1996 | 1315 | 04/24/1996 | 1402 |
Subject: | SGP/EBBR/E22 - 3-month check |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during this period, at least, do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. The period of incorrect data for the EBBR is listed in the times indicated; data times affected do not include the beginning time and do include the ending time (data files are denoted by the end time). |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/30/1996 | 1430 | 07/30/1996 | 2100 |
07/31/1996 | 1400 | 07/31/1996 | 1430 |
Subject: | SGP/EBBR/E8 - Annual Calibration Changeout |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On 30 July 1996 the EBBR changeout for annual calibration occurred at E8, Coldwater, KS. SEBS unit 5 was removed and unit 4 was installed. Between 1430 and 2100 GMT no data or incorrect data was collected; most of the data is missing. Data for half hours 1400-1430 GMT on 31 July is incorrect; 3 month calibration checks were being performed. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/1996 | 0200 | 08/07/1996 | 1445 |
08/11/1996 | 1200 | 08/11/1996 | 1845 |
08/11/1996 | 2315 | 08/11/1996 | 2330 |
08/12/1996 | 0000 | 08/12/1996 | 0200 |
08/12/1996 | 1130 | 08/12/1996 | 1900 |
08/12/1996 | 2230 | 08/14/1996 | 2200 |
08/15/1996 | 1630 | 08/15/1996 | 1730 |
08/16/1996 | 2200 | 08/20/1996 | 1945 |
08/11/1996 | 1015 | 08/11/1996 | 1045 |
Subject: | SGP/EBBR/E8 - Incorrect Right Humidity Probe Temperature |
DataStreams: | sgp30ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The temperature portion of the right humidity probe produced incorrect values during the periods indicated. No site operations intervention occurred to correct the problem. The problem has not reoccurred. The following data values are incorrect for the periods listed below: 15 minute: rr_thum_r 30 minute: hum_top, hum_bot, vp_top, vp_bot, e, h, bowen August 5: 0200-2330 6: 0000-2330 7: 0000-1445 11: 1045, 1200-1845, 2315-2330 12: 0000-0200, 1130-1900, 2230-2330 13: 0000-2330 14: 0000-2200 15: 1630-1730 16: 2200-2330 17: 0000-2330 18: 0000-2330 19: 0000-2330 20: 0000-1945 |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/19/1996 | 0000 | 06/19/1996 | 0230 |
06/19/1996 | 0800 | 06/19/1996 | 1000 |
06/30/1996 | 0230 | 06/30/1996 | 1530 |
06/30/1996 | 2130 | 06/30/1996 | 2330 |
07/01/1996 | 0100 | 07/01/1996 | 0830 |
07/02/1996 | 0500 | 07/02/1996 | 1500 |
07/03/1996 | 0130 | 07/03/1996 | 0400 |
07/03/1996 | 1430 | 07/03/1996 | 1530 |
07/04/1996 | 0200 | 07/04/1996 | 0230 |
07/05/1996 | 0000 | 07/05/1996 | 0030 |
07/05/1996 | 0200 | 07/05/1996 | 0230 |
07/02/1996 | 0000 | 07/02/1996 | 0400 |
Subject: | SGP/EBBR/E8 - Reprocess: Soil Heat Flow Sensor 2 Incorrect Data |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil heat flow sensor #2 periodically outputted incorrect values during the period indicated. The output often indicated large negative heat flow. Problems with good connections of the sensor leads have caused this problem in the past. However, no maintenance was performed before the problem ceased on 15 July 1996. No data is available for 20-27 and 29 July, so it is not known if the problem occurred during that time. The data affected (and thus incorrect) includes: 15 minute: mv_hft2 30 minute: shf2, c_shf2, g2, ave_shf, e, h. The dates and times (GMT) of incorrect values are listed below. Only half hour times are listed. 06/19/96 0230, 0800-1000 06/30/96 0230-1530, 2130-2330 07/01/96 0100-0830 07/02/96 0400, 0500-1500 07/03/96 0130-0400, 1430-1530 07/04/96 0200-0230 07/05/96 0030, 0230 The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1 and 3 through 5 to calculate the average soil heat flow, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/1996 | 1300 | 11/06/1996 | 2300 |
Subject: | SGP/EBBR/E22 - Annual Calibration Changeout |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | On 6 November 1996 the EBBR changeout for annual calibration occurred at E22, Cordell, OK. SEBS unit 7 was removed and unit 3 was installed. Between 1300 and 2300 GMT no data or incorrect data was collected; most of the data is missing. Data for half hour 1930 GMT on 7 November is incorrect; 3 month calibration checks were being performed. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/30/1996 | 1726 | 10/30/1996 | 1801 |
Subject: | SGP/EBBR/E8 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp15ebbrE8.b1, sgp5ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/1996 | 1500 | 11/06/1996 | 1553 |
