Data Quality Reports for Session: 179151 User: dturner Completed: 08/22/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D021206.1TWP/SONDE/C2 - Broken Temperature SensortwpsondewnpnC2.b1Incorrect
D030419.1TWP/SONDE/C2 - Bad RH Data (AIRS Validation Sonde)twpsondewnpnC2.b1Incorrect
D030608.4TWP/SONDE/C2 - Failure of RH sensortwpsondewnpnC2.b1Suspect
D031008.1TWP/SONDE/C2 - Strange problem with data -twpsondewnpnC2.b1Incorrect
D031231.1TWP/SONDE/C2 - RH sensor badtwpsondewnpnC2.b1Suspect
D040127.3TWP/SONDE/C2 - Broken temperature sensorstwpsondewnpnC2.b1Incorrect
D040208.2TWP/SONDE/C2 - Questionable rH datatwpsondewnpnC2.b1Suspect
D040219.4TWP/SONDE/C2 - Bad RH sensortwpsondewnpnC2.b1Suspect
D040425.1TWP/SONDE/C2 - Bad RH values alofttwpsondewnpnC2.b1Incorrect
D040711.2TWP/SONDE/C2 - RH sensor failure near surfacetwpsondewnpnC2.b1Suspect
D040815.1TWP/SONDE/C2 - Operator used wrong calibration tapetwpsondewnpnC2.b1Incorrect
D040921.1TWP/SONDE/C2 - RH sensor failure below 9kmtwpsondewnpnC2.b1Suspect
D040921.2TWP/SONDE/C2 - Bad RH values alofttwpsondewnpnC2.b1Incorrect
D050913.3TWP/SONDE/C2 - Bad RH sensortwpsondewnpnC2.b1Suspect
D051122.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1Incorrect
D051219.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1Incorrect
D070920.2TWP/SONDE/C2 - Failure of RH sensortwpsondewnpnC2.b1Suspect


DQRID : D021206.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/11/2003234506/12/20030112Incorrect
04/24/2003235804/25/20030126Incorrect
12/24/2002233112/25/20020112Incorrect
06/02/2003113406/02/20031321Incorrect
11/25/2002233711/26/20020001Incorrect
09/27/2003232709/28/20030132Incorrect
06/06/2003233506/07/20030057Incorrect
06/26/2003233506/27/20030114Incorrect
04/23/2003112904/23/20031308Incorrect
09/18/2002113709/18/20021301Incorrect
12/16/2002113812/16/20021248Incorrect
06/09/2003234006/10/20030116Incorrect
04/22/2003234104/23/20030122Incorrect
11/23/2002233011/24/20020108Incorrect
12/27/2002234412/28/20020131Incorrect
09/24/2002012909/24/20020204Incorrect
12/22/2002013612/22/20020204Incorrect
06/14/2003113306/14/20031300Incorrect
03/28/2003021403/28/20030339Incorrect
09/24/2002022009/24/20020352Incorrect
Subject:
TWP/SONDE/C2  - Broken Temperature Sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensors on these radiosondes broke at launch.  All
the temperature-related data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D030419.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/13/2003012404/13/20030200Incorrect
Subject:
TWP/SONDE/C2 - Bad RH Data (AIRS Validation Sonde)
DataStreams:twpsondewnpnC2.b1
Description:
The RH sensor heating circuit failed on this sonde.  This results in
oscillation of reported RH between reasonable values and values near zero.
This was an AIRS validation sounding.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D030608.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/03/2003233906/04/20030016Suspect
Subject:
TWP/SONDE/C2  - Failure of RH sensor
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D031008.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/02/2003023210/02/20030300Incorrect
Subject:
TWP/SONDE/C2 - Strange problem with data -
DataStreams:twpsondewnpnC2.b1
Description:
This is very strange.  The data has what appears to be a pressure jump at launch (from 
1007.2 to 956.5 hPa) but the sounding temperature and RH data clearly look to be 
interpolated between the surface and tropopause.  By examining the absolutely raw data, I found that 
the system assigned an elapsed time of 408 seconds (since system start up) to the launch 
point, but there is no raw data between 410 and 532 seconds.  The raw pressure value at 
410 seconds is 1007.6 hPa and that at 532 seconds is 40.9 hPa!  Unless this sounding was 
done by using a rocket-powered balloon (maybe even then), it would be impossible for the 
sonde to get to 40.9 hPa in 2 minutes.  Because the apparent gap is only two minutes, the 
system was able to interpolate but is not sophisticated enough to understand that the 
limits of the interpolation are unreasonable.

