DQR ID | Subject | Data Streams Affected |
---|---|---|
D030527.2 | SGP/SIRS/E27 - Instrument problem, Shadowband Misalignment | sgpsirs20sE27.a0, sgpsirsE27.00, sgpsirsE27.b1 |
D030529.3 | SGP/SIRS/E21 - Shadowband Misalignment/Tracker Failure | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
D030612.1 | SGP/SIRS/E21 - Shadowband Misalignment/Tracker Failure | sgpsirsE21.b1, sgpsirs20sE21.a0, sgpsirsE21.00 |
D030701.1 | SGP/SIRS/E10 - Noisy US(upwelling shortwave) data stream problem | sgpsirs20sE10.a0, sgpsirsE10.00, sgpsirsE10.b1 |
D030715.1 | SGP/SIRS/E3 - Degraded upwelling longwave data | sgpsirs20sE3.a0, sgpsirsE3.00, sgpsirsE3.b1 |
D030715.2 | SGP/SIRS/E5 - Downwelling direct normal (NIP) reporting low | sgpsirs20sE5.a0, sgpsirsE5.00, sgpsirsE5.b1 |
D030722.2 | SGP/SIRS/E21 - Shadowband Misalignment | sgpsirsE21.b1, sgpsirs20sE21.a0, sgpsirsE21.00 |
D030901.1 | SGP/SIRS/E19 - Shadowband Misalignment | sgpsirsE19.b1, sgpsirs20sE19.a0, sgpsirsE19.00 |
D030922.4 | SGP/SIRS/E27 - Tracker Misalignment | sgpsirsE27.b1, sgpsirs20sE27.a0, sgpsirsE27.00 |
D030922.5 | SGP/SIRS/E2 - Ventilator Fan Failure | sgpsirs20sE2.a0, sgpsirsE2.00, sgpsirsE2.b1 |
D031009.2 | SGP/SIRS/E21 - Reprocessed: Incorrect PIR/UIR Calibration Coefficient | sgpsirsE21.b1 |
D031020.1 | SGP/SIRS/E1 - failed SIRS multiplexer | sgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1 |
D031020.2 | SGP/SIRS/E9 - Possible instrument drift | sgpsirs20sE9.a0, sgpsirsE9.00, sgpsirsE9.b1 |
D031020.3 | SGP/SIRS/E18 - Instrument noise problem | sgpsirs20sE18.a0, sgpsirsE18.00, sgpsirsE18.b1 |
D031020.4 | SGP/SIRS/E21 - Bird Poop Contamination | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
D031027.1 | SGP/SIRS/E10 - Noisy UIR (upwelling longwave)data stream problem | sgpsirs20sE10.a0, sgpsirsE10.00, sgpsirsE10.b1 |
D031118.1 | SGP/SIRS/E1 - Unknown - Downwelling shaded longwave measurement incorrect | sgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1 |
D031124.8 | SGP/SIRS/E7 - Unknown problem with downwelling shortwave instrument (Eppley PSP) | sgpsirs20sE7.a0, sgpsirsE7.00, sgpsirsE7.b1 |
D031208.6 | SGP/SIRS/E4 - Shadowband Misalignment | sgpsirs20sE4.a0, sgpsirsE4.00, sgpsirsE4.b1 |
D040126.3 | SGP/SIRS/E21 - False signal | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
D040525.12 | SGP/SIRS/E19 - Metadata errors | sgpsirsE19.b1 |
D040712.2 | SGP/SIRS/E24 - NIP shaded near sunrise | sgpsirs20sE24.a0, sgpsirs60sE24.a1, sgpsirsE24.a2, sgpsirsE24.b1 |
D040712.3 | SGP/SIRS/E24 - NIP shaded near sunrise | sgpsirs20sE24.a0, sgpsirsE24.b1 |
D060630.17 | SGP/SIRS/E9 - Reprocessed: Longwave Calibration error | sgpsirsE9.b1 |
D060630.20 | SGP/SIRS/E12 - Reprocessed: Longwave Calibration error | sgpsirsE12.b1 |
D060630.21 | SGP/SIRS/E13 - Reprocessed: Longwave Calibration error | sgpsirsE13.b1 |
D060630.22 | SGP/SIRS/E15 - Reprocessed: Longwave Calibration error | sgpsirsE15.b1 |
D060630.23 | SGP/SIRS/E16 - Reprocessed: Longwave Calibration error | sgpsirsE16.b1 |
D060630.24 | SGP/SIRS/E18 - Reprocessed: Longwave Calibration error | sgpsirsE18.b1 |
D060630.25 | SGP/SIRS/E19 - Reprocessed: Longwave Calibration error | sgpsirsE19.b1 |
D060630.26 | SGP/SIRS/E20 - Reprocessed: Longwave Calibration error | sgpsirsE20.b1 |
D060630.27 | SGP/SIRS/E21 - Reprocessed: Longwave Calibration error | sgpsirsE21.b1 |
D060630.28 | SGP/SIRS/E22 - Reprocessed: Longwave Calibration error | sgpsirsE22.b1 |
D060630.29 | SGP/SIRS/E24 - Reprocessed: Longwave Calibration error | sgpsirsE24.b1 |
D060630.30 | SGP/SIRS/E27 - Reprocessed: Longwave Calibration error | sgpsirsE27.b1 |
D060630.8 | SGP/SIRS/C1 - Reprocessed: Longwave Calibration error | sgpsirsC1.b1 |
D080610.1 | SGP/SIRS/E20 - Missing data | sgpsirs20sE20.a0, sgpsirsE20.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/16/2003 | 0000 | 06/18/2003 | 1845 |
Subject: | SGP/SIRS/E27 - Instrument problem, Shadowband Misalignment |
DataStreams: | sgpsirs20sE27.a0, sgpsirsE27.00, sgpsirsE27.b1 |
