Subject: | PVC/GNDRAD/M1 - Incorrect calibration coefficients, Incorrect Configuration |
DataStreams: | pvcgndrad20sM1.a0, pvcgndrad60sM1.b1, pvcgndradM1.00
|
Description: | Incorrect programs were used in both Skyrad and Gndrad since PVC deployment period began.
The Gndrad program was corrected 10/04/2012 approx 20:17GMT. Reprocesing will be
required of the upwelling PSP (up_short_hemsip). This will require applying the following
corrections for the time range specified.
Serial number:
serial number(original) = 30938
serial number(corrected) = 37297
Data:
up_short_hemip(corrected) = up_short_hemisp(original)*(113.25(new cal)/129.42(old cal)) |
Measurements: | pvcgndradM1.00: - Raw data stream - documentation not supported
pvcgndrad60sM1.b1: - Upwelling Shortwave Hemispheric Irradiance, Unventilated Pyranometer, Minima(up_short_hemisp_min)
- Upwelling Shortwave Hemispheric Irradiance, Unventilated Pyranometer, Maxima(up_short_hemisp_max)
- Radiation, shortwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_short_hemisp)
- up_short_hemisp_std
pvcgndrad20sM1.a0:
|