DQR ID | Subject | Data Streams Affected |
---|---|---|
D070117.1 | NSA/GVR/C1 - Missing data | nsagvrC1.b1, nsagvrC1.a0, nsagvrC1.00 |
D070515.2 | NSA/GVR/C1 - Instrument calibration | nsagvrC1.a0, nsagvrC1.b1 |
D070710.3 | NSA/GVR/C1 - Temperature sensor wire broken | nsagvrC1.a0, nsagvrC1.b1 |
D070713.1 | NSA/GVR/C1 - Incorrect auxiliary measurements | nsagvrC1.a0, nsagvrC1.b1 |
D071214.1 | NSA/GVR/C1 - Incorrect ambient temperature readings | nsagvrC1.a0, nsagvrC1.b1 |
D100809.2 | NSA/GVR/C1 - Saturated brightness temperature | nsagvrC1.a0, nsagvrC1.b1, nsagvrrawC1.00 |
D100909.2 | NSA/GVR/C1 - Saturated brightness temperatures | nsagvrC1.a0, nsagvrC1.b1 |
D110210.1 | NSA/GVRP/C1 - Reprocessed: Calibration drift | nsagvrpC1.b1 |
D110210.2 | NSA/GVRP/C1 - Missing data | nsagvrpC1.b1 |
D110304.2 | NSA/GVRP/C1 - Reprocessed: Calibration drift | nsagvrpC1.b1 |
D110405.1 | NSA/GVRP/C1 - Incorrect surface temp and RH | nsagvrpC1.b1 |
D110615.1 | NSA/GVRP/C1 - Reprocess: Calibration drift | nsagvrpC1.b1 |
D110927.3 | NSA/GVRP/C1 - High noise in the data | nsagvrpC1.b1 |
D120418.2 | NSA/GVRP/C1 - Reprocess: Radiometer Calibration | nsagvrpC1.b1 |
D120423.2 | NSA/GVRP/C1 - Incorrect surface temperature and humidity | nsagvrpC1.b1 |
D121015.1 | NSA/GVR/C1 - Saturated brightness temperatures | nsagvrC1.a0, nsagvrC1.b1 |
D121217.1 | NSA/GVRP/C1 - Calibration drift | nsagvrpC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/30/2012 | 1200 | 07/02/2012 | 1700 |
04/14/2012 | 2100 | 04/14/2012 | 1353 |
03/03/2012 | 0800 | 03/05/2012 | 2000 |
02/26/2012 | 1500 | 02/28/2012 | 1700 |
01/11/2007 | 1500 | 01/15/2007 | 1600 |
04/13/2007 | 2122 | 04/18/2007 | 1325 |
04/30/2007 | 2020 | 05/02/2007 | 1256 |
06/09/2008 | 0500 | 06/20/2008 | 1852 |
11/10/2008 | 1903 | 11/14/2008 | 1616 |
08/03/2009 | 1700 | 10/26/2009 | 1800 |
05/08/2010 | 2200 | 05/11/2010 | 1800 |
09/27/2010 | 0300 | 10/01/2010 | 2200 |
10/08/2010 | 1500 | 10/11/2010 | 1900 |
12/25/2010 | 0800 | 01/03/2011 | 2300 |
01/12/2011 | 0600 | 01/15/2011 | 1700 |
04/26/2011 | 0500 | 05/02/2011 | 1804 |
05/17/2011 | 0300 | 05/19/2011 | 1900 |
06/18/2011 | 0900 | 06/20/2011 | 1900 |
07/09/2011 | 2300 | 07/11/2011 | 2300 |
12/10/2011 | 0400 | 12/14/2011 | 2000 |
Subject: | NSA/GVR/C1 - Missing data |
DataStreams: | nsagvrC1.b1, nsagvrC1.a0, nsagvrC1.00 |
Description: | Data are missing and unrecoverable. |
Measurements: | nsagvrC1.b1:
nsagvrC1.a0:
nsagvrC1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/11/2007 | 1741 | 05/17/2007 | 0000 |
Subject: | NSA/GVR/C1 - Instrument calibration |
DataStreams: | nsagvrC1.a0, nsagvrC1.b1 |
Description: | During the period of time specified, the instrument was undergoing calibration. The antenna pointed constantly to a hot load placed on the instrument window. All brightness temperatures during this time have temperatures > 300 K. |
Measurements: | nsagvrC1.b1:
nsagvrC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/30/2007 | 1900 | 07/13/2007 | 2200 |
Subject: | NSA/GVR/C1 - Temperature sensor wire broken |
DataStreams: | nsagvrC1.a0, nsagvrC1.b1 |
Description: | The temp_ext temperatures were occasionally spiking to very large values. Comparisons with the co-located 2m METTWR suggest the GVR exterior temperatures were not reporting correctly. Site ops found the temperature sensor had a broken wire and repaired it 7/13. |
Measurements: | nsagvrC1.b1:
nsagvrC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2007 | 1700 | 06/19/2007 | 1700 |
Subject: | NSA/GVR/C1 - Incorrect auxiliary measurements |
DataStreams: | nsagvrC1.a0, nsagvrC1.b1 |
Description: | Between 6/16 and 6/19 there was a software problem and the auxiliary measurements all were reading constant values. The brightness temperature measurements do not appear to be affected. One of the incorrect measurements that may affect the use of the data is the surface temperature. This measurement should be disregarded. Measurements from a different instrument (such as the MWR) can be used to replace the GVR readings. |
Measurements: | nsagvrC1.b1:
nsagvrC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/03/2007 | 2000 | 05/22/2008 | 2200 |
Subject: | NSA/GVR/C1 - Incorrect ambient temperature readings |
DataStreams: | nsagvrC1.a0, nsagvrC1.b1 |
Description: | The ambient temperature readings are incorrect. The temperature sensor was replaced on 5/13, however the temperature readings were incorrect until 5/22 due to problems with corrosion in the cables |
Measurements: | nsagvrC1.b1:
nsagvrC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/16/2010 | 1200 | 07/22/2010 | 2000 |
Subject: | NSA/GVR/C1 - Saturated brightness temperature |
DataStreams: | nsagvrC1.a0, nsagvrC1.b1, nsagvrrawC1.00 |
