DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160512.2 | SGP/EBBR/E13 - Sensible and Latent Heat Flux Incorrect | sgp30ebbrE13.b1 | Suspect |
D160715.8 | SGP/EBBR/E13 - Sensible and Latent Heat Fluxes Incorrect | sgp30ebbrE13.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/08/2016 | 1030 | 05/05/2016 | 2000 | Suspect |
Subject: | SGP/EBBR/E13 - Sensible and Latent Heat Flux Incorrect |
DataStreams: | sgp30ebbrE13.b1 |
Description: | Sensible and Latent Heat Flux were sometimes incorrect when Soil Heat Flow 5 was incorrect. The EBBR unit was replaced not only to resolve problems with the unit, but to provide a newly calibrated unit for an IOP. |
Measurements: | sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/02/2016 | 1030 | 07/12/2016 | 1430 | Incorrect |
Subject: | SGP/EBBR/E13 - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp30ebbrE13.b1 |
Description: | Sensible and latent fluxes were incorrect when soil heat flow 3 was incorrect. Soil heat flow probe 3 was damaged, apparently by rodents. |
Measurements: | sgp30ebbrE13.b1:
|