Data Quality Reports for Session: 147814 User: nickd Completed: 02/13/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D021211.1TWP/SMET/C3 - Lower wind direction sensor misaligned twpsmet60sC3.a1, twpsmet60sC3.b1
D030212.1TWP/SMET/C3 - Intermittent Pressure Sensor Failuretwpsmet60sC3.a1, twpsmet60sC3.b1
D031103.2TWP/SMET/C3 - Data reported at 15sec rather than 60s intervalstwpsmet60sC3.a1, twpsmet60sC3.b1
D040722.4TWP/SMET/C3 - Rain blocked pressure porttwpsmet60sC3.b1
D040726.3TWP/SMET/C3 - Negative rainfall ratestwpsmet60sC3.b1
D050210.1TWP/SMET/C3 - Data non-representativetwpsmet60sC3.b1
D050418.2TWP/SMET/C3 - Lower wind direction sensor failuretwpsmet60sC3.b1
D050808.3TWP/SMET/C3 - Data not representativetwpsmet60sC3.b1
D051115.1TWP/SMET/C3 - Upper wind sensor failuretwpsmet60sC3.b1
D051117.2TWP/SMET/C3 - RH Values Artificially Hightwpsmet60sC3.b1
D051117.3TWP/SMET/C3 - Instrument failuretwpsmet60sC3.b1
D060426.8TWP/SMET/C3 - ORG data noisy due to degraded groundtwpsmet60sC3.b1
D060502.3TWP/SMET/C3 - Reprocessed: Lower windspeed sensor calibration errortwpsmet60sC3.b1
D060627.1TWP/SMET/C3 - Data not representativetwpsmet60sC3.b1
D061011.4TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge addedtwpsmet60sC3.b1
D061221.2TWP/RAIN/C3 - faulty precipitation result from blocked funneltwprainC3.b1
D070306.2TWP/SMET/C3 - Barometer Data Incorrecttwpsmet60sC3.b1
D070306.3TWP/SMET/C3 - Water in enclosuretwpsmet60sC3.b1
D070803.1TWP/SMET/C3 - Upper wind speed artificially slowtwpsmet60sC3.b1
D071105.1TWP/SMET/C3 - ORG overestimated rainfalltwpsmet60sC3.b1
D080225.1TWP/SMET/C3 - Upper Wind Sensor Data Questionabletwpsmet60sC3.b1
D080428.3TWP/SMET/C3 - Reprocess: ORG Data has very small rainrate valuestwpsmet60sC3.b1
D080611.2TWP/SMET/C3 - Lower Wind Direction Intermittently Wrongtwpsmet60sC3.b1
D080715.1TWP/SMET/C3 - Upper wind speed data incorrecttwpsmet60sC3.b1
D080922.5TWP/SMET/C3 - Missing datatwpsmet60sC3.b1
D081022.8TWP/SMET/C3 - Wind Speed data contain offset due to induced voltagetwpsmet60sC3.b1
D090129.4TWP/SMET/C3 - Reprocess: Station Pressure values reported in hPa instead of kPa.twpsmet60sC3.b1
D090225.3TWP/SMET/C3 - Reprocess: Temperature data biased lowtwpsmet60sC3.b1
D090515.1TWP/SMET/C3 - Reprocess: Temperature Biased low by 1.06 Degreestwpsmet60sC3.b1
D091023.1TWP/SMET/C3 - Tower Lowered for Maintenancetwpsmet60sC3.b1
D120614.1TWP/MET/C3 - Incorrect ORG datatwpmetC3.b1
D130201.1TWP/DISDROMETER/C3 - Precip not measuredtwpdisdrometerC3.b1