Subject: | SGP/EBBR/E22 - EBBR 3 Month Checks - 10 EFs |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/1996 | 1500 | 11/01/1996 | 1530 |
11/10/1996 | 0900 | 11/10/1996 | 0930 |
11/09/1996 | 0800 | 11/09/1996 | 0830 |
Subject: | SGP/EBBR/E8 - Incorrect Soil Moisture #1 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Moisture probe #1 went offscale at these half hour times, for an unknown reason. This problem may reoccur, and at present is extremely intermittent. A work request has been submitted to site operations to check the soil and electrical contacts of the probe. Affected data values are: 15 minute: r_sm1 30 minute: sm1, c_shf1, ces1, cs1, g1, ave_shf, e, h |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/1996 | 0300 | 08/03/1996 | 0830 |
08/03/1996 | 2030 | 08/03/1996 | 2100 |
08/05/1996 | 0230 | 08/07/1996 | 1430 |
09/20/1996 | 1030 | 09/20/1996 | 1800 |
09/20/1996 | 2100 | 09/20/1996 | 2130 |
09/20/1996 | 2230 | 09/20/1996 | 2330 |
09/21/1996 | 2330 | 09/22/1996 | 0130 |
09/22/1996 | 0530 | 09/23/1996 | 0130 |
09/23/1996 | 0400 | 09/26/1996 | 0700 |
09/26/1996 | 1900 | 09/26/1996 | 2030 |
09/27/1996 | 0100 | 09/29/1996 | 2330 |
09/30/1996 | 0000 | 09/30/1996 | 1930 |
10/07/1996 | 1400 | 10/07/1996 | 1830 |
10/07/1996 | 2000 | 10/15/1996 | 1900 |
08/03/1996 | 1400 | 08/03/1996 | 1700 |
09/21/1996 | 1500 | 09/21/1996 | 1530 |
09/30/1996 | 2130 | 10/07/1996 | 0530 |
Subject: | SGP/EBBR/E8 - Incorrect Soil Moisture #2, Soil Temperature #2 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil Moisture probe #2 and soil temperature probe #2 went offscale at times soon after installation of a new EBBR unit at E8. The problem may have been poor soil contact to the probes or poor electrical contact at the j- panel. No site operation intervention occurred to correct the problem; the problem has not reoccurred since 15 October 1996. Affected data values were: 15 minute: r_sm2, rr_ts2 30 minute: sm2, ts2, c_shf2, ces2, cs2, g2, ave_shf, e, h The affected times (GMT) and dates are listed below: Aug 3: 0300-0830, 1400-1700, 2030-2100 5: 0230-2330 6: 0000-2330 7: 0000-1430 Sep 20: 1030-1800, 2130, 2230-2330 21: 1530, 2330 22: 0000-0130, 0530-2330 23: 0000-0130, 0400-2330 24: 0000-2330 25: 0000-2330 26: 0000-0700, 1900-2030 27: 0100-2330 28: 0000-2330 29: 0000-2330 30: 0000-1930, 2130-2330 Oct 1: 0000-2330 2: 0000-2330 3: 0000-2330 4: 0000-2330 5: 0000-2330 6: 0000-2330 7: 0000-0530, 1400-1830, 2000-2330 8: 0000-2330 9: 0000-2330 10: 0000-2330 11: 0000-2330 12: 0000-2330 13: 0000-2330 14: 0000-2330 15: 0000-1900 |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/1996 | 0000 | 06/10/1997 | 2200 |
Subject: | SGP/EBBR/E8 - Incorrect Soil Temperature #s 1 and 2 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil temperature probes #1 and #2 went offscale at times for many months. Replacement probes were not available until recently. The probes were replaced on 16 June, 1997 at 2200 GMT. Affected data values were: 15 minute: r_sm1, r_sm2, rr_ts1, rr_ts2 30 minute: sm1, sm2, ts1, ts2, c_shf1, c_shf2, ces1, ces2, cs1, cs2, g1, g2, ave_shf, e, h |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/15/1997 | 1500 | 06/24/1997 | 2000 |
Subject: | SGP/EBBR/E8 - Incorrect Soil Temperature #s 1 and 2 |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil temperature probes #1 and #2 were replaced on June 10, 1997. However, the leads were apparently not connected tightly enough, resulting in some large or offscale resistance measurements and therefore temperature values for those probes. The problem occurred only occasionally until June 20, 1997. On June 20 - 24, 1997 probes 1 and 2 outputs were offscale for several hours per day, both not always at the same time. On June 24, 1997, site operations checked and tightened the connections. No offscale values have occurred since then. Affected data values were: 15 minute: r_sm1, r_sm2, rr_ts1, rr_ts2 30 minute: sm1, sm2, ts1, ts2, c_shf1, c_shf2, ces1, ces2, cs1, cs2, g1, g2, ave_shf, e, h |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/14/1997 | 0000 | 04/09/1997 | 1730 |
