DQR ID | Subject | Data Streams Affected |
---|---|---|
D150123.9 | OLI/AMC/M1 - Data missing from loss of power | oliamcM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/06/2015 | 1430 | 01/08/2015 | 2000 |
Subject: | OLI/AMC/M1 - Data missing from loss of power |
DataStreams: | oliamcM1.b1 |
Description: | Data is missing due to logger battery voltage dropping below operational range of about 9 VDC, and the logger shutting down automatically in order to preserve battery life. Starting with 12/31/2014 04:00:00, AMF3 facility lost power due to failure of microturbines. The system was operating normally until the logger shut down at this time. |
Measurements: | oliamcM1.b1:
|