DQR ID | Subject | Data Streams Affected |
---|---|---|
D120419.2 | SGP/QCRAD/E19 - Ventilator failure | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2, sgpqcrad1longE19.s1, sgpqcrad1longE19.s2 |
D120510.29 | SGP/QCRAD/E19 - Noisy Tc/Td | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2, sgpqcrad1longE19.s1, sgpqcrad1longE19.s2 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2006 | 0000 | 04/13/2006 | 1522 |
Subject: | SGP/QCRAD/E19 - Ventilator failure |
DataStreams: | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2, sgpqcrad1longE19.s1, sgpqcrad1longE19.s2 |
Description: | Dowenlling shortwave hemispheric instrument signal (Eppley PSP) is very erratic due to failed ventilator fan which was replaced during PM on 04/13/2006. Discussed in instrument DQR D060403.3. |
Measurements: | sgpqcrad1longE19.c1:
sgpqcrad1longE19.c2:
sgpqcrad1longE19.s1:
sgpqcrad1longE19.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/25/2002 | 0000 | 10/11/2002 | 1800 |
09/22/2010 | 0000 | 10/10/2010 | 0100 |
Subject: | SGP/QCRAD/E19 - Noisy Tc/Td |
DataStreams: | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2, sgpqcrad1longE19.s1, sgpqcrad1longE19.s2 |
Description: | Case and dome temperatures for longwave fluxes are sporadically noisy during these periods corresponding to additional noise in the longwave fluxes. No instrument DQRs exist documenting this problem. |
Measurements: | sgpqcrad1longE19.c1:
sgpqcrad1longE19.c2:
sgpqcrad1longE19.s1:
sgpqcrad1longE19.s2:
|