DQR ID | Subject | Data Streams Affected |
---|---|---|
D120420.3 | SGP/QCRAD/C1 - Reprocess: SW data require -1 multipler | sgpqcrad1longC1.s2 |
D120420.4 | SGP/QCRAD/C1 - Upwelling SW cable problem | sgpqcrad1longC1.s2 |
D120420.5 | SGP/QCRAD/C1 - Sensor cable problem | sgpqcrad1longC1.s2 |
D120420.6 | SGP/QCRAD/C1 - SW direct normal data drop out | sgpqcrad1longC1.s2 |
D120420.7 | SGP/QCRAD/C1 - Failed upwelling SW cable | sgpqcrad1longC1.s2 |
D120420.8 | SGP/QCRAD/C1 - NIP failure | sgpqcrad1longC1.s2 |
D120510.5 | SGP/QCRAD/C1 - Case and dome temperatures not available | sgpqcrad1longC1.s2 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/24/2000 | 1900 | 05/08/2000 | 2359 |
Subject: | SGP/QCRAD/C1 - Reprocess: SW data require -1 multipler |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | Magnitude of data are correct, however, 2 wires were crossed producing negative values for SW Global Down, SW Global Up, SW Direct Normal, and SW Diffuse Down. Problem has been fixed at site. Users of this data should multiply by -1 to get the correct values for SW Global Down, SW Global Up, SW Direct Normal, and SW Diffuse Down. Data will be reprocessed. Documented in instrument DQR D000919.1. |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2006 | 0900 | 01/23/2007 | 1930 |
Subject: | SGP/QCRAD/C1 - Upwelling SW cable problem |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | Starting 11/13/2006, the C1 SIRS Upwelling Shortwave (up_short_hemisp) started to show large differences in comparison to co-located E13 SIRS and C1 IRT25m. After much troubleshooting and gnashing of teeth it was discovered that rodents had chewed through the instrument cables at the bottom of the tilt tower. Cables were replaced 01/23/2007 1930GMT. Documented in instrument DQR D061121.2 |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/03/2007 | 1400 | 04/22/2008 | 2052 |
Subject: | SGP/QCRAD/C1 - Sensor cable problem |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | Beginning October 3, 2007, the upwelling longwave irradiance (up_long_hemisp, aka "UIR") started to spike to unreasonable values positively offset from the co-located E13 by over 100W/m2. On January 17 2008 the UIR values become consistently bad due to bad UIR dome thermistor wiring. Problem appears as a resistive offset in the dome thermistor but due to nature of failure (unknown cable problem) data correction would not yield any better data than the co-located E13 UIR instrument that is known to be operating correctly. Instrument cable replacement on 4/22 fixed the problem. Thermopile data (netIR) from this instrument is good. However, incoming irradiance data is not correctable due to bad UIR dome thermistor wiring within the instrument cable. Documented in instrument DQR D071022.4. |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/18/2009 | 1751 | 07/31/2009 | 1645 |
Subject: | SGP/QCRAD/C1 - SW direct normal data drop out |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | NIP cable had failed during this time period. Short_direct_normal data are not recoverable during this time period. However, it can be reconstructed using the following formula for each one minute data: short_direct_normal = (down_short_hemisp-down_short_diffuse_hemisp)/cos(zenith angle) Documented in instrument DQR D090623.1 |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/20/2010 | 0414 | 06/30/2010 | 1442 |
Subject: | SGP/QCRAD/C1 - Failed upwelling SW cable |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | The upwelling shortwave hemispheric (up_short_hemisp) data were missing or incorrect due to a bad cable. Cable was replaced on 6/30 and the problem was resolved. Documented in instrument DQR D100701.1 |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/12/2011 | 0000 | 06/24/2011 | 1740 |
Subject: | SGP/QCRAD/C1 - NIP failure |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | The short_direct_normal (NIP instrument) cable failed during this time period. Data are not recoverable. Users can derive an estimate of direct normal values using the following formula: short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle) Documented in instrument DQR D110628.2 |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/1999 | 0000 | 12/31/2000 | 2359 |
Subject: | SGP/QCRAD/C1 - Case and dome temperatures not available |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | Longwave downwelling and upwelling case and dome temperatures are not available during this time period because they were not kept in the sirs.a1 files and sirs.b1 files are not available. This means that some longwave tests and corrections can't be applied. |
Measurements: | sgpqcrad1longC1.s2:
|