DQR ID | Subject | Data Streams Affected |
---|---|---|
D041123.1 | SGP/THWAPS/C1 - RH Sensor Calibration | sgpthwapsC1.b1 |
D050812.12 | SGP/THWAPS/C1 - Minor bias in saturation vapor pressure | sgpthwapsC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/16/2004 | 2135 | 11/29/2004 | 1930 |
Subject: | SGP/THWAPS/C1 - RH Sensor Calibration |
DataStreams: | sgpthwapsC1.b1 |
Description: | The analog output for the T/RH probe was incorrectly calibrated. This caused excessively high readings of RH. |
Measurements: | sgpthwapsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/21/1999 | 0000 | 08/10/2005 | 1445 |
Subject: | SGP/THWAPS/C1 - Minor bias in saturation vapor pressure |
DataStreams: | sgpthwapsC1.b1 |
Description: | The saturation vapor correction when temps are below 0C contained an error in the multiplier. The incorrect multiplier only equates to an error of 0.01 kPa which is less than the stated uncertainty of the measurement (+/- 0.035 kPa). |
Measurements: | sgpthwapsC1.b1:
|