DQR ID | Subject | Data Streams Affected |
---|---|---|
D051219.1 | NSA/VCEIL/C1 - poor sensitivity at low altitude | nsavceil25kC1.b1 |
D091210.1 | NSA/VCEIL/C1 - Degraded window | nsavceil25kC1.b1 |
D100215.3 | NSA/VCEIL/C1/C2 - Reprocess: Incorrect serial_number and sum_backscatter fields | nsavceil25kC1.b1, nsavceil25kC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/27/2005 | 1930 | 05/02/2006 | 0000 |
Subject: | NSA/VCEIL/C1 - poor sensitivity at low altitude |
DataStreams: | nsavceil25kC1.b1 |
Description: | The ceilometer has an internal amplifier that it sets to low or high depending on prevalant conditions. Normally, the gain is automatically set high during clear skies. It appears that during these times, a slight degradation in either the transmitter or receiver is being amplified and produces a "ripple" at low altitudes in the backscatter plot. Cloud base height detection is not affected. |
Measurements: | nsavceil25kC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/02/2006 | 0000 | 12/03/2009 | 2107 |
Subject: | NSA/VCEIL/C1 - Degraded window |
DataStreams: | nsavceil25kC1.b1 |
Description: | The ceilometer continuously gave warnings for window contamination because the window had deteriorated. Cloud base height detection did not appear to be significantly affected. The window conditioner was replaced on 8/11/09 but this did not clear the warning. The instrument was replaced. |
Measurements: | nsavceil25kC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/22/1999 | 0000 | 10/31/2009 | 2359 |
Subject: | NSA/VCEIL/C1/C2 - Reprocess: Incorrect serial_number and sum_backscatter fields |
DataStreams: | nsavceil25kC1.b1, nsavceil25kC2.b1 |
Description: | Attribute serial_number: In most cases, this field is not representative of the actual serial number of the instrument because the value did not change when an instrument changed locations or was replaced with a spare. Variable sum_backscatter: An out-of-range condition for the "sum of detected and normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 0.0999 /srad. This DQR also applies to a1 and b1 level data collected prior to the begin date. |
Measurements: | nsavceil25kC2.b1:
nsavceil25kC1.b1:
|