I can't imagine what went wrong - I've not seen anything like this before.
Measurements:twpsondewnpnC2.b1:
  • East longitude(lon)
  • Wind speed, W-to-E component, at altitude, 915 MHz RWP(u_wind)
  • Wind speed, S-to-N component, at altitude, derived from ETA model(v_wind)
  • asc
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)
  • base_time
  • time_offset
  • wstat
  • Wind Direction(deg)
  • Altitude above mean sea level(alt)
  • Wind speed, surface, 1-min avg(wspd)
  • Pressure, atmospheric, constant value of model level, RUC model output(pres)
  • North latitude(lat)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D031231.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/22/2003002510/22/20030155Suspect
10/22/2003234010/23/20030121Suspect
Subject:
TWP/SONDE/C2 - RH sensor bad
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuits failed during these
soundings.  The RH oscillates rapidly between low (~0 %RH) and not
so low (ambient %RH) and the oscillation seems to stop when the
temperature is below ~35 degC.  According to Vaisala these are
symptoms of failure in the ASIC (application specific integrated
circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D040127.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/29/2004233203/30/20040047Incorrect
05/13/2004233405/14/20040112Incorrect
01/19/2004022301/19/20040346Incorrect
11/18/2003120511/18/20031340Incorrect
12/23/2003234212/24/20030132Incorrect
12/21/2003115512/21/20031344Incorrect
05/14/2004140605/14/20041411Incorrect
01/18/2004113501/18/20041256Incorrect
03/09/2004114503/09/20041315Incorrect
12/20/2003122512/20/20031408Incorrect
02/26/2004122102/26/20041350Incorrect
12/23/2003113712/23/20031259Incorrect
06/07/2004022406/07/20040351Incorrect
11/19/2003115611/19/20031346Incorrect
06/25/2004114606/25/20041328Incorrect
11/08/2003233511/09/20030215Incorrect
07/02/2004005907/02/20040217Incorrect
01/16/2004233701/17/20040109Incorrect
01/09/2004115701/09/20041342Incorrect
12/21/2003005612/21/20030126Incorrect
09/08/2004142509/08/20041605Incorrect
02/17/2004140602/17/20041536Incorrect
01/12/2004234201/13/20040214Incorrect
10/18/2004234610/19/20040128Incorrect
03/30/2004233403/31/20040059Incorrect
01/08/2004113501/08/20041309Incorrect
01/26/2004120201/26/20040146Incorrect
12/08/2004114312/08/20041306Incorrect
03/22/2004113403/22/20041300Incorrect
01/23/2004113701/23/20041319Incorrect
11/21/2003004411/21/20030113Incorrect
11/29/2004232111/30/20040105Incorrect
01/19/2004232901/20/20040157Incorrect
10/13/2003004010/13/20030203Incorrect
03/14/2004113103/14/20041300Incorrect
05/01/2004233505/02/20040110Incorrect
12/30/2004113012/30/20041239Incorrect
07/24/2004123507/24/20041355Incorrect
09/08/2004233809/08/20042345Incorrect
01/16/2004140901/16/20041548Incorrect
08/22/2004114408/22/20041207Incorrect
04/11/2004234804/12/20040120Incorrect
01/22/2004005201/22/20040227Incorrect
07/20/2004022007/20/20040358Incorrect
12/23/2003004212/23/20030212Incorrect
05/02/2004233605/03/20040109Incorrect
01/21/2004125201/21/20041400Incorrect
01/09/2004135901/09/20041537Incorrect
Subject:
TWP/SONDE/C2 - Broken temperature sensors
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensors on these radiosondes broke after launch.  All
temperature-related data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D040208.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/04/2004114702/04/20041337Suspect
Subject:
TWP/SONDE/C2  - Questionable rH data
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D040219.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/18/2004003502/18/20040205Suspect
Subject:
TWP/SONDE/C2  - Bad RH sensor
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D040425.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/21/2004123204/21/20041248Incorrect
Subject:
TWP/SONDE/C2 - Bad RH values aloft
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit on this sonde failed about 24 minutes into 
the flight.  The RH values look reasonable, though somewhat noisy, from the surface to 
approximately 6.9 km (439 hPa) at which point they begin to oscillate between 0 and 100 %.  
This behavior stops when the sonde reaches approximately 11.5km (232 hPa) when the air 
temperature drops below -40 degC.  The oscillation stops at that point, though the RH 
remains higher than might be expected even under supersaturated conditions.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D040711.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/10/2004123807/10/20041315Suspect
Subject:
TWP/SONDE/C2  - RH sensor failure near surface
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D040815.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/12/2004004408/12/20040233Incorrect
Subject:
TWP/SONDE/C2 - Operator used wrong calibration tape
DataStreams:twpsondewnpnC2.b1
Description:
It appears as though the opperator used the wrong calibration tape for this sounding.  The 
primary symptom is a jump in temperature from a surface value of 30.1 degC to the first 
value aloft of 61.0 degC.  Secondly, the recorded serial number for this sounding is 
Z1430179 which is the same serial number recorded for the previous sounding.

