DQR ID | Subject | Data Streams Affected |
---|---|---|
D020313.1 | SGP/THWAPS/C1 - THWAPS C1 miscalibrated | sgpthwapsC1.a1, sgpthwapsC1.b1 |
D030917.21 | SGP/THWAPS/B1/B5/B6/C1 - Missing data due to Ice storm and clock problems | sgpthwapsB1.00, sgpthwapsB1.a1, sgpthwapsB1.b1, sgpthwapsB5.00, sgpthwapsB5.a1, sgpthwapsB5.b1, sgpthwapsB6.00, sgpthwapsB6.a1, sgpthwapsB6.b1, sgpthwapsC1.00, sgpthwapsC1.a1, sgpthwapsC1.b1 |
D041123.1 | SGP/THWAPS/C1 - RH Sensor Calibration | sgpthwapsC1.b1 |
D050812.10 | SGP/THWAPS/C1 - Minor bias in humidity | sgpthwapsC1.a1, sgpthwapsC1.b1 |
D050812.12 | SGP/THWAPS/C1 - Minor bias in saturation vapor pressure | sgpthwapsC1.b1 |
D070613.2 | SGP/RAIN/C1 - Funnel clogged | sgprainC1.b1 |
D080116.24 | SGP/THWAPS/C1 - RH Data Incorrect | sgpthwapsC1.b1 |
D090211.2 | SGP/THWAPS/C1 - Barometer intermittent | sgpthwapsC1.b1 |
D090306.1 | SGP/THWAPS/C1 - Time stamp of data record incorrect | sgpthwapsC1.b1 |
D090824.2 | SGP/THWAPS/C1 - Wind speed incorrect | sgpthwapsC1.b1 |
D110724.1 | SGP/RAIN/C1 - Incorrect precipitation measurements | sgprainC1.b1 |
D111027.1 | SGP/RAIN/C1 - Missing data | sgprainC1.b1 |
D120113.5 | SGP/RAIN/C1 - Reprocess: Incorrect datestamp | sgprainC1.b1 |
D120301.2 | SGP/RAIN/C1 - Lightning strike | sgprainC1.b1 |
D130617.2 | SGP/RAIN/C1 - Instrument problem | sgprainC1.00, sgprainC1.b1, sgprainauxC1.00, sgprainauxC1.b1 |
D130711.1 | SGP/THWAPS/C1 - T/RH sensor failure | sgpthwapsC1.b1 |
D991008.1 | SGP/THWAPS - RH values low by factor of 1.1 | sgpthwapsC1.a1, sgpthwapsC1.b1, sgpthwapsC1.00, DsgpthwapsC1.a1, DsgpthwapsC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/09/2002 | 0340 | 04/23/2002 | 1418 |
Subject: | SGP/THWAPS/C1 - THWAPS C1 miscalibrated |
DataStreams: | sgpthwapsC1.a1, sgpthwapsC1.b1 |
Description: | Power outages apparently led to the THWAPS HMP233 temperature and RH sensor being replaced at C1. The power went out at 0340 on 3/9 was restored at 0635 and went out again at 2025. The new HMP233 appears to be incorrectly calibrated and all temperature data since its installation are wrong. If I am correct about the nature of the calibration problem, the true average temperature can be recovered from that recorded by a linear conversion, but I don't think its worth the trouble. |
Measurements: | sgpthwapsC1.b1:
sgpthwapsC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/09/2002 | 2055 | 03/11/2002 | 2005 |
03/13/2002 | 1755 | 03/14/2002 | 1505 |
Subject: | SGP/THWAPS/B1/B5/B6/C1 - Missing data due to Ice storm and clock problems |
DataStreams: | sgpthwapsB1.00, sgpthwapsB1.a1, sgpthwapsB1.b1, sgpthwapsB5.00, sgpthwapsB5.a1, sgpthwapsB5.b1, sgpthwapsB6.00, sgpthwapsB6.a1, sgpthwapsB6.b1, sgpthwapsC1.00, sgpthwapsC1.a1, sgpthwapsC1.b1 |
Description: | Data are missing due to power failures caused by an ice storm and resulting system/instrument clock problems. |
Measurements: | sgpthwapsC1.b1:
sgpthwapsB5.00:
sgpthwapsC1.a1:
sgpthwapsB6.a1:
sgpthwapsC1.00:
sgpthwapsB6.b1:
sgpthwapsB1.00:
sgpthwapsB1.b1:
sgpthwapsB1.a1:
sgpthwapsB6.00:
sgpthwapsB5.a1:
sgpthwapsB5.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/16/2004 | 2135 | 11/29/2004 | 1930 |
Subject: | SGP/THWAPS/C1 - RH Sensor Calibration |
DataStreams: | sgpthwapsC1.b1 |
Description: | The analog output for the T/RH probe was incorrectly calibrated. This caused excessively high readings of RH. |
Measurements: | sgpthwapsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/21/1999 | 0000 | 12/07/1999 | 1100 |
Subject: | SGP/THWAPS/C1 - Minor bias in humidity |
DataStreams: | sgpthwapsC1.a1, sgpthwapsC1.b1 |
