DQR ID | Subject | Data Streams Affected |
---|---|---|
D121003.1 | SGP/SIRS/E13 - Instrument problem | sgpsirs20sE13.a0, sgpsirs60sE13.a1, sgpsirsE13.00, sgpsirsE13.01, sgpsirsE13.a0, sgpsirsE13.a1, sgpsirsE13.a2, sgpsirsE13.b1 |
D121003.2 | SGP/SIRS/E13 - Reprocess: Incorrect UIR calibration | sgpsirs20sE13.a0, sgpsirsE13.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2012 | 0000 | 09/20/2012 | 2045 |
Subject: | SGP/SIRS/E13 - Instrument problem |
DataStreams: | sgpsirs20sE13.a0, sgpsirs60sE13.a1, sgpsirsE13.00, sgpsirsE13.01, sgpsirsE13.a0, sgpsirsE13.a1, sgpsirsE13.a2, sgpsirsE13.b1 |
Description: | up_long_hemisp measurements are intermittant and remain so after extensive troubleshooting. This problem will likly remain open ended until site ops can replace the cable which serves the up_long_hemisp instrument (i.e. the Eppley PIR). |
Measurements: | sgpsirs20sE13.a0:
sgpsirsE13.01:
sgpsirsE13.00:
sgpsirs60sE13.a1:
sgpsirsE13.a0:
sgpsirsE13.b1:
sgpsirsE13.a2:
sgpsirsE13.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/2012 | 0000 | 09/20/2012 | 2045 |
Subject: | SGP/SIRS/E13 - Reprocess: Incorrect UIR calibration |
DataStreams: | sgpsirs20sE13.a0, sgpsirsE13.b1 |
Description: | For the time range specified the UIR had incorrect calibration coefficients applied. The k1 coefficient applied was 0.25042 when it should have been 0.2558. This resulted in the instrument under reporting up_long_hemsip by approximately 2.1%. Data will be reprocessed. |
Measurements: | sgpsirs20sE13.a0:
sgpsirsE13.b1:
|