DQRID : D021211.1
Start DateStart TimeEnd DateEnd Time
08/21/2002000012/09/20020000
Subject:
TWP/SMET/C3 - Lower wind direction sensor misaligned
DataStreams:twpsmet60sC3.a1, twpsmet60sC3.b1
Description:
The lower wind direction sensor (sensor #2) was misaligned and off by 15 degrees.
Measurements:twpsmet60sC3.a1:
  • wind2_dir_sd
  • Wind #2 direction vector average(wind2_dir_vec_avg)

twpsmet60sC3.b1:
  • lo_dir_sd
  • lo_dir_vec_avg


Back To Table of Contents

DQRID : D030212.1
Start DateStart TimeEnd DateEnd Time
10/23/2002000002/06/20032203
Subject:
TWP/SMET/C3 - Intermittent Pressure Sensor Failure
DataStreams:twpsmet60sC3.a1, twpsmet60sC3.b1
Description:
The Barometric Pressure sensor was intermittently failing at TWP site C3 Darwin.  At times 
there were large pressure fluctuations occurring on the order of 24 mb in a few minutes. 
 Times of large pressure fluctuations are readily apparent in the data.  At these times 
the data should be carefully analyzed before determining it's usefulness.  In between the 
periods of large fluctuations the readings appear to be stable.  It is still suggested 
that independent pressure data be obtained for comparison before use.
Measurements:twpsmet60sC3.a1:
  • Barometric Pressure(atmos_pressure)

twpsmet60sC3.b1:
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D031103.2
Start DateStart TimeEnd DateEnd Time
03/12/2002062806/03/20022359
Subject:
TWP/SMET/C3 - Data reported at 15sec rather than 60s intervals
DataStreams:twpsmet60sC3.a1, twpsmet60sC3.b1
Description:
Data in these data streams should be reported at 60sec intervals, but
the logger program running during this time had the interval set to
15s.  Users will have to average the 15sec data to get 60sec data.
Measurements:twpsmet60sC3.a1:
  • Wind #1 speed arithmetic average(wind1_spd_arith_avg)
  • Wind #2 speed maximum(wind2_spd_max)
  • Wind #2 speed maximum gust(wind2_spd_max_gust)
  • wind2_spd_sd
  • Wind #1 speed maximum gust(wind1_spd_max_gust)
  • Precipitation mean(precip_mean)
  • time_offset
  • vappress_sd
  • Wind #1 speed minimum(wind1_spd_min)
  • wind1_dir_sd
  • Temperature maximum(temp_max)
  • logger_temp
  • wind2_dir_sd
  • Wind #2 speed vector average(wind2_spd_vec_avg)
  • internal_voltage
  • bit_data
  • Wind #2 speed minimum(wind2_spd_min)
  • Precipitation maximum(precip_max)
  • Wind #1 speed vector average(wind1_spd_vec_avg)
  • Barometric Pressure(atmos_pressure)
  • Relative humidity maximum(relh_max)
  • Vapor pressure mean(vappress_mean)
  • Relative humidity minimum(relh_min)
  • temp_sd
  • Temperature minimum(temp_min)
  • relh_sd
  • Wind #1 direction vector average(wind1_dir_vec_avg)
  • Mean Air Temperature or Hardware Error(temp_mean)
  • external_voltage
  • Precipitation minimum(precip_min)
  • Wind #2 direction vector average(wind2_dir_vec_avg)
  • Wind #2 speed arithmetic average(wind2_spd_arith_avg)
  • wind1_spd_sd
  • precip_sd
  • Relative humidity mean(relh_mean)
  • Wind #1 speed maximum(wind1_spd_max)

twpsmet60sC3.b1:
  • lo_dir_sd
  • Vapor pressure mean(vappress_mean)
  • logger_temp
  • up_spd_max
  • up_spd_min
  • precip_sd
  • lo_spd_min
  • temp_sd
  • relh_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • up_spd_sd
  • lo_dir_vec_avg
  • lo_spd_arith_avg
  • tbrg_total
  • Mean Air Temperature or Hardware Error(temp_mean)
  • Precipitation mean(precip_mean)
  • up_dir_vec_avg
  • tbrg_count_total
  • time_offset
  • Relative humidity mean(relh_mean)
  • lo_spd_sd
  • lo_spd_vec_avg
  • Barometric Pressure(atmos_pressure)
  • up_spd_arith_avg
  • lo_spd_max
  • vappress_sd
  • up_spd_vec_avg
  • up_dir_sd
  • time
  • internal_voltage