This problem points to three failures of procedure (I have notified TWP site ops by 
e-mail).  The first failure is using the wrong tape. The second failure is not checking the 
radiosonde values before launch which relates to the third failure of not entering the 
radiosonde values as the surface values of the sounding.
Measurements:twpsondewnpnC2.b1:
  • East longitude(lon)
  • Wind speed, W-to-E component, at altitude, 915 MHz RWP(u_wind)
  • Wind speed, S-to-N component, at altitude, derived from ETA model(v_wind)
  • asc
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)
  • base_time
  • time_offset
  • wstat
  • Wind Direction(deg)
  • Altitude above mean sea level(alt)
  • Wind speed, surface, 1-min avg(wspd)
  • Pressure, atmospheric, constant value of model level, RUC model output(pres)
  • North latitude(lat)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D040921.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/16/2004113509/16/20041210Suspect
Subject:
TWP/SONDE/C2  - RH sensor failure below 9km
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D040921.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/19/2004140809/19/20041447Incorrect
09/18/2004005809/18/20040133Incorrect
09/17/2004114509/17/20041216Incorrect
Subject:
TWP/SONDE/C2 - Bad RH values aloft
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit on these sondes failed after launch but 
resume proper operation when the ambient temperature dropped below ~40 degC.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D050913.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/03/2005235007/04/20050024Suspect
07/19/2005233207/20/20050016Suspect
07/10/2005234007/11/20050021Suspect
Subject:
TWP/SONDE/C2 - Bad RH sensor
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below -50 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents

DQRID : D051122.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/21/2005233211/22/20050140Incorrect
11/19/2005233411/20/20050153Incorrect
Subject:
TWP/SONDE/C2 - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D051219.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/07/2005115312/07/20051323Incorrect
Subject:
TWP/SONDE/C2  - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D070920.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/19/2003122103/19/20031253Suspect
08/20/2003234008/21/20030002Suspect
Subject:
TWP/SONDE/C2 - Failure of RH sensor
DataStreams:twpsondewnpnC2.b1
Description:
The RH sensor ASIC that controls the sensor heating failed.  As a result the RH at the 
beginning of the sounding oscillates between near-zero and ambient values.  The RH heating 
cycle stops when the ambient temperature reaches below -30 degC; above this level RH 
values should be valid.
Measurements:twpsondewnpnC2.b1:
  • Humidity, relative, at 21-m height, 30-min avg(rh)
  • Dewpoint Temperature(dp)


Back To Table of Contents