Data Quality Reports for Session: 143622 User: riccardo Completed: 07/12/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D051028.5TWP/MPL/C3 - Reprocess: Global metadata errortwpmplC3.a1
D060224.3TWP/MPL/C3 - Temporary laser failuretwpmplC3.00, twpmplC3.a1
D060227.1TWP/MPL/C3 - Backscatter profiles are inaccurate due to computer HD corruption.twpmplC3.00, twpmplC3.a1
D060621.1TWP/MPL/C3 - Missing data due to instrument removaltwpmplC3.00, twpmplC3.a1
D080529.2TWP/MPL/C3 - Intermittent Double Pulse from lasertwpmplC3.a1


DQRID : D051028.5
Start DateStart TimeEnd DateEnd Time
08/08/2002004107/22/20031447
Subject:
TWP/MPL/C3 - Reprocess: Global metadata error
DataStreams:twpmplC3.a1
Description:
During the time period specified, the global metadata attribute "averaging_int" is 
reported as 30 seconds when the value for this attribute should be 60 seconds.
Measurements:twpmplC3.a1:
  • time_offset


Back To Table of Contents

DQRID : D060224.3
Start DateStart TimeEnd DateEnd Time
02/20/2006060002/23/20060600
Subject:
TWP/MPL/C3 - Temporary laser failure
DataStreams:twpmplC3.00, twpmplC3.a1
Description:
No backscatter information is in the data. The backscatter data in the files during this 
time is solar background information only. 

The laser turned it's self off, reason: unknown. A power cycle of the entire system 
corrected the problem.
Measurements:twpmplC3.00:
  • Raw data stream - documentation not supported

twpmplC3.a1:
  • Preliminary cloud base height above ground level(preliminary_cbh)
  • detector_counts


Back To Table of Contents

DQRID : D060227.1
Start DateStart TimeEnd DateEnd Time
02/24/2006060002/27/20060700
Subject:
TWP/MPL/C3 - Backscatter profiles are inaccurate due to computer HD corruption.
DataStreams:twpmplC3.00, twpmplC3.a1
Description:
The data shows a consistent cloud deck at 7.5km with what appears to be a duplication of 
the data from ground zero. The temperature and energy monitor values are also affected. 
The computer was found with multiple corruption problems. These corruptions may have cause 
the ghost data at 7.5km.  A chkdsk was performed which appears to have corrected the 
problem.
Measurements:twpmplC3.00:
  • Raw data stream - documentation not supported

twpmplC3.a1:
  • instrument_temp
  • background_signal
  • energy_monitor
  • laser_temp
  • detector_temp
  • Preliminary cloud base height above ground level(preliminary_cbh)
  • detector_counts


Back To Table of Contents

DQRID : D060621.1
Start DateStart TimeEnd DateEnd Time
07/23/2003000003/03/20050000
Subject:
TWP/MPL/C3 - Missing data due to instrument removal
DataStreams:twpmplC3.00, twpmplC3.a1
Description:
The instrument was removed from this site to replace a failed instrument at TWP-C2 Nauru.
Measurements:twpmplC3.00:
  • Raw data stream - documentation not supported

twpmplC3.a1:
  • range
  • voltage_10
  • Pressure, associated geometric height, at altitude, NGM model output(height)
  • voltage_15
  • range_bin_width
  • time_offset
  • detector_counts
  • lat
  • assumed_zero_range_bin
  • base_time
  • shots_summed
  • laser_temp
  • range_bin_time
  • time
  • detector_temp
  • range_offset
  • instrument_temp
  • background_signal
  • alt
  • energy_monitor
  • lon
  • pulse_rep
  • max_altitude
  • Preliminary cloud base height above ground level(preliminary_cbh)
  • filter_temp
  • range_bins
  • voltage_05


Back To Table of Contents

DQRID : D080529.2
Start DateStart TimeEnd DateEnd Time
02/01/2005000008/24/20062359
Subject:
TWP/MPL/C3 - Intermittent Double Pulse from laser
DataStreams:twpmplC3.a1
Description:
When Spectra Physics discontinued support for their laser diode power supply, an 
alternative was need to extend the usefulness of the Spectra Physics laser head inside the MPLs.  
The solution was to use a Coherent laser diode. Unfortunately, the Coherent diode lasers 
do not have the exact same characteristics as the Spectra Physics diode lasers. As a 
result, the output of the Spectra Physics laser head sometimes produces a second pulse close 
in time but lower in intensity from the first main pulse. This is known as ?double pulsing?.

Not all of the Spectra Physics/Coherent combination systems produce a double pulse and the 
systems that do produce a double pulse do not always exhibit this behavior. It can be 
identified by a second peak at ~.2km. At first glance the signal in the data will resemble 
a low aerosol or cloud layer. When the double pulse is weak, the atmospheric signal will 
obscure the double pulse signal. At the ARM sites with a sunshade such as TWP and AMF, 
the double pulse can be seen when the sunshade closes during solar noon. When the sunshade 
closes, the MPL does not see atmospheric signal so the second pulse is evident. At SGP, a 
daily window cleaning is part of the routine. During this time the double pulse can be 
seen. Without a sunshade or window cleaning it is impossible to say definitively if double 
pulse is present. 

The first system to produce a double pulse was MPL s/n 008 located at TWP ? C2 Nauru in 
February 2005. The second system was MPL s/n 004 located at SGP in August 2005. The 
potential for the other systems to produce a double pulse is high. However, as of March 2006, 
the MPLs at NSA, TWP ? C1, TWP ? C3 and AMF have not recorded data with the double pulse 
artifact.  

The ARM program is in the process of replacing all the MPLs with a newer model. After the 
new MPLs are deployed the double pulse problem should not be an issue. The old systems 
will be relegated as spares. The double pulse problem will return if an old system is 
needed to replace a failed new system.
Measurements:twpmplC3.a1:
  • detector_counts


Back To Table of Contents



END OF DATA