Data Quality Reports for Session: 179724 User: greened Completed: 09/15/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160610.15MCQ/SKYRAD/S1 - IRT not installedmcqskyrad60sS1.b1, mcqskyrad20sS1.a0Incorrect
D160708.2SGP/MWR3C/E32 - Incorrect datasgpmwr3cE32.b1Incorrect
D160728.1SGP/MWR3C/E32 - Missing datasgpmwr3cE32.b1Missing
D160801.4ASI/SONDE/S1 - Reprocess: Incorrect lat and lonasisonderawS1.00Incorrect
D160802.6SGP/MFRSR/E41 - Heater board failuresgpmfrsrE41.00Incorrect
D160811.1SGP/EBBR/E11 - Soil Temperature 1 Spikingsgp30ebbrE11.b1Suspect
D160819.1AWR/RWP/M1 - Data collectionawr1290bsrwpspeclowprecipM1.a0, awr1290bsrwpspeclolowwindM1.a0,
awr1290bsrwpspeclolowprecipM1.a0, awr1290bsrwpspeclowwindM1.a0
Missing
D160824.9OLI/AOSMET/M1 - Reprocess: Wind direction incorrectoliaosmetM1.a1Incorrect
D160826.3SGP/EBBR/E11 - Sensible and Latent Heat Fluxes Incorrectsgp30ebbrE11.b1Incorrect
D160826.5SGP/EBBR/E11 - Soil temperature 2 Spikingsgp30ebbrE11.b1Suspect
D160830.2OLI/UHSAS/M1 - High counts below 100 nmoliaosuhsasM1.a1Incorrect
D160907.14ASI/SMPS/M1 - Instrument maintenceasiaossmpstrhM1.a1Incorrect


DQRID : D160610.15
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/03/2016054303/31/20182359Incorrect
Subject:
MCQ/SKYRAD/S1 - IRT not installed
DataStreams:mcqskyrad60sS1.b1, mcqskyrad20sS1.a0
Description:
An IRT was not installed so the datalogger sampled a floating signal.
Measurements:mcqskyrad60sS1.b1:
  • Temperature, Sky Infra-Red(sky_ir_temp)
  • sky_ir_temp_max
  • sky_ir_temp_std
  • sky_ir_temp_min

mcqskyrad20sS1.a0:
  • inst_sky_ir_signal


Back To Table of Contents

DQRID : D160708.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/29/2016000006/29/20161920Incorrect
Subject:
SGP/MWR3C/E32 - Incorrect data
DataStreams:sgpmwr3cE32.b1
Description:
The instrument had incorrect receiver parameters.
Measurements:sgpmwr3cE32.b1:
  • Liquid water path best-estimate value, hourly mean(lwp)
  • 89.0 GHz sky brightness temperature(tbsky89)
  • Column precip water vapor(pwv)


Back To Table of Contents

DQRID : D160728.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/17/2016000007/18/20162359Missing
Subject:
SGP/MWR3C/E32 - Missing data
DataStreams:sgpmwr3cE32.b1
Description:
Data are missing
Measurements:sgpmwr3cE32.b1:
  • hail_duration
  • surface_pressure
  • qc_surface_temperature
  • qc_surface_relative_humidity
  • qc_wind_direction_avg
  • qc_wind_direction_max
  • qc_wind_direction_min
  • 30.0 GHz sky brightness temperature(tbsky30)
  • rain_accumulation
  • Column precip water vapor(pwv)
  • qc_hail_intensity
  • wind_direction_min
  • wind_direction_max
  • rain_peak_intensity
  • lwp_err
  • qc_tbsky89
  • azimuth
  • elevation
  • hail_intensity
  • 23.8 GHz sky brightness temperature(tbsky23)
  • wind_speed_max
  • qc_time
  • qc_rain_duration
  • qc_hail_accumulation
  • 89.0 GHz sky brightness temperature(tbsky89)
  • qc_wind_speed_min
  • time
  • qc_wind_speed_max
  • qc_rain_peak_intensity
  • qc_azimuth
  • qc_surface_pressure
  • qc_infrared_temperature
  • infrared_temperature
  • Liquid water path best-estimate value, hourly mean(lwp)
  • qc_hail_duration
  • pwv_err
  • wind_direction_avg
  • base_time
  • Rain Intensity(rain_intensity)
  • qc_elevation
  • North latitude(lat)
  • wind_speed_min
  • Altitude above mean sea level(alt)
  • qc_wind_speed_avg
  • qc_tbsky23
  • qc_rain_intensity
  • wind_speed_avg
  • surface_relative_humidity
  • hail_accumulation
  • qc_rain_accumulation
  • time_offset
  • surface_temperature
  • East longitude(lon)
  • qc_tbsky30
  • Rain Duration(rain_duration)