Back To Table of Contents

DQRID : D040722.4
Start DateStart TimeEnd DateEnd Time
01/01/2004000005/17/20042259
Subject:
TWP/SMET/C3 - Rain blocked pressure port
DataStreams:twpsmet60sC3.b1
Description:
Atmospheric pressure values were suspect due to anomolous readings during heavy rain 
events.  Suspect blockage of port due to rainfall.  New logger and enclosure corrected problem.
Measurements:twpsmet60sC3.b1:
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D040726.3
Start DateStart TimeEnd DateEnd Time
03/12/2002062805/17/20042300
Subject:
TWP/SMET/C3 - Negative rainfall rates
DataStreams:twpsmet60sC3.b1
Description:
Due to limitations of the Zeno dataloggers all voltage measurements
from the optical rain gauge (ORG) were converted to rainfall rates. 
This resulted in negative values for rain rates.  Any value below
0.10 mm/hr should be considered to be 0 mm/hr.  New dataloggers 
alleviated this problem.
Measurements:twpsmet60sC3.b1:
  • precip_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • Precipitation mean(precip_mean)


Back To Table of Contents

DQRID : D050210.1
Start DateStart TimeEnd DateEnd Time
02/09/2005231002/09/20052320
Subject:
TWP/SMET/C3 - Data non-representative
DataStreams:twpsmet60sC3.b1
Description:
Tower mast was lowered for inspection.  Data during this time is non-representative.
Measurements:twpsmet60sC3.b1:
  • lo_dir_sd
  • tbrg_count_total
  • Vapor pressure mean(vappress_mean)
  • Relative humidity mean(relh_mean)
  • up_spd_max
  • up_spd_min
  • lo_spd_sd
  • lo_spd_vec_avg
  • precip_sd
  • lo_spd_min
  • Barometric Pressure(atmos_pressure)
  • temp_sd
  • up_spd_arith_avg
  • lo_spd_max
  • vappress_sd
  • up_spd_vec_avg
  • relh_sd
  • up_dir_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • up_spd_sd
  • lo_dir_vec_avg
  • lo_spd_arith_avg
  • tbrg_total
  • Mean Air Temperature or Hardware Error(temp_mean)
  • Precipitation mean(precip_mean)
  • up_dir_vec_avg


Back To Table of Contents

DQRID : D050418.2
Start DateStart TimeEnd DateEnd Time
04/14/2005023004/19/20050050
Subject:
TWP/SMET/C3 - Lower wind direction sensor failure
DataStreams:twpsmet60sC3.b1
Description:
The lower wind direction sensor failed on 4/14 and disagrees significantly with the upper 
wind direction sensor.  The lower wind sensor reports a constant value around 0-15 
degrees since it failed.  Testing showed that excitation channel two on the logger was the 
problem.  Logger was replaced by spare.
Measurements:twpsmet60sC3.b1:
  • lo_dir_sd
  • lo_dir_vec_avg


Back To Table of Contents

DQRID : D050808.3
Start DateStart TimeEnd DateEnd Time
08/05/2005003008/05/20050330
Subject:
TWP/SMET/C3 - Data not representative
DataStreams:twpsmet60sC3.b1
Description:
Tower was lowered for radiometer change and characterization. Data are not representative 
of local conditions.
Measurements:twpsmet60sC3.b1:
  • lo_dir_sd
  • tbrg_count_total
  • Vapor pressure mean(vappress_mean)
  • Relative humidity mean(relh_mean)
  • up_spd_max
  • up_spd_min
  • lo_spd_vec_avg
  • lo_spd_sd
  • precip_sd
  • lo_spd_min
  • Barometric Pressure(atmos_pressure)
  • temp_sd
  • up_spd_arith_avg
  • lo_spd_max
  • vappress_sd
  • up_spd_vec_avg
  • relh_sd
  • up_dir_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • up_spd_sd
  • lo_dir_vec_avg
  • lo_spd_arith_avg
  • tbrg_total
  • Precipitation mean(precip_mean)
  • Mean Air Temperature or Hardware Error(temp_mean)
  • up_dir_vec_avg


