DQR ID | Subject | Data Streams Affected |
---|---|---|
D040408.1 | SGP/ECOR/E16 - Data suspect due to inability to run QC | sgp30ecorE16.b1, sgpecorE16.00 |
D041103.2 | SGP/ECOR/E16 - Failed IRGA | sgp30ecorE16.b1 |
D041223.2 | SGP/ECOR/E16 - Missing data | sgp30ecorE16.b1 |
D050421.2 | SGP/ECOR/E16 - Reprocess: Incorrectly flagged (marked as "bad") H2O/CO2 data | sgp30ecorE16.b1 |
D060104.1 | SGP/ECOR/E16 - Flux Measurements Incorrect | sgp30ecorE16.b1 |
D060619.4 | SGP/ECOR/E16 - LI-7500 Pressure and Temperature missing | sgp30ecorE16.b1 |
D060707.4 | SGP/ECOR/E16 - Latent Heat and CO2 Fluxes Incorrect | sgp30ecorE16.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2004 | 0000 | 04/28/2004 | 1500 |
Subject: | SGP/ECOR/E16 - Data suspect due to inability to run QC |
DataStreams: | sgp30ecorE16.b1, sgpecorE16.00 |
Description: | Some QC checks were unsuccessful due to a communication failure that resulted in contaminated QC field data. All the data that are based on water vapor density and CO2 concentration are "suspect" because many of the standard QC checks were not employed. |
Measurements: | sgpecorE16.00:
sgp30ecorE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/02/2004 | 0130 | 11/09/2004 | 1530 |
Subject: | SGP/ECOR/E16 - Failed IRGA |
DataStreams: | sgp30ecorE16.b1 |
Description: | The infrared gas analyzer (IRGA) failed on Nov. 2, 2004, 02:00; the system continued to operate without H2O and CO2 data until Nov. 9, 2004, 15:30; then it stopped due to another hardware failure. The system was returned to service on Nov. 12, 2004, 18:00. The latent heat flux and CO2 flux data are irrevocably lost for the period from Nov. 2, 2004, 02:00 to Nov. 12, 2004, 18:00. |
Measurements: | sgp30ecorE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/28/2006 | 1530 | 05/24/2006 | 1500 |
07/14/2006 | 0630 | 07/19/2006 | 1930 |
07/14/2005 | 1430 | 07/18/2005 | 1330 |
01/05/2005 | 0730 | 01/10/2005 | 1900 |
04/14/2004 | 1800 | 04/16/2004 | 1630 |
11/09/2004 | 1530 | 11/12/2004 | 1730 |
Subject: | SGP/ECOR/E16 - Missing data |
DataStreams: | sgp30ecorE16.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgp30ecorE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/23/2004 | 0000 | 03/13/2005 | 0030 |
Subject: | SGP/ECOR/E16 - Reprocess: Incorrectly flagged (marked as "bad") H2O/CO2 data |
DataStreams: | sgp30ecorE16.b1 |
Description: | Due to system clock instability on ECOR computers their system time is adjusted approximately once per hour; sometimes this adjustment leads to backward time jumps over midnight, into the previous day. In this case all the data based on CO2/H2O measurements might be erroneously marked as bad. Below is a list of times when that might have happen. SITE YEAR MONTH DAY HOUR MINUTE 16 2004 02 23 00 00 16 2004 02 28 00 00 16 2004 03 01 00 00 16 2004 03 09 00 00 16 2004 03 15 00 00 16 2004 03 17 00 00 16 2004 03 22 00 00 16 2004 03 30 00 00 16 2004 04 02 00 00 16 2004 07 22 00 00 16 2004 08 06 00 00 16 2004 09 05 00 00 16 2004 09 20 00 00 16 2005 03 13 00 00 |
Measurements: | sgp30ecorE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/09/2005 | 1630 | 01/18/2006 | 1230 |
Subject: | SGP/ECOR/E16 - Flux Measurements Incorrect |
DataStreams: | sgp30ecorE16.b1 |
Description: | The ECOR datastream was corrupted; this happens when the ECOR program produces a datastream of the wrong length (i.e., leaves one or more measurements out). Fluxes and some other measurements are incorrect. Reprocessing will not correct this problem; the raw data are corrupted. |
Measurements: | sgp30ecorE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/17/2006 | 0230 | 08/02/2006 | 1530 |
Subject: | SGP/ECOR/E16 - LI-7500 Pressure and Temperature missing |
DataStreams: | sgp30ecorE16.b1 |
Description: | The pressure and temperature measurements were not recorded; the LI-7500 serial datastream was not being captured by the ECOR system because of a serial port failure. Some degradation of the latent and sensible heat fluxes may have occurred during high summer temperatures. The error in the fluxes would be no more than 5 to 10% of the measurement, still well within the measurement error of the system. |
Measurements: | sgp30ecorE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/21/2006 | 1500 | 06/23/2006 | 1600 |
Subject: | SGP/ECOR/E16 - Latent Heat and CO2 Fluxes Incorrect |
DataStreams: | sgp30ecorE16.b1 |
Description: | The LI-7500 CO2/H2O water vapor sensor was removed during this period. |
Measurements: | sgp30ecorE16.b1:
|