Description: | The THWAPS humidity sensor output was scaled so that the instrument output voltage range (0-1VDC) represented 0-100 %RH. Comparisons with the co-located chilled mirror sensor at the CF suggested that the THWAPS RH sensor was reading artifically low at high humidities because of this 100 %RH maximum. The scaling of the system was changed so that 0-1 VDC now represents 0-110 %RH. This resulted in a slight loss of precision in the measurements, but since the RH measurement is only accurate to 2-3 %RH at best, this isn't a significant problem. |
Measurements: | sgpthwapsC1.b1:
sgpthwapsC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/21/1999 | 0000 | 08/10/2005 | 1445 |
Subject: | SGP/THWAPS/C1 - Minor bias in saturation vapor pressure |
DataStreams: | sgpthwapsC1.b1 |
Description: | The saturation vapor correction when temps are below 0C contained an error in the multiplier. The incorrect multiplier only equates to an error of 0.01 kPa which is less than the stated uncertainty of the measurement (+/- 0.035 kPa). |
Measurements: | sgpthwapsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/01/2007 | 0000 | 06/02/2007 | 2359 |
Subject: | SGP/RAIN/C1 - Funnel clogged |
DataStreams: | sgprainC1.b1 |
Description: | The funnel for the instrument was clogged and the results for this time period are not reliable. |
Measurements: | sgprainC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/22/2007 | 1750 | 12/22/2007 | 2130 |
Subject: | SGP/THWAPS/C1 - RH Data Incorrect |
DataStreams: | sgpthwapsC1.b1 |
Description: | The RH data was incorrect during the times listed. Reason unknown. The sensor recovered on its own. |
Measurements: | sgpthwapsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/2009 | 0000 | 02/13/2009 | 2359 |
Subject: | SGP/THWAPS/C1 - Barometer intermittent |
DataStreams: | sgpthwapsC1.b1 |
Description: | Intermittent dropouts on the barometer data were occurring, with data reading low or missing. Loose wiring was the probable culprit. In most instances, only 1 minute of data was missing. Linear interpolation between the last known good point and the first good point following the dropout can be accomplished. |
Measurements: | sgpthwapsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/2008 | 0000 | 01/23/2009 | 2359 |
Subject: | SGP/THWAPS/C1 - Time stamp of data record incorrect |
DataStreams: | sgpthwapsC1.b1 |
Description: | Due to problems with the collector system, the datalogger was being time synched inappropriately. This problem was intermittent and did not affect every time stamp. The worst case time adjustments were approximately +/- 7 minutes. These were infrequent and most of the time adjustments were less than 7 minutes. |
Measurements: | sgpthwapsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/2009 | 0000 | 11/27/2009 | 1745 |
Subject: | SGP/THWAPS/C1 - Wind speed incorrect |
DataStreams: | sgpthwapsC1.b1 |
Description: | The wind sensor was reporting 0 m/s winds for excessive periods of time and appeared to measure lower wind speeds than nearby co-located wind sensors. Sensor was replaced. |
Measurements: | sgpthwapsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/24/2010 | 0000 | 04/24/2010 | 2359 |
04/26/2010 | 0000 | 04/26/2010 | 2359 |
04/30/2010 | 0000 | 04/30/2010 | 2359 |
05/03/2010 | 0000 | 05/06/2010 | 2359 |
05/09/2010 | 0000 | 05/10/2010 | 2359 |
05/12/2010 | 0000 | 05/17/2010 | 2359 |
05/19/2010 | 0000 | 05/25/2010 | 2359 |
05/29/2010 | 0000 | 05/29/2010 | 2359 |
05/31/2010 | 0000 | 05/31/2010 | 2359 |
06/04/2010 | 0000 | 06/04/2010 | 2359 |
06/07/2010 | 0000 | 06/07/2010 | 2359 |
06/09/2010 | 0000 | 06/10/2010 | 2359 |
06/14/2010 | 0000 | 06/16/2010 | 2359 |
06/20/2010 | 0000 | 06/20/2010 | 2359 |