Back To Table of Contents

DQRID : D051115.1
Start DateStart TimeEnd DateEnd Time
11/14/2005045311/15/20052215
Subject:
TWP/SMET/C3 - Upper wind sensor failure
DataStreams:twpsmet60sC3.b1
Description:
A lightning strike destroyed the upper wind sensor.
Measurements:twpsmet60sC3.b1:
  • up_spd_vec_avg
  • up_dir_sd
  • up_spd_max
  • up_spd_sd
  • up_spd_min
  • up_spd_arith_avg
  • up_dir_vec_avg


Back To Table of Contents

DQRID : D051117.2
Start DateStart TimeEnd DateEnd Time
11/14/2005045311/15/20050050
Subject:
TWP/SMET/C3 - RH Values Artificially High
DataStreams:twpsmet60sC3.b1
Description:
Lightning strike caused RH readings to be higher than reasonable (130%).
Measurements:twpsmet60sC3.b1:
  • relh_sd
  • Vapor pressure mean(vappress_mean)
  • vappress_sd
  • Relative humidity mean(relh_mean)


Back To Table of Contents

DQRID : D051117.3
Start DateStart TimeEnd DateEnd Time
11/14/2005045311/23/20050000
Subject:
TWP/SMET/C3 - Instrument failure
DataStreams:twpsmet60sC3.b1
Description:
A lightning strike caused a failure in the ORG.  Rainfall was being continually recorded.
Measurements:twpsmet60sC3.b1:
  • precip_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • Precipitation mean(precip_mean)


Back To Table of Contents

DQRID : D060426.8
Start DateStart TimeEnd DateEnd Time
12/01/2005000008/24/20062359
Subject:
TWP/SMET/C3 - ORG data noisy due to degraded ground
DataStreams:twpsmet60sC3.b1
Description:
SMET Rain Data began to be noisy during December 2005 and the problem slowly became worse 
over time so that in February 2006 the ORG was reporting with a constant background value 
of around 0.10 mm/hr rain rate.  ORG voltage showed a near constant value above the 85mV 
threshold.  The ground for the ORG had become degraded and was lost due to corrosion.
Measurements:twpsmet60sC3.b1:
  • precip_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • Precipitation mean(precip_mean)


Back To Table of Contents

DQRID : D060502.3
Start DateStart TimeEnd DateEnd Time
04/27/2006233005/02/20060201
Subject:
TWP/SMET/C3 - Reprocessed: Lower windspeed sensor calibration error
DataStreams:twpsmet60sC3.b1
Description:
The lower anemometer was replaced but the calibration coefficients were not changed during 
this period.  The data were reprocessed to apply the correct calibrations and the 
reprocessed data were archived in August 2008.
Measurements:twpsmet60sC3.b1:
  • lo_spd_vec_avg
  • lo_spd_sd
  • lo_spd_min
  • lo_spd_arith_avg
  • lo_spd_max


Back To Table of Contents

DQRID : D060627.1
Start DateStart TimeEnd DateEnd Time
06/27/2006011506/27/20060344
Subject:
TWP/SMET/C3 - Data not representative
DataStreams:twpsmet60sC3.b1
Description:
The Anemometer readings were incorrect due to the tower being lowered for maintenance.
Measurements:twpsmet60sC3.b1:
  • lo_spd_sd
  • lo_spd_vec_avg
  • lo_spd_min
  • lo_dir_sd
  • up_spd_arith_avg
  • lo_spd_max
  • up_spd_vec_avg
  • up_dir_sd
  • up_spd_max
  • lo_dir_vec_avg
  • up_spd_min
  • up_spd_sd
  • lo_spd_arith_avg
  • up_dir_vec_avg


Back To Table of Contents

DQRID : D061011.4
Start DateStart TimeEnd DateEnd Time
03/12/2002000009/26/20061435
Subject:
TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added
DataStreams:twpsmet60sC3.b1
Description:
A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. 
 Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 
datastream.  When reprocessing of this datastream occurs these variables will be added 
with -9999 (missing) as the value for dates prior to gauge installation.
Measurements:twpsmet60sC3.b1:
  • base_time


