Data Quality Reports for Session: 153129 User: dumka Completed: 09/20/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D120807.1PGH/AOS/M1 - Trailer temp too low, CCN affectedpghaosccn100M1.a1
D120914.3PGH/CPC/M1 - Global metadata errorpghaoscpcM1.a1


DQRID : D120807.1
Start DateStart TimeEnd DateEnd Time
12/11/2011180002/10/20120600
Subject:
PGH/AOS/M1 - Trailer temp too low, CCN affected
DataStreams:pghaosccn100M1.a1
Description:
The trailer temperature wasn't regulating very well. Night time temperatures fell to 10 C. 
The CCN doesn't operate well at low temperatures. As a consequence the instrument 
temperatures were below the acceptable range for the model calculation of %ss. The instrument 
still reported a %ss values, but this value is likely off by as much as 0.1%, especially 
at night when the trailer temperatures were low.
Measurements:pghaosccn100M1.a1:
  • CCN_ss_set


Back To Table of Contents

DQRID : D120914.3
Start DateStart TimeEnd DateEnd Time
06/09/2011095003/27/20122359
Subject:
PGH/CPC/M1 - Global metadata error
DataStreams:pghaoscpcM1.a1
Description:
The global metadata in the pghaoscpcM1.a1 file incorrectly specifies the CPC model number 
as "3372". The correct model number is "3010".

This is a metadata error only. The data are not affected. A low priority reprocessing task 
has been submitted.
Measurements:pghaoscpcM1.a1:
  • base_time


Back To Table of Contents



END OF DATA