Description: | Downwelling and Upwelling longwave values are somewhat noisy due to noisy PIR (pygeometer) case and dome thermistor measurements on both the upwelling and downwelling PIRs. This problem dates back to the beginning of the SIRS E27 data stream. |
Measurements: | sgpsirsE27.b1:
sgpsirsE27.00:
sgpsirs20sE27.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/2003 | 1000 | 05/28/2003 | 1430 |
Subject: | SGP/SIRS/E21 - Shadowband Misalignment/Tracker Failure |
DataStreams: | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
Description: | The solar tracker failed beginning around sunrise on 05/19/03 most likely during the previous nights reset. Failure time was approximate time of sunrise at 1000 on 05/19/03. The tracker was power cycled and RTS at 1415 GMT on 05/28/03. |
Measurements: | sgpsirs20sE21.a0:
sgpsirsE21.00:
sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/03/2003 | 1000 | 06/16/2003 | 1835 |
Subject: | SGP/SIRS/E21 - Shadowband Misalignment/Tracker Failure |
DataStreams: | sgpsirsE21.b1, sgpsirs20sE21.a0, sgpsirsE21.00 |
Description: | Suntracker appears to have failed beginning on 6/3/03, most likely during previous night reset. Start time listed is approximate time of sunrise. |
Measurements: | sgpsirs20sE21.a0:
sgpsirsE21.00:
sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/22/2003 | 0146 | 07/02/2003 | 1400 |
Subject: | SGP/SIRS/E10 - Noisy US(upwelling shortwave) data stream problem |
DataStreams: | sgpsirs20sE10.a0, sgpsirsE10.00, sgpsirsE10.b1 |
Description: | Beginning 6/22, upwelling shortwave values frequently dropped well below 0 W/m^2 (occasionally dropping below -50 W/m^2) during the nighttime hours. Daytime values were excessively high and noisy. Instrument wiring was suspected and was corrected 07/02/2003. |
Measurements: | sgpsirsE10.b1:
sgpsirsE10.00:
sgpsirs20sE10.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2003 | 0400 | 07/30/2003 | 1600 |
Subject: | SGP/SIRS/E3 - Degraded upwelling longwave data |
DataStreams: | sgpsirs20sE3.a0, sgpsirsE3.00, sgpsirsE3.b1 |
Description: | Upwelling longwave values had been showing spikes in the late evening beginning around 7/5. On 7/16/03 technicians found a rodent nest and cables chewed at pivot point of upwelling instrument tower. Only the upwelling longwave (UIR) instrument case temperature was affected but this degrades all of the UIR longwave irradiance measurement. Both UIR and US cables replaced on 7/30/03 1600 GMT which fixed the problem. |
Measurements: | sgpsirs20sE3.a0:
sgpsirsE3.00:
sgpsirsE3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/2003 | 1630 | 07/16/2003 | 1840 |
Subject: | SGP/SIRS/E5 - Downwelling direct normal (NIP) reporting low |
DataStreams: | sgpsirs20sE5.a0, sgpsirsE5.00, sgpsirsE5.b1 |
Description: | There were large differences between the measured downwelling shortwave (DS) and the downwelling shortwave calculated from direct (NIP) and diffuse (DD) readings. It appears that the direct normal shortwave readings were much too low, particularly on 7/14 and 7/15. It appears that the diffuse instrument was being shaded properly, but the NIP was not tracking the sun. The problem was corrected during the preventative maintenance visit on 7/16/03 at 1840GMT. The technician reported doing the quarterly connection corrosion checks and ventilator fan cleaning, the NIP was described as having "dry-light dust" on the window, no mention of tracker alignment was made. Although uncertain, a bad connection at some point in the instrument wiring was likely to blame and was eliminated during the corrosion inspection process. |
Measurements: | sgpsirs20sE5.a0:
sgpsirsE5.b1:
sgpsirsE5.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/17/2003 | 1000 | 08/06/2003 | 1350 |
Subject: | SGP/SIRS/E21 - Shadowband Misalignment |
DataStreams: | sgpsirsE21.b1, sgpsirs20sE21.a0, sgpsirsE21.00 |
Description: | Solar tracker stopped functioning on 7/17. Tracker power cycled on 07/24 and returned to normal tracking but failed again 7/25. Power was cycled again 8/6 and tracker performance was sustained thereafter. |