Back To Table of Contents

DQRID : D061221.2
Start DateStart TimeEnd DateEnd Time
11/01/2006000011/30/20062359
Subject:
TWP/RAIN/C3 - faulty precipitation result from blocked funnel
DataStreams:twprainC3.b1
Description:
The funnel was partially blocked and the water was flowing so slowly through the system 
that the timing of precipitation was wrong.
Measurements:twprainC3.b1:
  • Rainfall rate(rain_rate)
  • Precipitation(precip_tbrg)


Back To Table of Contents

DQRID : D070306.2
Start DateStart TimeEnd DateEnd Time
03/02/2007172303/03/20071228
Subject:
TWP/SMET/C3 - Barometer Data Incorrect
DataStreams:twpsmet60sC3.b1
Description:
The barometer data was incorrect or missing due to water intrusion.
Measurements:twpsmet60sC3.b1:
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D070306.3
Start DateStart TimeEnd DateEnd Time
03/03/2007005603/03/20070123
Subject:
TWP/SMET/C3 - Water in enclosure
DataStreams:twpsmet60sC3.b1
Description:
Data became erratic and unreasonable after water intrusion into the enclosure.
Measurements:twpsmet60sC3.b1:
  • relh_sd
  • precip_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • temp_sd
  • Vapor pressure mean(vappress_mean)
  • Mean Air Temperature or Hardware Error(temp_mean)
  • Precipitation mean(precip_mean)
  • vappress_sd
  • Relative humidity mean(relh_mean)


Back To Table of Contents

DQRID : D070803.1
Start DateStart TimeEnd DateEnd Time
07/01/2007000009/18/20072315
Subject:
TWP/SMET/C3 - Upper wind speed artificially slow
DataStreams:twpsmet60sC3.b1
Description:
The upper wind measurement was consistently slower by around 0.5 - 1 m/s than the lower 
wind measurement.  The bearings on the upper wind measurement were binding.
Measurements:twpsmet60sC3.b1:
  • up_spd_vec_avg
  • up_spd_max
  • up_spd_sd
  • up_spd_min
  • up_spd_arith_avg


Back To Table of Contents

DQRID : D071105.1
Start DateStart TimeEnd DateEnd Time
11/02/2007185511/02/20072010
Subject:
TWP/SMET/C3 - ORG overestimated rainfall
DataStreams:twpsmet60sC3.b1
Description:
Optical rain gauge (ORG) measured approximately 10mm higher than two co-located tipping 
bucket rain gauges and a co-located disdrometer.  The ORG is typically higher but usually 
not by an order of magnitude.
Measurements:twpsmet60sC3.b1:
  • precip_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • Precipitation mean(precip_mean)


Back To Table of Contents

DQRID : D080225.1
Start DateStart TimeEnd DateEnd Time
01/05/2008000002/27/20080315
Subject:
TWP/SMET/C3 - Upper Wind Sensor Data Questionable
DataStreams:twpsmet60sC3.b1
Description:
The wind_up vs. wind_low wind directions diverged, with the differences between the two 
eventually reached 45 degrees. The Upper sensor was found to be mis-aligned and loose.
Measurements:twpsmet60sC3.b1:
  • up_spd_vec_avg
  • up_spd_max
  • up_spd_sd
  • up_spd_min
  • up_spd_arith_avg


Back To Table of Contents

DQRID : D080428.3
Start DateStart TimeEnd DateEnd Time
05/01/2004000006/30/20102359
Subject:
TWP/SMET/C3 - Reprocess: ORG Data has very small rainrate values
DataStreams:twpsmet60sC3.b1
Description:
The ORG data is collected via analog signal and not digital.  Therefore, we
continuously collect data even when no rain is occurring.  We also capture small events 
such as leaves, large bugs, dust, etc blowing through the measurement path.  Because this 
discrete event occurs quickly the value is usually captured for 1 or 2 seconds.  This 
value is then averaged to the minute producing rainrate on the order of .009 mm/hr or less.  