Subject: | SGP/EBBR/E22 - Reprocess: E22 Soil Heat Flow Sensor # 5 Incorrect Data |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil heat flow sensor #5 frequently produced large negative (and incorrect) values during the period above. The problem ended during a PM visit by site operations. However, there is no documentation by Site Operations in the instrumentation report or corrective maintenance report that Site Operations personnel took action to correct the problem. Site Operations also told me that they have no record of any changes or corrective maintenance. Apparently, something fortuitous happened during the PM to correct the problem. The start date and time of the problem cannot be determined exactly, but it did begin during a period of missing data from Feb. 14-25, 1997. Values affected (and thus incorrect) include: 15 minute: mv_hft5 30 minute: shf5, c_shf5, g5, ave_shf, e, h. The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1, 2, 3, and 4 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/19/1997 | 0130 | 07/08/1997 | 1915 |
Subject: | SGP/EBBR/E8 - AEM Inoperative |
DataStreams: | sgp30ebbrE8.b1 |
Description: | Beginning on 19 June 1997, the set screw holding one of the AEM sprockets to the drive motor became stripped, preventing the aspirator units from exchanging every time. The problem was intermittent until 24 June when site operations personnel determined that a new sprocket (therefore a replacement AEM) was needed. The AEM was therefore left in the position such that the right aspirator was in the bottom position continuously, with home signals indicating values near -1. The AEM was replaced on 8 July 1997. For most of the time period above, home signals, bowen, e, h, and all thermocouple and probe tmperatures, probe relative humidities, and vapor pressures do not reflect a switching AEM. However, temperatures, RH and vapor pressures are accurate for their height. Obviously, bowen, e, and h are incorrect. |
Measurements: | sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/13/1997 | 0305 | 07/16/1997 | 1500 |
Subject: | SGP/EBBR/E22 - Right Aspirator Unit Stolen |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | On 13 June 1997, vandals stole the right AEM aspirator unit, including the thermocouple and T/RH probe inside. These were replaced on 16 July 1997. Incorrect data fields during the period above are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, mv_hum_r, tair_r 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/1997 | 2030 | 07/14/1997 | 0915 |
Subject: | SGP/EBBR/E8 - Reprocess: E8 EBBR Program Compile Error, Effect On Net Radiation |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | An error in CR10X program compilation occurred when new software was used to perform the compilation. The new software created an erroneous data table listing, which the program works off of. An incorrect data location was used in the adjustment of net radiation for wind speed, causing all net radiation values to be divided by 2. The following data was incorrect during the period above: 5 minute: q 15 minute: mv_q 30 minute: q, e, h The latent and sensible heat fluxes can be corrected by recomputing as follows: e = -(2q + ave_shf)/(1 + bowen) h = -(e + ave_shf + 2q) |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/1997 | 0915 | 08/05/1997 | 2300 |
Subject: | E8 EBBR Lightning Strike |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | DQR No: Platform: EBBR Subject: E8 EBBR Lightning Strike Date Submitted: October 16, 1997 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other Platform/Measurement: EBBR/Numerous What level data: (raw,a0,a1,b1,c1 etc): a1 (15 and 30 minute) a0 (5 minute) What location was the data collected at: E8, Coldwater, KS Period of time in question Begin Date 07/14/97 Time 0915 GMT End Date 08/05/97 Time 2300 GMT Discussion of Problem: A lightning "strike" to the EF 8 Facility shortly after 0900 GMT on 14 July 1997 caused the failure of several EBBR sensors, the solar panel voltage regulator, and the CR10X datalogger. All data is missing for the period above, with degraded operations thereafter for a couple of months while replacement parts slowly became available (a separate DQR will cover the degraded data period). |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/30/1997 | 1700 | 07/30/1997 | 1830 |
Subject: | SGP/EBBR/E22 - Reprocess: E22 EBBR Soil Heat Flow Probe #5 Incorrect Data |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Soil heat flow probe #5 produced large negative (and incorrect) values during the period above. The probe was replaced. Somehow, two half hour records were produced at 1900 GMT as a result, the first record having 9s for the data fields; the second record is correct. Values affected (and thus incorrect) include: 15 minute: mv_hft5 30 minute: shf5, c_shf5, g5, ave_shf, e, h. The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1, 2, 3, and 4 to calculate the average soil heat flow, ave_shf = (g1 + g2 + g3 + g4)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/1997 | 2330 | 10/15/1997 | 1830 |
