DQR ID | Subject | Data Streams Affected |
---|---|---|
D130307.1 | SGP/CCN/C1 - Reprocess: Decline in CCN at low %ss values | sgpaosccn100C1.a1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/09/2012 | 0000 | 10/25/2012 | 2300 |
Subject: | SGP/CCN/C1 - Reprocess: Decline in CCN at low %ss values |
DataStreams: | sgpaosccn100C1.a1 |
Description: | The CCN number concentrations at low %ss values started to decline. DAta at low %ss values are suspect. Reeprocess to apply the instrument calibration done in November 2012 to the data Sep/12-Oct/12. Better %ss values can be obtained by applying %ss = (dT-2.63)/16.00, where dT = T3-T1. |
Measurements: | sgpaosccn100C1.a1:
|