DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D040615.5 | SGP/SIRS/E12 - Incorrect Times | sgp1swfanalsirs1longE12.c1 | Suspect |
D050502.7 | SGP/SIRS/E12 - SWFANAL Data description error | sgp1swfanalsirs1longE12.c1 | Does not affect quality |
D120420.3 | SGP/QCRAD/C1 - Reprocess: SW data require -1 multipler | sgpqcrad1longC1.s2 | Incorrect |
D120420.4 | SGP/QCRAD/C1 - Upwelling SW cable problem | sgpqcrad1longC1.s2 | Incorrect |
D120420.5 | SGP/QCRAD/C1 - Sensor cable problem | sgpqcrad1longC1.s2 | Incorrect |
D120420.6 | SGP/QCRAD/C1 - SW direct normal data drop out | sgpqcrad1longC1.s2 | Incorrect |
D120420.7 | SGP/QCRAD/C1 - Failed upwelling SW cable | sgpqcrad1longC1.s2 | Incorrect |
D120420.8 | SGP/QCRAD/C1 - NIP failure | sgpqcrad1longC1.s2 | Incorrect |
D120510.5 | SGP/QCRAD/C1 - Case and dome temperatures not available | sgpqcrad1longC1.s2 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/19/2004 | 2212 | 03/10/2004 | 2247 | Suspect |
Subject: | SGP/SIRS/E12 - Incorrect Times |
DataStreams: | sgp1swfanalsirs1longE12.c1 |
Description: | The ntpd server (time adjustment) stopped working on Feb. 19th at 22:12:16 GMT, and the clock was corrected by -5600.177269 seconds on March 10th at 22:47 GMT. The recorded times for observations are off by amounts ranging from 0 seconds on Feb. 19th to 5600 seconds on March 10th. |
Measurements: | sgp1swfanalsirs1longE12.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/21/1996 | 0000 | 10/28/2003 | 2359 | Does not affect quality |
Subject: | SGP/SIRS/E12 - SWFANAL Data description error |
DataStreams: | sgp1swfanalsirs1longE12.c1 |
Description: | The long name attribute for the cloud effect fields incorrectly defines the cloud effect to be the difference between the clearskyfit and measured values: "Difference: ***fluxdn_clearskyfit - ***fluxdn_measured (***fcg)" ; The cloud effects are correctly calculated by subtracting the irradiance calculated in the clear sky fit from the measured irradiance. "Difference: ***fluxdn_measured - ***fluxdn_clearskyfit (***fcg)" ; NOTE: this is just a documentation error. The data are correctly calculated. |
Measurements: | sgp1swfanalsirs1longE12.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/24/2000 | 1900 | 05/08/2000 | 2359 | Incorrect |
Subject: | SGP/QCRAD/C1 - Reprocess: SW data require -1 multipler |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | Magnitude of data are correct, however, 2 wires were crossed producing negative values for SW Global Down, SW Global Up, SW Direct Normal, and SW Diffuse Down. Problem has been fixed at site. Users of this data should multiply by -1 to get the correct values for SW Global Down, SW Global Up, SW Direct Normal, and SW Diffuse Down. Data will be reprocessed. Documented in instrument DQR D000919.1. |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/13/2006 | 0900 | 01/23/2007 | 1930 | Incorrect |
Subject: | SGP/QCRAD/C1 - Upwelling SW cable problem |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | Starting 11/13/2006, the C1 SIRS Upwelling Shortwave (up_short_hemisp) started to show large differences in comparison to co-located E13 SIRS and C1 IRT25m. After much troubleshooting and gnashing of teeth it was discovered that rodents had chewed through the instrument cables at the bottom of the tilt tower. Cables were replaced 01/23/2007 1930GMT. Documented in instrument DQR D061121.2 |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/03/2007 | 1400 | 04/22/2008 | 2052 | Incorrect |
Subject: | SGP/QCRAD/C1 - Sensor cable problem |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | Beginning October 3, 2007, the upwelling longwave irradiance (up_long_hemisp, aka "UIR") started to spike to unreasonable values positively offset from the co-located E13 by over 100W/m2. On January 17 2008 the UIR values become consistently bad due to bad UIR dome thermistor wiring. Problem appears as a resistive offset in the dome thermistor but due to nature of failure (unknown cable problem) data correction would not yield any better data than the co-located E13 UIR instrument that is known to be operating correctly. Instrument cable replacement on 4/22 fixed the problem. Thermopile data (netIR) from this instrument is good. However, incoming irradiance data is not correctable due to bad UIR dome thermistor wiring within the instrument cable. Documented in instrument DQR D071022.4. |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/18/2009 | 1751 | 07/31/2009 | 1645 | Incorrect |
Subject: | SGP/QCRAD/C1 - SW direct normal data drop out |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | NIP cable had failed during this time period. Short_direct_normal data are not recoverable during this time period. However, it can be reconstructed using the following formula for each one minute data: short_direct_normal = (down_short_hemisp-down_short_diffuse_hemisp)/cos(zenith angle) Documented in instrument DQR D090623.1 |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/20/2010 | 0414 | 06/30/2010 | 1442 | Incorrect |
Subject: | SGP/QCRAD/C1 - Failed upwelling SW cable |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | The upwelling shortwave hemispheric (up_short_hemisp) data were missing or incorrect due to a bad cable. Cable was replaced on 6/30 and the problem was resolved. Documented in instrument DQR D100701.1 |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/12/2011 | 0000 | 06/24/2011 | 1740 | Incorrect |
Subject: | SGP/QCRAD/C1 - NIP failure |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | The short_direct_normal (NIP instrument) cable failed during this time period. Data are not recoverable. Users can derive an estimate of direct normal values using the following formula: short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle) Documented in instrument DQR D110628.2 |
Measurements: | sgpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/01/1999 | 0000 | 12/31/2000 | 2359 | Suspect |
Subject: | SGP/QCRAD/C1 - Case and dome temperatures not available |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | Longwave downwelling and upwelling case and dome temperatures are not available during this time period because they were not kept in the sirs.a1 files and sirs.b1 files are not available. This means that some longwave tests and corrections can't be applied. |
Measurements: | sgpqcrad1longC1.s2:
|