Measurements: | sgpsirs20sE21.a0:
sgpsirsE21.00:
sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/28/2003 | 1200 | 09/04/2003 | 1526 |
Subject: | SGP/SIRS/E19 - Shadowband Misalignment |
DataStreams: | sgpsirsE19.b1, sgpsirs20sE19.a0, sgpsirsE19.00 |
Description: | The solar tracker appears to have failed on 8/28. Fuse replaced on 09/04/03 and tracker returned to service at 15:26GMT |
Measurements: | sgpsirsE19.00:
sgpsirs20sE19.a0:
sgpsirsE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2003 | 1200 | 10/15/2003 | 1730 |
Subject: | SGP/SIRS/E27 - Tracker Misalignment |
DataStreams: | sgpsirsE27.b1, sgpsirs20sE27.a0, sgpsirsE27.00 |
Description: | Suntracker failed at this facility. The suntracker was reset on 9/17, but this only temporarily fixed the problem and the tracker began failing again 9/18. Tracker was reset and realigned on 10/15/03 17:30GMT and has been operating properly since. |
Measurements: | sgpsirsE27.b1:
sgpsirsE27.00:
sgpsirs20sE27.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2003 | 1000 | 09/25/2003 | 1543 |
Subject: | SGP/SIRS/E2 - Ventilator Fan Failure |
DataStreams: | sgpsirs20sE2.a0, sgpsirsE2.00, sgpsirsE2.b1 |
Description: | Measured downwelling shortwave data was very questionable, reporting nighttime values as low as -50 W/m^2, while differing from calculated downwelling shortwave by as much as 50 W/m^2 lower during the day. The first indication of a problem was around 10 UTC, 9/14, when measured DS values suddenly dropped below 10 W/m^2 for about a 2 hour period before sunrise, then were consistently lower than calculated DS by as much as 40 W/m^2 during the day. The problem got progressively worse with time until the ventilator fan was replace on 9/25. Ventilator fan failure changed the fundamental operating characteristics of this PSP and resulted in unpredictable biased measurements which cannot be corrected without additional empirical characterization of the radiometer in question. |
Measurements: | sgpsirsE2.b1:
sgpsirs20sE2.a0:
sgpsirsE2.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1330 | 08/08/2003 | 1650 |
Subject: | SGP/SIRS/E21 - Reprocessed: Incorrect PIR/UIR Calibration Coefficient |
DataStreams: | sgpsirsE21.b1 |
Description: | Note: This reprocessing was completed 20061016. Incorrect coefficients were applied to the upwelling longwave radiation measurement instrument (i.e. pyrgeometer, SIRS acroynm "UIR") during this time period. For a PIR mounted in an upwelling measurement orientation (UIR), the thermopile output is extremely low compared to a downwelling PIR (DIR). In addition, the UIR case and dome temperature differences are much smaller than a DIR. This results in the pyrgeometer receiver radiation term, Wr (NREL/Hickey equation) or Wc (Albrecht/Cox equation), dominating all other terms of the infrared radiant flux equation. Since the receiver radiation term uses coefficients in both forms of the pyrgeometer equations nearly equal to 1.0 +- <1% the result is a measurement of incoming longwave radiant flux relatively insensitive to the installed instrument - so long as it is an Eppley PIR. Evaluation of the problem data shows the expected error of the data (before correction) will be within 1% - less than the uncertainty of the instrument (Eppley PIR greater of 5% or 5 W/m2) To correct the one minute data found in the a1 or b1 datastream follow this procedure: 1) Obtain the following variables from the b1 datafile: inst_up_long_dome_temp inst_up_long_case_temp up_long_netir 2) For each one minute data entry in b1 tpUIR[uV] = up_long_netir[uV]/0.2523876[W/m2/uV] 3) Apply for each one minute data entry: WinUIR(W/m2)=k*V+ Sigma*Tc^4 -4*Sigma*(Td^4 - Tc^4) =[0.28902*tpUIR]+[Sigma*inst_up_long_case_temp^4] -[4*Sigma*(inst_up_long_dome_temp^4-inst_up_long_case^4)] where, Sigma = 5.6697E-8 (Stefan-Boltzman Const) Other datastream corrections that use the UIR measurements will require similar corrections. See the following web site for additional information on the pyrgeometer equations, specifically sections 2.1(Single Calibration Factor Approach) and 2.2 (Four Coefficient Calibration Approach) http://www.arm.gov/docs/instruments/static/pyrg.html |