Subject: | SGP/EBBR/E8 - Reprocess: E8 EBBR Lightning Strike / sensor Replacements |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | DQR No: Platform: EBBR Subject: E8 EBBR Lightning Strike Date Submitted: October 17, 1997 Submitted By: _x_ Instrument Mentor ___ EST Member ___ Science Team Member ___ Other Platform/Measurement: EBBR/Numerous What level data: (raw,a0,a1,b1,c1 etc): a1 (15 and 30 minute) a0 (5 minute) What location was the data collected at: E8, Coldwater, KS Period of time in question Begin Date 08/05/97 Time 2330 GMT End Date 10/15/97 Time 1830 GMT Discussion of Problem: A lightning "strike" to the EF 8 Facility shortly after 0900 GMT on 14 July 1997 caused the failure of several EBBR sensors, the solar panel voltage regulator, and the CR10 datalogger. A previous DQR indicated this. During the period listed above, all but one of the damaged parts were replaced. Replacements occurred on 22 July, 5 August, 19 August, and 16 September. Some sensors began producing correct data immediately after their replacement; however the temperature measurements were incorrect until 15 October, when the reference temperature probe, with which they are corrected, was reconnected. Only the soil heat flow probe #2 remains to be replaced. Average soil, sensible, and latent heat fluxes are therefore now ,still, incorrect, but can be recomputed using only four sets of soil heat flow probes, as below. ave_shf = (g1 +g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/29/1997 | 2003 | 04/29/1997 | 2030 |
Subject: | SGP/EBBR/E8 - EBBR Biannual Checks |
DataStreams: | sgp15ebbrE8.b1, sgp5ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/1997 | 1525 | 04/23/1997 | 1605 |
Subject: | SGP/EBBR/E22 - EBBR Biannual Checks |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During EBBR 6 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for periods indicated. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/1997 | 0900 | 10/29/1997 | 2000 |
Subject: | SGP/EBBR/E8 - Reprocess: SHF2 Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil heat flow probe #2 outputted incorrect, large positive values during the period indicated. Replacement of the probe on 29 October 1997 corrected the problem. The serial number of the probe replaced is 923025 and of the replacement probe is 923110. The data affected (and thus incorrect) includes: 15 minute: mv_hft2 30 minute: shf2, c_shf2, g2, ave_shf, e, h. The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1, and 3 through 5 to calculate the average soil heat flow, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/1997 | 0000 | 04/21/2000 | 1640 |
Subject: | SGP/EBBR/E22 - Reprocess: E22 Soil Moisture Values Off End of Calibration at High Soil |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Intermittently, since November 1997 soil moisture probes 1, 2, and 3 indicate small positive or else negative values when soil moisture is high (greater than 35%). This situation is similar to that at the Cement EBBR, although the soil at Cordell is not capable of holding as much water as that at Cement. The cause of the problem is inadequate range in the calibration curve equation. Very high soil moisture cannot be obtained with the calibration soil used by REBS. The soil moisture equations could be manipulated to allow larger soil moistures to be calculated. However, the infrequency of occurence does not justify doing so. The soil moisture data must be inspected by the data user to see when soil moisture values for one or more probes is inconsistent with the others. Recalculation of ave_shf, e, and h must be made using only the soil moisture sets with correct soil moisture values. Affected data values are (# for probe number): 15 minute: rr_sm# 30 minute: sm#, c_shf#, cs#, ces#, g#, ave_shf, e, h Sensible, latent, and soil heat fluxes can be recalculated using the soil probe sets that provide correct data. For example, if soil moisture 2 is incorrect, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), If soil moistures 1 and 2 are incorrect, ave_shf = (g3 + g4 + g5)/3 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), If soil moistures 1, 2, and 3 are incorrect, ave_shf = (g4 + g5)/2 e = -(q + ave_shf)/(1 +bowen) h = -(e + ave_shf +q), etc. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/21/1997 | 0830 | 12/22/1997 | 1700 |
