Subject: | GRW/WACR/M1 - Reprocessed: Reflectivity and Height Offsets |
DataStreams: | grwwacrM1.b1, grwwacrspeccmaskcopolM1.a0, grwwacrspeccmaskxpolM1.a0
|
Description: | The instrument vendor informed us of a range (height) offset of 13 m and provided updated
reflectivity calibration. All values reported for 'Reflectivity' were approximately 3 dB
too low. In addition, the Reflectivity in the first and fourth range gates (starting
from the surface) needed to be further corrected by an additional 2.7 and -0.9 dB,
respectively. The 'height' field values were 13.33 m too low. The data were reprocessed to
correct these problems.
The grwwacrM1.b1 data were reprocessed and archived in October 2010. The
grwwacrspeccmaskcopolM1.a0 and grwwacrspeccmaskxpolM1.a0 data were reprocessed and archived in January
2012. |
Measurements: | grwwacrspeccmaskcopolM1.a0: - Cloud, reflectivity, at altitude(Reflectivity)
- heights
grwwacrM1.b1: - Cloud, reflectivity, at altitude(Reflectivity)
- heights
grwwacrspeccmaskxpolM1.a0: - Cloud, reflectivity, at altitude(Reflectivity)
- heights
|
Subject: | GRW/WACR/M1 - Transmit power drop outs |
DataStreams: | grwwacrM1.b1, grwwacrspeccmaskcopolM1.a0, grwwacrspeccmaskxpolM1.a0
|
Description: | At approximately 16:27 GMT on 20090727, txpower began intermittently dropping out and
reporting slightly negative values. This issue continued intermittently through 20090807
when it stopped occurring. The problem resumed 20091006 and continued to occur on a daily
basis until the monitor detector was replaced 20101019.
The transmit power as measured by the detector after replacement was approximately 1.3 dB
less than when it failed. Reflectivity measurements have a +0/-1.3 dBZ of additional
uncertainty for this period of time. |
Measurements: | grwwacrspeccmaskcopolM1.a0: - txpower
- Cloud, reflectivity, at altitude(Reflectivity)
grwwacrM1.b1: - txpower
- Cloud, reflectivity, at altitude(Reflectivity)
grwwacrspeccmaskxpolM1.a0: - Cloud, reflectivity, at altitude(Reflectivity)
- txpower
|