These values are not indicative of rain and should be removed from the record or summing 
over long periods can bias the climate record.  We will reprocess the data to remove these 
small values.
Measurements:twpsmet60sC3.b1:
  • precip_sd
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • Precipitation mean(precip_mean)


Back To Table of Contents

DQRID : D080611.2
Start DateStart TimeEnd DateEnd Time
05/27/2008194506/24/20080113
Subject:
TWP/SMET/C3 - Lower Wind Direction Intermittently Wrong
DataStreams:twpsmet60sC3.b1
Description:
The lower wind sensor cabling had a problem causing an intermittent connection for the 
wind direction data.
Measurements:twpsmet60sC3.b1:
  • lo_dir_sd
  • lo_dir_vec_avg


Back To Table of Contents

DQRID : D080715.1
Start DateStart TimeEnd DateEnd Time
07/08/2008204507/17/20080144
Subject:
TWP/SMET/C3 - Upper wind speed data incorrect
DataStreams:twpsmet60sC3.b1
Description:
Beginning on 07/08, the upper wind sensor started reporting significantly lower wind 
speeds than the lower wind sensor.  The cable connections weather proofing failed allowing 
corrosion to occur and increasing he resistance of the cables.  This in turn lowered the 
wind speeds artificially.
Measurements:twpsmet60sC3.b1:
  • up_spd_vec_avg
  • up_spd_max
  • up_spd_sd
  • up_spd_min
  • up_spd_arith_avg


Back To Table of Contents

DQRID : D080922.5
Start DateStart TimeEnd DateEnd Time
05/17/2004220205/19/20040559
04/23/2006235704/27/20060348
Subject:
TWP/SMET/C3 - Missing data
DataStreams:twpsmet60sC3.b1
Description:
Data are missing and unrecoverable.
Measurements:twpsmet60sC3.b1:
  • alt
  • lo_dir_sd
  • lon
  • base_time
  • tbrg_count_total
  • time_offset
  • Vapor pressure mean(vappress_mean)
  • Relative humidity mean(relh_mean)
  • logger_temp
  • up_spd_max
  • up_spd_min
  • lo_spd_vec_avg
  • lo_spd_sd
  • precip_sd
  • lo_spd_min
  • Barometric Pressure(atmos_pressure)
  • temp_sd
  • lat
  • up_spd_arith_avg
  • lo_spd_max
  • vappress_sd
  • up_spd_vec_avg
  • relh_sd
  • up_dir_sd
  • time
  • Precipitation maximum(precip_max)
  • Precipitation minimum(precip_min)
  • up_spd_sd
  • lo_dir_vec_avg
  • internal_voltage
  • lo_spd_arith_avg
  • tbrg_total
  • Precipitation mean(precip_mean)
  • Mean Air Temperature or Hardware Error(temp_mean)
  • up_dir_vec_avg


Back To Table of Contents

DQRID : D081022.8
Start DateStart TimeEnd DateEnd Time
03/12/2002000010/17/20080627
Subject:
TWP/SMET/C3 - Wind Speed data contain offset due to induced voltage
DataStreams:twpsmet60sC3.b1
Description:
The marine model of the RM Young wind sensor needs twisted pair shielded cable.  The 
manufacturer did not originally supply the pigtail with twisted pair shielded cable.  Use of 
twisted pair shielded cable for the rest of the cable run to the logger was also not 
specified.  This caused an induced voltage in the wind speed line from the wind direction 
excitation.  The induced voltage varied from sensor to sensor and probably changed anytime 
the cable was replaced or parts of the wind sensor were repaired or replaced. The induced 
voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s.
Measurements:twpsmet60sC3.b1:
  • up_spd_vec_avg
  • lo_spd_vec_avg
  • lo_spd_sd
  • lo_spd_min
  • up_spd_max
  • up_spd_sd
  • up_spd_min
  • lo_spd_arith_avg
  • up_spd_arith_avg
  • lo_spd_max


Back To Table of Contents

DQRID : D090129.4
Start DateStart TimeEnd DateEnd Time
01/27/2009185001/30/20091535
Subject:
TWP/SMET/C3 - Reprocess: Station Pressure values reported in hPa instead of kPa.
DataStreams:twpsmet60sC3.b1
Description:
An incorrect program was entered into the datalogger.  This program did not convert the 
station pressure from hPa to kPa.  This made the values larger by a factor of 10 and caused 
the QC bit to be tripped.
Measurements:twpsmet60sC3.b1:
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D090225.3
Start DateStart TimeEnd DateEnd Time
05/19/2004051904/19/20050419
Subject:
TWP/SMET/C3 - Reprocess: Temperature data biased low
DataStreams:twpsmet60sC3.b1
Description:
Due to a datalogger problem, the temperature data were biased low by 2.13 degrees C.