Back To Table of Contents

DQRID : D160801.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/01/2016022908/01/20160323Incorrect
Subject:
ASI/SONDE/S1 - Reprocess: Incorrect lat and lon
DataStreams:asisonderawS1.00
Description:
Somehow the initial Lat/Lon settings were corrupted.  Data has been reprocessed and needs 
to be collected and ingested.
Measurements:asisonderawS1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D160802.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/07/2016184606/30/20161844Incorrect
Subject:
SGP/MFRSR/E41 - Heater board failure
DataStreams:sgpmfrsrE41.00
Description:
The heater board failed completely.  It was replaced on 20160630 and doing so resolved the 
problem.
Measurements:sgpmfrsrE41.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D160811.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/05/2016100008/08/20161600Suspect
Subject:
SGP/EBBR/E11 - Soil Temperature 1 Spiking
DataStreams:sgp30ebbrE11.b1
Description:
Soil temperature 1 spiked at times, the result of loose lead connections.

This did not have a significant effect on the calculation of sensible and latent heat 
fluxes.
Measurements:sgp30ebbrE11.b1:
  • energy_storage_change_1
  • Soil temperature mean 1(soil_temp_1)


Back To Table of Contents

DQRID : D160819.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/10/2016001108/11/20161700Missing
Subject:
AWR/RWP/M1 - Data collection
DataStreams:awr1290bsrwpspeclowprecipM1.a0, awr1290bsrwpspeclolowwindM1.a0,
awr1290bsrwpspeclolowprecipM1.a0, awr1290bsrwpspeclowwindM1.a0
Description:
Instrument computer was put on stand by since it had reached 90% storage capacity. 
Therefore, data is missing/was not recorded during the DQPR period.
Measurements:awr1290bsrwpspeclowprecipM1.a0:
  • pulse_width
  • num_time_domain_integrations
  • first_gate_height
  • time
  • beam_azimuth
  • num_fft_points
  • East longitude(lon)
  • interpulse_period
  • base_time
  • gate_interval
  • beam_zenith
  • num_frequency_domain_integrations
  • time_offset
  • Altitude above mean sea level(alt)
  • Spectrum of raw counts from spectrometer(spectra)
  • North latitude(lat)
  • mode_number
  • beam_number
  • noise_threshold
  • fft_frequency_spacing
  • first_reported_spectra_bin

awr1290bsrwpspeclowwindM1.a0:
  • first_reported_spectra_bin
  • beam_number
  • gate_interval
  • fft_frequency_spacing
  • num_frequency_domain_integrations
  • pulse_width
  • first_gate_height
  • noise_threshold
  • num_fft_points
  • time
  • mode_number
  • base_time
  • time_offset
  • East longitude(lon)
  • Altitude above mean sea level(alt)
  • North latitude(lat)
  • interpulse_period
  • beam_azimuth
  • num_time_domain_integrations
  • Spectrum of raw counts from spectrometer(spectra)
  • beam_zenith

awr1290bsrwpspeclolowwindM1.a0:
  • North latitude(lat)
  • first_gate_height
  • base_time
  • beam_number
  • Spectrum of raw counts from spectrometer(spectra)
  • mode_number
  • beam_azimuth
  • Altitude above mean sea level(alt)
  • time_offset
  • beam_zenith
  • fft_frequency_spacing
  • num_frequency_domain_integrations
  • pulse_width
  • East longitude(lon)
  • noise_threshold
  • first_reported_spectra_bin
  • gate_interval
  • interpulse_period
  • time
  • num_fft_points
  • num_time_domain_integrations

