DQR ID | Subject | Data Streams Affected |
---|---|---|
D080502.2 | SGP/RWP/C1 - Reprocess: Instrument in non-standard mode for IOP | sgp915rwptempconC1.a1, sgp915rwptempmomC1.a0, sgp915rwptempspecC1.a0, sgp915rwpwindconC1.a1, sgp915rwpwindmomC1.a0, sgp915rwpwindspecC1.a0 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/06/2007 | 1900 | 10/03/2007 | 1500 |
Subject: | SGP/RWP/C1 - Reprocess: Instrument in non-standard mode for IOP |
DataStreams: | sgp915rwptempconC1.a1, sgp915rwptempmomC1.a0, sgp915rwptempspecC1.a0, sgp915rwpwindconC1.a1, sgp915rwpwindmomC1.a0, sgp915rwpwindspecC1.a0 |
Description: | The SGP.C1 915RWP was operated in a non-standard mode during the Nocturnal Flux IOP (approximately the month of September). The profiler was placed into a constant high temporal and spatial resolution mode rather than alternating between modes as it does routinely. 30 min averages were produced (usually 60 min) at a 60 m gate resolution for both winds and RASS. Data were processed and archived, but the processing algorithm interpreted this data as 2 different power levels per hourly sample rather than 30 minute samples at the same power level. Data will be reprocessed. |
Measurements: | sgp915rwptempmomC1.a0:
sgp915rwpwindmomC1.a0:
sgp915rwpwindconC1.a1:
sgp915rwpwindspecC1.a0:
sgp915rwptempconC1.a1:
sgp915rwptempspecC1.a0:
|