DQR ID | Subject | Data Streams Affected |
---|---|---|
D050105.1 | TWP/SONDE/C1 - Bad RH at beginning of sounding | twpsondewnpnC1.b1 |
D050201.2 | TWP/SONDE/C1 - Bad RH at beginning of sounding | twpsondewnpnC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/30/2004 | 2330 | 12/31/2004 | 0108 |
Subject: | TWP/SONDE/C1 - Bad RH at beginning of sounding |
DataStreams: | twpsondewnpnC1.b1 |
Description: | It appears that the RH sensor heating circuit failed during this portion of the sounding (approximately between 0km and 11km). The RH oscillates rapidly between low (~5 %RH) and not so low (~70 %RH) and the oscillation seems to stop when the temperature is below ~40 degC. According to Vaisala these are symptoms of failure in the ASIC (application specific integrated circuit) that controls the heating. |
Measurements: | twpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/07/2005 | 2330 | 01/08/2005 | 0102 |
Subject: | TWP/SONDE/C1 - Bad RH at beginning of sounding |
DataStreams: | twpsondewnpnC1.b1 |
Description: | It appears that the RH sensor heating circuit failed during this portion of the sounding (approximately between 0km and 12km). The RH oscillates rapidly between low (~2 %RH) and not so low (~80 %RH) and the oscillation seems to stop when the temperature is below ~40 degC. According to Vaisala these are symptoms of failure in the ASIC (application specific integrated circuit) that controls the heating. |
Measurements: | twpsondewnpnC1.b1:
|