awr1290bsrwpspeclolowprecipM1.a0:
  • beam_zenith
  • time
  • base_time
  • interpulse_period
  • num_frequency_domain_integrations
  • noise_threshold
  • first_gate_height
  • beam_azimuth
  • North latitude(lat)
  • num_fft_points
  • fft_frequency_spacing
  • time_offset
  • first_reported_spectra_bin
  • Spectrum of raw counts from spectrometer(spectra)
  • beam_number
  • East longitude(lon)
  • mode_number
  • gate_interval
  • num_time_domain_integrations
  • pulse_width
  • Altitude above mean sea level(alt)


Back To Table of Contents

DQRID : D160824.9
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/06/2016182108/07/20160036Incorrect
Subject:
OLI/AOSMET/M1 - Reprocess: Wind direction incorrect
DataStreams:oliaosmetM1.a1
Description:
The wind sensor was not aligned to true north during this time. It was aligned 59 degrees 
west of north.
Measurements:oliaosmetM1.a1:
  • Wind direction, relative to true North(WindDirection)


Back To Table of Contents

DQRID : D160826.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/19/2016103008/22/20161700Incorrect
Subject:
SGP/EBBR/E11 - Sensible and Latent Heat Fluxes Incorrect
DataStreams:sgp30ebbrE11.b1
Description:
Soil temperature 1 and soil moisture 1 were incorrect, resulting in the sensible and 
latent heat fluxes being incorrect.
Measurements:sgp30ebbrE11.b1:
  • Latent heat flux(latent_heat_flux)
  • Average soil heat flow at the surface(surface_soil_heat_flux_avg)
  • Sensible heat flux(sensible_heat_flux)
  • energy_storage_change_1
  • Soil temperature mean 1(soil_temp_1)
  • surface_soil_heat_flux_1
  • soil_moisture_1


Back To Table of Contents

DQRID : D160826.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/15/2016070008/22/20161200Suspect
Subject:
SGP/EBBR/E11 - Soil temperature 2 Spiking
DataStreams:sgp30ebbrE11.b1
Description:
Sensible and latent heat flux was not deleteriously affected when small spikes
occurred in soil temperature 2.
Measurements:sgp30ebbrE11.b1:
  • Soil temperature mean 2(soil_temp_2)
  • energy_storage_change_2


Back To Table of Contents

DQRID : D160830.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/24/2016165608/24/20162345Incorrect
Subject:
OLI/UHSAS/M1 - High counts below 100 nm
DataStreams:oliaosuhsasM1.a1
Description:
High counts below 100 nm. Most likely some debris stuck to the jet nozzle. No valid data.
Measurements:oliaosuhsasM1.a1:
  • lower_size_limit
  • MET air temperature(temperature)
  • upper_size_limit
  • flow
  • current
  • sample_pressure
  • sampling_volume
  • unit_temperature
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)
  • bin_num
  • purge_flow
  • reference
  • auxiliary
  • differential_voltage
  • sheath_flow
  • scattering_voltage


Back To Table of Contents

DQRID : D160907.14
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/06/2016103009/06/20161154Incorrect
08/29/2016152708/29/20161549Incorrect
09/06/2016172509/06/20161744Incorrect
08/29/2016160508/31/20161719Incorrect
09/02/2016130509/02/20161334Incorrect
09/06/2016180009/06/20161824Incorrect
09/01/2016130909/01/20161814Incorrect
08/29/2016125508/29/20161344Incorrect
Subject:
ASI/SMPS/M1 - Instrument maintence
DataStreams:asiaossmpstrhM1.a1
Description:
During these time periods, a HEPA filter was connected to different portions of the system 
for investigation of leaks.
Measurements:asiaossmpstrhM1.a1:
  • Sample relative humidity(sample_rh)
  • Sample temperature(sample_temperature)


Back To Table of Contents