01/09/1998 | 0000 | 01/09/1998 | 1300 |
Subject: | SGP/EBBR/E8- EBBR Frozen Wind Speed |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Wind speed cups were frozen during the indicated time periods. All 5, 15, and 30 minute data values indicated are incorrect for the periods listed. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/04/1997 | 0000 | 01/06/1998 | 1840 |
Subject: | SGP/EBBR/E8 - Battery Low - Data Affected |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | The battery condition deteriorated over time so that the solar panel and AC charger were not always able to keep the battery voltage high enough to produce correct data after 3 December 1997, especially at night. The condition became particularly bad after 8 December 1997. Data for conditions of battery voltage below 10.4 volts should be considered incorrect; inspection of the battery voltage data is required to determine this. The battery was replaced on 6 January 1998 at 1840 GMT, after which data was again correct. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/1997 | 1900 | 10/29/1997 | 2010 |
Subject: | SGP/EBBR/E8 - Reprocess: E8 Soil Heat Flow #2 Failure - Replaced |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1 |
Description: | Soil heat flow probe #2 values were positive and in the thousands after an apparently catastrophic failure at 1900 GMT on 11 October 1997. The probe was replaced on 29 October 1997 at 2010 GMT. The data affected (and thus incorrect) includes: 15 minute: mv_hft2 30 minute: shf2, c_shf2, g2, ave_shf, e, h. The latent and sensible heat fluxes can be recalculated by using only soil heat flows 1 and 3 through 5 to calculate the average soil heat flow, ave_shf = (g1 + g3 + g4 + g5)/4 e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q). |
Measurements: | sgp15ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/1998 | 0745 | 01/14/1998 | 1600 |
Subject: | SPG/EBBR/E8 - Wind Speed Sensor and AEM Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The Automatic Exchange Mechanism was frozen during the entire period and did not exchange. Data values that were affected by the inoperative AEM are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot 15 minute: rr_thum_r, rr_thum_l, mv_hum_r, mv_hum_l, tair_r, tair_l 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, e, h, bowen The wind speed sensor was frozen stalled during part of this period (1100-1600 GMT). Data values that were affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/15/1998 | 0300 | 01/15/1998 | 1415 |
Subject: | SGP/EBBR/E22 - Low Battery - Resulting In Incorrect Data |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | A low battery condition (less than 10.5 volts) caused a number of data fields to be incorrect at some times and all data fields to be incorrect at a more limited number of times. All data fields are suspect when the battery voltage drops below 10.5 volts because the CR10 cannot operate properly. Some sensors are affected slightly (i.e. pressure) at slightly larger voltages. The low battery condition resulted from an initially weak battery and a week of extremely low solar radiation (from heavy cloudiness) on the solar panel that provides most of the charging of the battery. Comments added by Marv Wesely: David Cook has been working with Dave Breedlove to try to assess the age of the batteries. The dates on the batteries needs to be read during site visits and provided to David C. It's quite possible that they need to be replaced. It's work in progress, though slow at times. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/08/1997 | 1830 | 10/08/1997 | 1902 |
Subject: | SGP/EBBR/E22 - 3-Month Check |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1 |
Description: | During EBBR 3 month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/14/1998 | 1820 | 03/15/1998 | 1625 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During the period above the wind speed sensor cups were frozen stalled by freezing rain. The data fields affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/16/1998 | 1530 | 03/16/1998 | 1830 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During the period above the wind speed sensor cups were frozen stalled by freezing rain. The data fields affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/1998 | 1510 | 03/17/1998 | 1555 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During the period above the wind speed sensor cups were frozen stalled by freezing rain. The data fields affected are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/29/1997 | 1930 | 10/29/1997 | 2125 |