Description: | Beginning 7/16 all sky temperature data become nearly equal and abnormally constant. The sky temperature data returned to normal on 7/22 ~20:00 UTC. The saturated brightness temperatures were probably due to water or dew deposit on the window. |
Measurements: | nsagvrC1.b1:
nsagvrrawC1.00:
nsagvrC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/26/2010 | 1200 | 08/30/2010 | 1500 |
Subject: | NSA/GVR/C1 - Saturated brightness temperatures |
DataStreams: | nsagvrC1.a0, nsagvrC1.b1 |
Description: | Brightness temperatures are saturated probably because of water on the radome. Water vapor retrievals are affected as well. |
Measurements: | nsagvrC1.b1:
nsagvrC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2010 | 0000 | 01/19/2011 | 1400 |
Subject: | NSA/GVRP/C1 - Reprocessed: Calibration drift |
DataStreams: | nsagvrpC1.b1 |
Description: | The most transparent channels (170-176 GHz) exhibited signs of a drift in the calibration. The radiometer was calibrated on Jan 17 2011. Data for 20101101-20110117 were recalibrated and reprocessed. The reprocessed data were archived in May 2011. After the calibration there was a problem restarting the software. The software was successfully restarted on Jan 19. |
Measurements: | nsagvrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2011 | 2200 | 03/02/2012 | 1400 |
01/17/2011 | 1900 | 01/19/2011 | 1400 |
04/27/2011 | 1707 | 05/02/2011 | 1700 |
Subject: | NSA/GVRP/C1 - Missing data |
DataStreams: | nsagvrpC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | nsagvrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/24/2011 | 2300 | 03/04/2011 | 0000 |
Subject: | NSA/GVRP/C1 - Reprocessed: Calibration drift |
DataStreams: | nsagvrpC1.b1 |
Description: | On Jan 24 2011 there was a power outage that caused another drift in the calibration. The radiometer was put in scanning mode to collect tip data. The tip data were processed for the most transparent channels. New calibration values were uploaded and took effect on March 4 2011. Data for 20110124-20110304 were recalibrated and reprocessed. The reprocessed data were archived in May 2011. |
Measurements: | nsagvrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/23/2011 | 0538 | 03/24/2011 | 2000 |
Subject: | NSA/GVRP/C1 - Incorrect surface temp and RH |
DataStreams: | nsagvrpC1.b1 |
Description: | The instrument experienced a period of missing data on 3/23/2011 from 05:38:30 GMT until 08:21:44 GMT. When data returned the surface temperature and surface relative humidity were incorrect. Surface temperature flat-lined at 999.990 K and surface RH flat-lined at 999.990%. On 3/24 the sensor was removed and cleaned resolving the problem. |
Measurements: | nsagvrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/02/2011 | 1700 | 05/19/2011 | 0000 |
Subject: | NSA/GVRP/C1 - Reprocess: Calibration drift |
DataStreams: | nsagvrpC1.b1 |
Description: | After the power outage there was a shift in the calibration of the instrument. Data will be reprocessed with updated coefficients. |
Measurements: | nsagvrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/01/2011 | 0000 | 10/13/2011 | 2200 |
Subject: | NSA/GVRP/C1 - High noise in the data |
DataStreams: | nsagvrpC1.b1 |
Description: | All channels of the radiometer became suddenly very noisy. The receiver noise figure was computed and was found anomalous. Data are unusable during the indicated times. On 10/13 the instrument was removed and will be sent to the vendor for repairs. |
Measurements: | nsagvrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2012 | 1400 | 04/13/2012 | 1816 |
Subject: | NSA/GVRP/C1 - Reprocess: Radiometer Calibration |
DataStreams: | nsagvrpC1.b1 |
Description: | The radiometer was calibrated with LN2 on 4/13/2012. Data between 3/2 and 4/13 will be recalibrated with the most current coefficients. |
Measurements: | nsagvrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2012 | 1408 | 04/19/2012 | 0020 |
Subject: | NSA/GVRP/C1 - Incorrect surface temperature and humidity |
DataStreams: | nsagvrpC1.b1 |
Description: | Surface Relative Humidity and Temperate values are incorrect or missing between 3/2/2012 and 4/19/2012 |
Measurements: | nsagvrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/23/2012 | 0000 | 09/27/2012 | 2300 |
Subject: | NSA/GVR/C1 - Saturated brightness temperatures |
DataStreams: | nsagvrC1.a0, nsagvrC1.b1 |
Description: | The brightness temperatures were saturated between 9/23 and 9/27. Data should be disregarded. |
Measurements: | nsagvrC1.b1:
nsagvrC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/2012 | 0000 | 11/15/2012 | 2000 |
Subject: | NSA/GVRP/C1 - Calibration drift |
DataStreams: | nsagvrpC1.b1 |
Description: | The radiometer showed a progressive drift in calibration starting approximately on 10/10/2012 and becoming worst in November. The instrument was calibrated with liquid nitrogen on 11/15/2012 |
Measurements: | nsagvrpC1.b1:
|