DQR ID | Subject | Data Streams Affected |
---|---|---|
D111207.1 | GAN/MWR3C/M1 - Intermittent data | ganmwr3cM1.b1 |
D120124.6 | GAN/MWR3C/M1 - 30-Ghz channel unstable | ganmwr3cM1.b1 |
D120220.1 | GAN/MWR3C/M1 - Radiometer instability | ganmwr3cM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2011 | 1200 | 11/01/2011 | 1800 |
Subject: | GAN/MWR3C/M1 - Intermittent data |
DataStreams: | ganmwr3cM1.b1 |
Description: | The data stops every day at 12 pm because of an error in the schedules tasks. The problem was fixed on 11/1/2011. |
Measurements: | ganmwr3cM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/2011 | 0000 | 12/08/2011 | 0000 |
Subject: | GAN/MWR3C/M1 - 30-Ghz channel unstable |
DataStreams: | ganmwr3cM1.b1 |
Description: | The 30-GHz channel was unstable between Dec 1 and Dec 5. Calibration was restored on Dec 9. |
Measurements: | ganmwr3cM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/04/2012 | 0000 | 01/08/2012 | 2359 |
01/12/2012 | 0000 | 01/20/2012 | 0000 |
01/31/2012 | 0000 | 02/01/2012 | 2359 |
02/07/2012 | 2100 | 02/08/2012 | 0500 |
Subject: | GAN/MWR3C/M1 - Radiometer instability |
DataStreams: | ganmwr3cM1.b1 |
Description: | The radiometer showed intermittent instability in the 30 GHz and 89 GHz channel. There was a calibration shift between Jan 4 and Jan 8 and a calibration instability between Jan 12 and Jan 20. It may be possible to recalibrate the data when a recalibration algorithm is developed. |
Measurements: | ganmwr3cM1.b1:
|