07/02/2010 | 0000 | 07/07/2010 | 2359 |
07/10/2010 | 0000 | 07/10/2010 | 2359 |
07/16/2010 | 0000 | 07/16/2010 | 2359 |
07/24/2010 | 0000 | 07/27/2010 | 2359 |
08/02/2010 | 0000 | 08/02/2010 | 2359 |
08/15/2010 | 0000 | 08/31/2010 | 2359 |
09/01/2010 | 0000 | 09/09/2010 | 2359 |
Subject: | SGP/RAIN/C1 - Incorrect precipitation measurements |
DataStreams: | sgprainC1.b1 |
Description: | RAIN TBRG precipitation measurements were significantly lower than accumulated precipitation measured by co-located ORG and MET instruments. This occurred in both light and heavy rain events. The tipping bucket sensor was replaced by a weighing bucket rain gauge. |
Measurements: | sgprainC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2011 | 1837 | 09/28/2011 | 0120 |
10/07/2011 | 1427 | 10/18/2011 | 1527 |
09/01/2011 | 0012 | 09/03/2011 | 2206 |
09/17/2011 | 1238 | 09/19/2011 | 1516 |
Subject: | SGP/RAIN/C1 - Missing data |
DataStreams: | sgprainC1.b1 |
Description: | The datalogger connection was disrupted resulting in no data from the weighing bucket rain gauge. |
Measurements: | sgprainC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/21/2011 | 0000 | 11/21/2011 | 2359 |
Subject: | SGP/RAIN/C1 - Reprocess: Incorrect datestamp |
DataStreams: | sgprainC1.b1 |
Description: | The datestamp was off by one day for a several day period. Precipitation was only reported for one day during this time period. The data for 20111121 should be reported as 20111122. |
Measurements: | sgprainC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/12/2011 | 0500 | 06/15/2011 | 0000 |
Subject: | SGP/RAIN/C1 - Lightning strike |
DataStreams: | sgprainC1.b1 |
Description: | A lightning strike took out the instrument on 6/12. Error_latch and stat_latch flags were tripped and the instrument failed to record precip measured by co-located instruments (e.g. on 6/13). Voltage and other data fields were also very erratic. Instrument was replaced and normal operations resumed on 20110615. |
Measurements: | sgprainC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/30/2013 | 0000 | 06/13/2013 | 1900 |
Subject: | SGP/RAIN/C1 - Instrument problem |
DataStreams: | sgprainC1.00, sgprainC1.b1, sgprainauxC1.00, sgprainauxC1.b1 |
Description: | Instrument has shown very large jumps in accumulated rain during the past week when there is no rain or very little rain. This is coupled with a known issue that is being worked on with the filtering out of noise in the instrument. This filtering leads to removal of low rain rate data resulting low accumulation totals on days with low maximum rain rates (<60 mm/hr) |
Measurements: | sgprainC1.b1:
sgprainauxC1.00:
sgprainauxC1.b1:
sgprainC1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/2013 | 0130 | 07/03/2013 | 1510 |
Subject: | SGP/THWAPS/C1 - T/RH sensor failure |
DataStreams: | sgpthwapsC1.b1 |
Description: | Storm damage caused failure of the temperature/humidity sensor. Sensor replaced. |
Measurements: | sgpthwapsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/1999 | 1445 | 10/14/1999 | 2230 |
Subject: | SGP/THWAPS - RH values low by factor of 1.1 |
DataStreams: | sgpthwapsC1.a1, sgpthwapsC1.b1, sgpthwapsC1.00, DsgpthwapsC1.a1, DsgpthwapsC1.b1 |
Description: | Because of instrument mentor stupidity, all C1 THWAPS RH data collected from the begin date of this DQR are low by a factor of 1.1. This error also will affect the vapor pressure data. The problem is that the CSI data logger is expecting the data to be scaled 0-100%RH on the range 0-1V, but it is now being scaled 0-110%RH on the range 0-1V. |
Measurements: | sgpthwapsC1.b1:
sgpthwapsC1.a1:
sgpthwapsC1.00:
DsgpthwapsC1.a1:
DsgpthwapsC1.b1:
|