D120124.5TWP/MWR3C/C3 - Incorrect calibrationtwpmwr3cC3.b1


DQRID : D120124.5
Start DateStart TimeEnd DateEnd Time
12/01/2011000012/10/20110000
Subject:
TWP/MWR3C/C3 - Incorrect calibration
DataStreams:twpmwr3cC3.b1
Description:
When the radiometer was relocated to the TWP site the 89-GHz detector non-linearity 
coefficient got somehow erased from the radiometer memory. The radiometer was therefore unable 
to calibrate properly.
Measurements:twpmwr3cC3.b1:
  • Radiation, longwave, brightness temperature, 23.8 GHz(tbsky23)
  • 89.0 GHz sky brightness temperature(tbsky89)
  • Water vapor, precipitable, column-integrated on isobaric grid, from RUC model(pwv)
  • Liquid water path best-estimate value, hourly mean(lwp)
  • 30.0 GHz sky brightness temperature(tbsky30)