Measurements: | sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2003 | 2023 | 10/21/2003 | 1930 |
Subject: | SGP/SIRS/E1 - failed SIRS multiplexer |
DataStreams: | sgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1 |
Description: | Beginning 10/14, a failed multiplexer resulted in missing downwelling shortwave data and downwelling and upwelling longwave data which were much too low. The multiplexer handles data for the longwave instruments (UIR and DIR) case and dome temperatures as well as the downwelling shortwave hemispheric instrument (DS). |
Measurements: | sgpsirsE1.00:
sgpsirs20sE1.a0:
sgpsirsE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2003 | 0000 | 10/21/2003 | 1600 |
Subject: | SGP/SIRS/E9 - Possible instrument drift |
DataStreams: | sgpsirs20sE9.a0, sgpsirsE9.00, sgpsirsE9.b1 |
Description: | Downwelling shortwave values were too low, dropping below -10 W/m^2 at night and reading 30-50 W/m^2 lower than DS calculated from direct and diffuse measurements during the day. This problem became more noticeable beginning on 10/14, when nighttime values began to drop below -10 W/m^2 consistently. The problem was corrected with the annual instrument changout on 10/23/03. Although the downwelling shortwave instrument (DS) appears to be the problem a more likely scenario is that the cable connections became thermopile. The instrument changeout possibly reseated and corrected any poor connections. |
Measurements: | sgpsirs20sE9.a0:
sgpsirsE9.b1:
sgpsirsE9.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2003 | 1000 | 10/20/2003 | 0000 |
Subject: | SGP/SIRS/E18 - Instrument noise problem |
DataStreams: | sgpsirs20sE18.a0, sgpsirsE18.00, sgpsirsE18.b1 |
Description: | Upwelling longwave data was somewhat noisy during this period. There was a direct relation to the case and dome thermistor measurements for the upwelling longwave PIR (UIR). Data quality does not appear to be greatly affected overall, though instantaneous values had greater than usual variability. The noise disappeared on 10/19. |
Measurements: | sgpsirs20sE18.a0:
sgpsirsE18.b1:
sgpsirsE18.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/18/2003 | 1600 | 10/29/2003 | 1630 |
Subject: | SGP/SIRS/E21 - Bird Poop Contamination |
DataStreams: | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
Description: | Downwelling shortwave values were erratic during the day on 10/18 and in very poor agreement at times with downwelling shortwave values calculated from direct and diffuse measurements. Bi-weekly preventative maintenace records indicate bird poop was cleaned from the downwelling shortwave instrument on 10/29 at 16:30. Review of maintenance records and instrument data indicate no bird contamination since 10/29. Turkey vultures are the usual suspects in data quality problems at this facility. The RBD(rotating bird deterrent) had failed the week of 10/13 and is awaiting parts. However, beginning in late October the birds had left the site. |
Measurements: | sgpsirs20sE21.a0:
sgpsirsE21.00:
sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/29/2003 | 1500 | 08/06/2003 | 0500 |
Subject: | SGP/SIRS/E10 - Noisy UIR (upwelling longwave)data stream problem |
DataStreams: | sgpsirs20sE10.a0, sgpsirsE10.00, sgpsirsE10.b1 |
Description: | Beginning 6/29, upwelling longwave (UIR) values frequently exceeded quality control boundaries and appeared excessively noisy. Instrument wiring was suspected. Data remained noisy until 08/06/2003 0500GMT when the UIR data returned to normal. |
Measurements: | sgpsirsE10.b1:
sgpsirsE10.00:
sgpsirs20sE10.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/11/2003 | 0438 | 11/18/2003 | 2200 |
Subject: | SGP/SIRS/E1 - Unknown - Downwelling shaded longwave measurement incorrect |
