DQR ID | Subject | Data Streams Affected |
---|---|---|
D040525.28 | SGP/MPL/C1 - Metadata errors | sgpmplC1.a1, sgpmplnor1campC1.c1 |
D980331.1 | Instrument misalignment report | sgpmplnor1campC1.c1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/1996 | 0000 | 01/19/2000 | 2359 |
Subject: | SGP/MPL/C1 - Metadata errors |
DataStreams: | sgpmplC1.a1, sgpmplnor1campC1.c1 |
Description: | The latitude, longitude and altitude of the Micropulse Lidar (MPL) were incorrectly entered into the ARM database. The correct location of the SGP.C1 MPL is: Lat: 36.606N Lon: 97.485W Alt: 316m |
Measurements: | sgpmplC1.a1:
sgpmplnor1campC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/15/1997 | 0000 | 09/10/1997 | 0000 |
Subject: | Instrument misalignment report |
DataStreams: | sgpmplnor1campC1.c1 |
Description: | A low-resolution (300 m) MPL system (unit number 04) was installed at the SGP site in April '97 upon the laser diode failure of the 02 unit in March. Calibration of this instrument in August found that the optics were misaligned causing a near range focusing error in the data. Calibration files produced for this data, while acceptable, do not completely correct for the problem. As a result, aerosol property retrievals using this data are not suggested. Cloud boundary heights, though appear satisfactory. This system was removed from SGP in early September |
Measurements: | sgpmplnor1campC1.c1:
|