DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160107.2 | AWR/MET/M1 - PWD Sensor Not Installed | awrmetM1.b1 | Missing |
D160107.3 | AWR/MET/M1 - Temperature, RH, Vapor Pressure Incorrect | awrmetM1.b1 | Incorrect |
D160107.4 | AWR/MET/M1 - PWD Incorrect | awrmetM1.b1 | Incorrect |
D160107.7 | AWR/SONDE/M1 - Suspect Surface Temperature or RH | awrsondewnpnM1.b1 | Suspect |
D160322.1 | AWR/RWP/M1 - Reprocess: Missing high power mode | awr1290bsrwpwindavgM1.b1 | Does not affect quality |
D160322.3 | AWR/TSI/M1 - Reprocess: Sky filter thresholds | awrtsiskycoverM1.b1, awrtsicldmaskM1.a1 | Incorrect |
D160407.5 | AWR/MWACR/M1 - Transmitter Down | awrmwacrspeccmaskcopolM1.a0 | Incorrect |
D160512.1 | AWR/MET/M1 - Missing or suspect PWD data due to heavy snow | awrmetM1.b1 | Suspect |
D160520.5 | AWR/MET/M1 - Probable instrument icing | awrmetM1.b1 | Incorrect |
D160608.2 | AWR/HSRL/M1 - Instrument malfunction | awrhsrlM1.01 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/17/2015 | 0100 | 11/18/2015 | 0337 | Missing |
Subject: | AWR/MET/M1 - PWD Sensor Not Installed |
DataStreams: | awrmetM1.b1 |
Description: | The PWD was not connected during this period. |
Measurements: | awrmetM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/17/2015 | 0100 | 11/18/2015 | 0350 | Incorrect |
Subject: | AWR/MET/M1 - Temperature, RH, Vapor Pressure Incorrect |
DataStreams: | awrmetM1.b1 |
Description: | Temperature and vapor pressure were missing and the RH was flatlined below 0. |
Measurements: | awrmetM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/18/2015 | 0338 | 11/24/2015 | 1959 | Incorrect |
Subject: | AWR/MET/M1 - PWD Incorrect |
DataStreams: | awrmetM1.b1 |
Description: | The PWD was reporting errors during the setup period and was reporting large amounts of precipitation. |
Measurements: | awrmetM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/15/2015 | 2225 | 12/15/2015 | 2341 | Suspect |
12/22/2015 | 1020 | 12/22/2015 | 1139 | Suspect |
12/22/2015 | 2238 | 12/22/2015 | 2356 | Suspect |
12/13/2015 | 2224 | 12/13/2015 | 2337 | Suspect |
12/03/2015 | 2232 | 12/03/2015 | 2343 | Suspect |
12/30/2015 | 1036 | 12/30/2015 | 1152 | Suspect |
12/27/2015 | 1027 | 12/27/2015 | 1142 | Suspect |
12/26/2015 | 1018 | 12/26/2015 | 1133 | Suspect |
12/10/2015 | 1027 | 12/10/2015 | 1140 | Suspect |
12/09/2015 | 1031 | 12/09/2015 | 1145 | Suspect |
01/13/2016 | 2233 | 01/13/2016 | 2348 | Suspect |
12/14/2015 | 2231 | 12/14/2015 | 2347 | Suspect |
12/06/2015 | 1031 | 12/06/2015 | 1146 | Suspect |
12/04/2015 | 1040 | 12/04/2015 | 1205 | Suspect |
12/11/2015 | 1025 | 12/11/2015 | 1142 | Suspect |
12/02/2015 | 2303 | 12/03/2015 | 0013 | Suspect |
12/26/2015 | 2231 | 12/26/2015 | 2346 | Suspect |
Subject: | AWR/SONDE/M1 - Suspect Surface Temperature or RH |
DataStreams: | awrsondewnpnM1.b1 |
Description: | For the times listed, surface values of either Temperature or Relative Humidity appear to be suspect. |
Measurements: | awrsondewnpnM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/15/2016 | 0300 | 01/15/2016 | 0359 | Does not affect quality |
Subject: | AWR/RWP/M1 - Reprocess: Missing high power mode |
DataStreams: | awr1290bsrwpwindavgM1.b1 |
Description: | The first hourly file (which is at 03:00 UTC) for some reason has data only for low power mode. Ingest seems to have processed this mode, and not encountering the high power mode, hasn't handled the subsequent hours of data. |
Measurements: | awr1290bsrwpwindavgM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/01/2015 | 0000 | 03/18/2016 | 2219 | Incorrect |
Subject: | AWR/TSI/M1 - Reprocess: Sky filter thresholds |
DataStreams: | awrtsiskycoverM1.b1, awrtsicldmaskM1.a1 |
Description: | The sky filter thresholds incorrect. |
Measurements: | awrtsicldmaskM1.a1:
awrtsiskycoverM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/19/2016 | 0149 | 03/19/2016 | 2359 | Incorrect |
01/28/2016 | 1658 | 01/28/2016 | 1807 | Incorrect |
02/17/2016 | 1946 | 02/29/2016 | 1759 | Incorrect |
02/02/2016 | 1521 | 02/02/2016 | 2016 | Incorrect |
01/29/2016 | 1729 | 01/30/2016 | 0633 | Incorrect |
Subject: | AWR/MWACR/M1 - Transmitter Down |
DataStreams: | awrmwacrspeccmaskcopolM1.a0 |
Description: | The MWACR was faulting during these periods causing the transmitter to turn off. Problem was a faulty HV modulator. |
Measurements: | awrmwacrspeccmaskcopolM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/06/2016 | 1104 | 05/06/2016 | 2338 | Suspect |
Subject: | AWR/MET/M1 - Missing or suspect PWD data due to heavy snow |
DataStreams: | awrmetM1.b1 |
Description: | Heavy snow interfered with the instrument lenses causing backscatter and missing data. Data from the present weather detector during this time is suspect or missing. |
Measurements: | awrmetM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/10/2016 | 1905 | 05/16/2016 | 0925 | Incorrect |
Subject: | AWR/MET/M1 - Probable instrument icing |
DataStreams: | awrmetM1.b1 |
Description: | The temperature and relative humidity sensor behaved suspiciously during this time. Temperature and relative humidity readings did not trend inversely, as expected. Ice buildup on the instrument is the probable cause. |
Measurements: | awrmetM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/04/2016 | 0923 | 06/06/2016 | 1100 | Incorrect |
Subject: | AWR/HSRL/M1 - Instrument malfunction |
DataStreams: | awrhsrlM1.01 |
Description: | All data collected during this period is bad. Problems with the output shutter/DAQ/Seed voltage were reported by the AMF2 operator. He restarted the computer on June 4th. The system lock point was lost at that time. The lock point was manually reset from ~-9V to 2.8V on Monday June 6 around 11 UTC. |
Measurements: | awrhsrlM1.01:
|