Subject: | SGP/EBBR/E8 - 3-Month Check |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During EBBR 3-month checks the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/1997 | 2130 | 05/12/1998 | 1850 |
Subject: | EBBR E8 Net Radiation Data Quality |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | On 14 July 1997 lightning struck the EBBR at E8. A number of sensors and components needed to be replaced. One component that was not replaced, a J-panel, suffered some damage that was not obvious at the time. Also, changes in sensors eventually required a change in the CR10 program to accomodate new calibration factors. A new program was installed on 14 October. Unfortunately, an incorrect nighttime calibration for the net radiation was contained in the new program provided by the EBBR vendor REBS. The calibration problem will be covered in a separate DQR. The J-panel problem began to surface on 14 October as well, with most net radiation values being smaller than correct values during the day, and large negative at night; the latter was partially from the J-panel problem and partially from the incorrect nighttime calibration. All net radiation values, nighttime and daytime from 14 October 1997 through 12 May 1998 need to be examined carefully (and compared with other nearby sites) to determine if they are correct, and nighttime values need to be corrected. Much of the latent and sensible heat flux data is corrupted by both problems. The correction to nightime values will be covered in two subsequent DQRs, as two net radiometers, with different calibrations, have been used at E8 since 14 October 1997. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/1998 | 1800 | 05/06/1998 | 1840 |
Subject: | SGP/EBBR/E22 - Six Month Checks |
DataStreams: | sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period indicated. |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/28/1998 | 1958 | 04/28/1998 | 2031 |
Subject: | SGP/EBBR/E8 - Six Month Checks |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | During EBBR 6 month checks (formally called 3 month checks) the Automatic Exchange Mechanism is disabled (reflected by home signals near zero) so that the two aspirator units can be placed at the same height. The measurements during these periods do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h), or latent heat flux (e), nor do the air temperature and relative humidity measurements indicate correct gradients. Checks performed just before or after the AEM is disabled also lead to some incorrect data. Incorrect 5, 15 minute and half hour data occurs for the period above. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/1997 | 2200 | 04/20/1998 | 1720 |
Subject: | EBBR E8 Adjustment of Negative Net Radiation Values |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Please see DQR D980526.2 for some background information on the cause of the problem described here. On 14 October 1997 a new CR10 program was installed at E8 EBBR. The program contained the wrong calibration coefficient for negative net radiations. A multiplier had been left out of the computation of the coefficient by the vendor. All negative net radiation values during the stated period need to be DIVIDED by 13.88. Only the 5 minute and 30 minute q values are affected. The 15 minute mv_q values are correct. Other values affected are 30 minute h and e, which would need to be recalculated from: e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/20/1998 | 1725 | 05/26/1998 | 1920 |
Subject: | EBBR E8 Adjustment of Negative Net Radiation Values |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Please see DQR D980526.2 for some background information on the cause of the problem described here, and D980603.1 for correction to the data obtained with the previous net radiometer. On 14 October 1997 a new CR10 program was installed at E8 EBBR. The program contained the wrong calibration coefficient for negative net radiations. A multiplier had been left out of the computation of the coefficient by the vendor. On 20 April 1998 a new net radiometer was installed. It was not noticed that the mutiplier was missing. A slightly different calibration was entered for the net radiometer, again without the multiplier applied. A few weeks later the mentor, in trying to diagnose the obvious incorrect negative net radiation values, discovered the missing multiplier. A new program with the multiplier included was installed on 26 May 1998 at 1917 GMT. All negative net radiation values during the stated period above need to be DIVIDED by 13.775. Only the 5 minute and 30 minute q values are affected. The 15 minute mv_q values are correct. Other values affected are 30 minute h and e, which would need to be recalculated from: e = -(q + ave_shf)/(1 + bowen) h = -(e + ave_shf + q) |