The data will be reprocessed.
Measurements:twpsmet60sC3.b1:
  • temp_sd
  • Mean Air Temperature or Hardware Error(temp_mean)


Back To Table of Contents

DQRID : D090515.1
Start DateStart TimeEnd DateEnd Time
04/19/2005042004/04/20090115
Subject:
TWP/SMET/C3 - Reprocess: Temperature Biased low by 1.06 Degrees
DataStreams:twpsmet60sC3.b1
Description:
The temperature of the SMET system was biased low by 1.06 Deg C due to a non-precision 
resistor being used int he measurement circuit.  The data will be reprocessed.  Saturation 
Vapor pressure and Vapor pressure will have to be reprocessed also.

Add 1.06 degrees to the temperature data.  

Correct Sat Vap Press using the Buck approximation:
   Es = 6.1121exp((17.502*T)/(240.97 + T) 
where T is the temperature in C.  

Correct Vapor pressure using: E = (RH*Es)/100
Measurements:twpsmet60sC3.b1:
  • temp_sd
  • Mean Air Temperature or Hardware Error(temp_mean)
  • Vapor pressure mean(vappress_mean)
  • vappress_sd


Back To Table of Contents

DQRID : D091023.1
Start DateStart TimeEnd DateEnd Time
09/30/2009004509/30/20090500
Subject:
TWP/SMET/C3 - Tower Lowered for Maintenance
DataStreams:twpsmet60sC3.b1
Description:
The tower was lowered for annual anemometer changes.  The wind data is suspect.
Measurements:twpsmet60sC3.b1:
  • lo_spd_sd
  • lo_spd_vec_avg
  • lo_spd_min
  • lo_dir_sd
  • up_spd_arith_avg
  • lo_spd_max
  • up_spd_vec_avg
  • up_dir_sd
  • up_spd_max
  • lo_dir_vec_avg
  • up_spd_min
  • up_spd_sd
  • lo_spd_arith_avg
  • up_dir_vec_avg


Back To Table of Contents

DQRID : D120614.1
Start DateStart TimeEnd DateEnd Time
06/14/2012000006/14/20120605
Subject:
TWP/MET/C3 - Incorrect ORG data
DataStreams:twpmetC3.b1
Description:
Sprinkler was in the path of the ORG, giving false precipitation data.
Measurements:twpmetC3.b1:
  • Precipitation mean(precip_mean)
  • qc_precip_mean
  • qc_precip_max
  • qc_precip_min
  • precip_std
  • Precipitation minimum(precip_min)
  • Precipitation maximum(precip_max)


Back To Table of Contents

DQRID : D130201.1
Start DateStart TimeEnd DateEnd Time
12/26/2012000001/26/20132359
Subject:
TWP/DISDROMETER/C3 - Precip not measured
DataStreams:twpdisdrometerC3.b1
Description:
The reason remains unknown but the impact disdrometer stopped working.  The device was 
almost completely unresponsive.

Researchers interested in the rainfall drop distribution can turn to the other disdrometer 
observations made by the 2dvd instrument.  Those data can be found in the VDIS 
datastreams.
Measurements:twpdisdrometerC3.b1:
  • num_drop
  • distribution_slope
  • distribution_intercept
  • number_density
  • mean_diam_drop_class
  • Precipitation(precip_dis)
  • fall_velocity
  • Rainfall rate(rain_rate)
  • energy_flux
  • delta_diam
  • radar_reflectivity
  • diam_max
  • liq_water
  • drop_class


Back To Table of Contents