DQR ID | Subject | Data Streams Affected |
---|---|---|
D130715.2 | PVC/ECOR/M1 - Vapor Pressure High | pvc30ecorM1.b1 |
D130718.5 | PVC/MET/M1 - Questionable humidity | pvcmetM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/01/2013 | 0000 | 06/29/2013 | 0930 |
Subject: | PVC/ECOR/M1 - Vapor Pressure High |
DataStreams: | pvc30ecorM1.b1 |
Description: | The vapor pressure data from the ECOR was somewhat high. This is obvious as the MET vapor pressure was determined to be too high and the ECOR vapor pressure was just slightly higher than the MET measurement. This probably resulted in slightly too large LE as measured by the ECOR, by about 10%. |
Measurements: | pvc30ecorM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/26/2013 | 1800 | 06/29/2013 | 1000 |
Subject: | PVC/MET/M1 - Questionable humidity |
DataStreams: | pvcmetM1.b1 |
Description: | A reference check of the humidity sensor showed that the instrument may have been reading as much as 20% high. |
Measurements: | pvcmetM1.b1:
|