Measurements: | sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/24/1998 | 0300 | 05/26/1998 | 1700 |
Subject: | SGP/EBBR/E8 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor showed the usual failure signature much of the time, an output of 135 degrees, beginning on 24 May 1998. Not all times showed 135 degrees, but all wind direction data during the stated period should be considered incorrect. The values affected are: 5 minute: res_ws, wind_d, sigma_d 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_d |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/20/1998 | 1505 | 12/22/1998 | 1800 |
Subject: | SGP/EBBR/E8 - Wind Direction and Speed Data Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction and speed sensors were iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws, wind_d, sigma_wd 15 minute: wind_s, mv_wind_d 30 minute: wind_s, res_ws, wind_d, sigma_wd |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/03/1998 | 0615 | 11/10/1998 | 1430 |
Subject: | SGP/EBBR/E22 - Battery Voltage Low, Data Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
Description: | Battery voltages less than 10.4 volts occurred much of the time. All data should be considered incorrect when the voltage is less than 10.4 volts. |
Measurements: | sgp5ebbrE22.00:
sgp30ebbrE22.b1:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/1999 | 2355 | 01/08/1999 | 1100 |
Subject: | SGP/EBBR/E22 - Wind Direction Data Incorrect |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp5ebbrE22.00 |
Description: | The wind direction sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp30ebbrE22.b1:
sgp5ebbrE22.00:
sgp5ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/08/1999 | 0835 | 01/08/1999 | 1500 |
Subject: | SGP/EBBR/E8 - Wind Speed Data Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind speed sensor was iced stopped by freezing rain, resulting in the following values being incorrect: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws, |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/1999 | 1605 | 01/29/1999 | 0830 |
Subject: | SGP/EBBR - EBBR E8 Wind Speed Iced, Data Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp5ebbrE8.00 |
Description: | The wind speed sensor was stopped by icing. Wind speed data is incorrect. The values are: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/1999 | 1530 | 02/02/1999 | 1800 |
Subject: | SGP/EBBR/E8 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction sensor output read 135 degrees for much of the period, an indication of sensor failure. Energy Balance fluxes are not affected by the problem. The values below were incorrect when wind direction was 135: 5 minute: res_ws, wind_d, sigma_wd 15 minute: mv_wind_d 30 minute: res_ws, wind_d, sigma_wd |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/30/1999 | 0220 | 01/31/1999 | 2030 |
Subject: | SGP/EBBR/E8 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp5ebbrE8.00 |
Description: | The wind speed sensor was iced stopped for the period. Energy Balance fluxes are not affected by the problem. The values below were incorrect for the period: 5 minute: wind_s, res_ws 15 minute: wind_s 30 minute: wind_s, res_ws |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
sgp5ebbrE8.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/1999 | 2045 | 03/10/1999 | 2100 |
Subject: | SGP/EBBR/E22 - Right Aspirator Fan Replaced |
DataStreams: | sgp15ebbrE22.b1, sgp30ebbrE22.b1 |
Description: | Replacement of the right aspirator fan on the EBBR AEM caused temperature, relative humidity, and vapor pressure data to be incorrect for the 2045 and 2100 GMT 15 minute data and Bowen ratio, sensible heat flux, and latent heat flux to be incorrect for the 2100 GMT data. |
Measurements: | sgp30ebbrE22.b1:
sgp15ebbrE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/1999 | 0955 | 03/14/1999 | 1800 |
Subject: | SGP/EBBR/E8 - Wind Speed Frozen |
DataStreams: | sgp30ebbrE8.b1, sgp5ebbrE8.b1, sgp15ebbrE8.b1 |
Description: | Wind speed and residual wind speed during this period are incorrect caused by an ice storm. |
Measurements: | sgp15ebbrE8.b1:
sgp5ebbrE8.b1:
sgp30ebbrE8.b1:
|