DQR ID | Subject | Data Streams Affected |
---|---|---|
D101222.2 | TWP/SWF/C2 - instrument grounding problem | twp15swfanalskyrad1longC2.c1, twp1swfanalskyrad1longC2.c1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2009 | 0000 | 04/23/2009 | 2359 |
05/04/2009 | 0000 | 05/08/2009 | 2359 |
05/15/2009 | 0000 | 05/15/2009 | 2359 |
05/21/2009 | 0000 | 05/25/2009 | 2359 |
Subject: | TWP/SWF/C2 - instrument grounding problem |
DataStreams: | twp15swfanalskyrad1longC2.c1, twp1swfanalskyrad1longC2.c1 |
Description: | A grounding problem with the skyrad data logger (discussed in DQR D090605.2) created intermittent anomalous voltages and incorrect flux data. Measured all-sky fluxes and derived cloud fraction during these periods are incorrect. The identification of clear sky periods during these days and estimated clear sky fluxes may also be incorrect. Days listed in this DQR are periods where bad data could be clearly identified; data from other days in this date range should be treated with caution. |
Measurements: | twp15swfanalskyrad1longC2.c1:
twp1swfanalskyrad1longC2.c1:
|