DataStreams: | sgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1 |
Description: | Downwelling longwave values were very noisy and out of range due to a missing PIR case temperature measurement. Faulty wiring to the case temperature thermistor on the downwelling longwave IR instrument (PIR) was the likely cause of the problem. On 11/18, the E1 tracker and associated instruments (direct shortwave -NIP, hemispheric shaded shortwave -8-48 and hemispheric shaded longwave - problem PIR) were moved to a new tracker stand. Rewiring the instrument corrected the problem. |
Measurements: | sgpsirsE1.00:
sgpsirs20sE1.a0:
sgpsirsE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/21/2003 | 0317 | 12/03/2003 | 0000 |
Subject: | SGP/SIRS/E7 - Unknown problem with downwelling shortwave instrument (Eppley PSP) |
DataStreams: | sgpsirs20sE7.a0, sgpsirsE7.00, sgpsirsE7.b1 |
Description: | Beginning around 0317 UTC, 11/21, downwelling shortwave values became anamolously high during the night, when they should have been around zero. Values returned to normal after SGP technicians reseated this instruments connections at 1215UTC 12/02/2003. The following day's data appeared normal. Appears that a poor connection was the culprit this time. |
Measurements: | sgpsirs20sE7.a0:
sgpsirsE7.b1:
sgpsirsE7.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/04/2003 | 1700 | 12/08/2003 | 1200 |
Subject: | SGP/SIRS/E4 - Shadowband Misalignment |
DataStreams: | sgpsirs20sE4.a0, sgpsirsE4.00, sgpsirsE4.b1 |
Description: | The solar tracker failed on 12/4, resulting in bad direct and diffuse measurements. The tracker started working again on 12/8 without Ops intervention. Technician maintenance checks on 12/3 and 12/17 reported correct tracker operation. |
Measurements: | sgpsirsE4.00:
sgpsirs20sE4.a0:
sgpsirsE4.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2003 | 0000 | 02/26/2004 | 2359 |
Subject: | SGP/SIRS/E21 - False signal |
DataStreams: | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
Description: | During clear-sky conditions, downwelling shortwave values spike downward approximately 50 to 200 W/m^2 around 22:15 UTC for about 15 minutes. This problem occurs at about the same time for the same duration each day during the indicated period of the year. This effect is due to a shadow from the Rotating Bird Deterrent (RBD) installed earlier in the year to prevent extensive bird droppings at the site. The RBD is installed at such a location that it will partially shade the downwelling shortwave PSP from 19 October to 23 February of the year. |
Measurements: | sgpsirs20sE21.a0:
sgpsirsE21.00:
sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2002 | 0000 | 05/06/2003 | 2359 |
Subject: | SGP/SIRS/E19 - Metadata errors |
DataStreams: | sgpsirsE19.b1 |
Description: | The latitude, longitude and altitude of the El Reno (E19) site were incorrectly entered into the ARM database. The correct location of the SIRS.E19 instrument is: Lat: 35.557N Lon: 98.017W Alt: 421m |
Measurements: | sgpsirsE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/08/2002 | 1200 | 05/20/2002 | 1300 |
07/26/2002 | 1300 | 09/06/2002 | 1200 |
04/08/2003 | 1200 | 04/16/2003 | 1300 |
Subject: | SGP/SIRS/E24 - NIP shaded near sunrise |
DataStreams: | sgpsirs20sE24.a0, sgpsirs60sE24.a1, sgpsirsE24.a2, sgpsirsE24.b1 |
Description: | The NIP is shaded by the EF laptop computer enclosure near sunrise during parts of the Spring and Summer. This has occurred from 04/08 to 05/20 and from 07/26 to 09/06 every year since 2001 and will occur during these date ranges in the future as well. The shading occurs for a few minutes each day between 1200 and 1300 GMT depending on the time of the year. For the early dates, shading occurs a little after 1200 GMT and by 05/20 shading happens nearer to 1300 GMT. In late July the shading occurs near 1300 GMT and closer to 1200 GMT by 09/06. |
Measurements: | sgpsirsE24.a2:
sgpsirs60sE24.a1:
sgpsirsE24.b1:
sgpsirs20sE24.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/2003 | 1200 | 05/20/2003 | 1300 |
07/26/2003 | 1300 | 09/06/2003 | 1200 |
04/08/2004 | 1200 | 05/20/2004 | 1300 |
07/26/2004 | 1300 | 09/06/2004 | 1200 |
Subject: | SGP/SIRS/E24 - NIP shaded near sunrise |
DataStreams: | sgpsirs20sE24.a0, sgpsirsE24.b1 |
Description: | The NIP is shaded by the EF laptop computer enclosure near sunrise during parts of the Spring and Summer. This has occurred from 04/08 to 05/20 and from 07/26 to 09/06 every year since 2001 and will occur during these date ranges in the future as well. The shading occurs for a few minutes each day between 1200 and 1300 GMT depending on the time of the year. For the early dates, shading occurs a little after 1200 GMT and by 05/20 shading happens nearer to 1300 GMT. In late July the shading occurs near 1300 GMT and closer to 1200 GMT by 09/06. |
Measurements: | sgpsirsE24.b1:
sgpsirs20sE24.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/2003 | 1640 | 02/07/2006 | 1730 |
Subject: | SGP/SIRS/E9 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE9.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instanteous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2003 | 1640 | 02/14/2006 | 1710 |
Subject: | SGP/SIRS/E12 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE12.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2003 | 1935 | 02/16/2006 | 1905 |
Subject: | SGP/SIRS/E13 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE13.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2003 | 1650 | 02/14/2006 | 1806 |
Subject: | SGP/SIRS/E15 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE15.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1605 | 02/15/2006 | 1533 |
Subject: | SGP/SIRS/E16 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE16.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2003 | 2200 | 02/14/2006 | 2100 |
Subject: | SGP/SIRS/E18 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE18.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2003 | 1600 | 02/16/2006 | 1558 |
Subject: | SGP/SIRS/E19 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE19.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 2100 | 02/15/2006 | 2030 |
Subject: | SGP/SIRS/E20 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE20.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in November 2006. |
Measurements: | sgpsirsE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1520 | 02/15/2006 | 1515 |
Subject: | SGP/SIRS/E21 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE21.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in November 2006. |
Measurements: | sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1830 | 02/15/2006 | 1741 |
Subject: | SGP/SIRS/E22 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE22.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in August 2006. |
Measurements: | sgpsirsE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2003 | 1640 | 02/16/2006 | 1705 |
Subject: | SGP/SIRS/E24 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE24.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE24.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/16/2003 | 0002 | 02/15/2006 | 1750 |
Subject: | SGP/SIRS/E27 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE27.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in May 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2003 | 1950 | 02/16/2006 | 1901 |
Subject: | SGP/SIRS/C1 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/24/2003 | 1302 | 01/28/2003 | 1949 |
01/30/2003 | 0226 | 02/05/2003 | 1951 |
07/11/2003 | 0002 | 07/15/2003 | 2001 |
05/18/2005 | 2132 | 05/26/2005 | 1827 |
05/27/2005 | 0000 | 05/31/2005 | 0646 |
12/10/2007 | 1803 | 12/13/2007 | 1850 |
04/17/2008 | 2246 | 04/23/2008 | 1440 |
05/04/2008 | 1823 | 05/17/2008 | 1447 |
Subject: | SGP/SIRS/E20 - Missing data |
DataStreams: | sgpsirs20sE20.a0, sgpsirsE20.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgpsirsE20.b1:
sgpsirs20sE20.a0:
|