DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140121.3 | TWP/RL/C3 - Missing T1 and/or T2 (rotational Raman) channels | twp10rlproftemp1newsC3.c0 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/15/2014 | 0000 | 02/11/2014 | 0000 | Incorrect |
12/28/2013 | 0000 | 01/09/2014 | 0059 | Incorrect |
Subject: | TWP/RL/C3 - Missing T1 and/or T2 (rotational Raman) channels |
DataStreams: | twp10rlproftemp1newsC3.c0 |
Description: | Data for the T1 and/or T2 (rotation Raman) channels are missing due to failures of multiple high voltage power supplies for PMTs. As a result, it is not be possible to derive temperature profiles from the Raman lidar during the effected period. Temperature profile data from the Raman lidar are incorrect. Do not use for analysis. |
Measurements: | twp10rlproftemp1newsC3.c0:
|