DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D131024.2 | OLI/GNDRAD/M1 - Instrument not properly grounded | oligndrad60sM1.b1 | Incorrect |
D131104.1 | OLI/SKYRAD/M1 - Tracker misaligned | oliskyrad60sM1.b1 | Incorrect |
D140805.3 | OLI/SKYRAD/M1 - Solar tracker not functioning correctly | oliskyrad60sM1.b1 | Incorrect |
D140905.4 | OLI/SKYRAD/M1 - Case and dome resistances near zero | oliskyrad60sM1.b1 | Suspect |
D140908.3 | OLI/SKYRAD/M1 - Missing data | oliskyrad60sM1.b1 | Missing |
D141212.2 | OLI/SKYRAD/M1 - Instrument problem | oliskyrad60sM1.b1 | Incorrect |
D150123.2 | OLI/SKYRAD/M1 - Environmental | oliskyrad60sM1.b1 | Missing |
D150123.3 | OLI/SKYRAD/M1 - Missing data resulting from loss of power | oliskyrad60sM1.b1 | Missing |
D150123.7 | OLI/GNDRAD/M1 - Environmental | oligndrad60sM1.b1 | Incorrect |
D150209.1 | OLI/SKYRAD/M1 - Instrument | oliskyrad60sM1.b1 | Incorrect |
D150327.3 | OLI/SKYRAD/M1 - Loss of power | oliskyrad60sM1.b1 | Missing |
D150327.7 | OLI/GNDRAD/M1 - Environmental | oligndrad60sM1.b1 | Incorrect |
D150526.1 | OLI/SKYRAD/M1 - Sun tracker not operational | oliskyrad60sM1.b1 | Incorrect |
D150531.1 | OLI/SKYRAD/M1 - Negative offset | oliskyrad60sM1.b1 | Suspect |
D150722.1 | OLI/SKYRAD/M1 - Loss of power caused tracker to stop | oliskyrad60sM1.b1 | Incorrect |
D150727.15 | OLI/SKYRAD/M1 - Spikes in data | oliskyrad60sM1.b1 | Incorrect |
D151102.3 | OLI/GNDRAD/M1 - PIR cable connector failure | oligndrad60sM1.b1 | Incorrect |
D160708.1 | OLI/SKYRAD/M1 - Intermittent spiking | oliskyrad60sM1.b1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/19/2013 | 0000 | 10/18/2013 | 1820 | Incorrect |
Subject: | OLI/GNDRAD/M1 - Instrument not properly grounded |
DataStreams: | oligndrad60sM1.b1 |
Description: | At the start of data collection all OLI GNDRAD measurements were corrupted by a grounding problem. Additional grounding methods were put into place on 10/18/2013 eliminating the problem. All GNDRAD data are suspect during the specified time range. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/19/2013 | 0000 | 11/03/2013 | 2210 | Incorrect |
Subject: | OLI/SKYRAD/M1 - Tracker misaligned |
DataStreams: | oliskyrad60sM1.b1 |
Description: | At the beginning of Oliktok SKYRAD data collection the sun tracker was misaligned. The down_short_diffuse_hemisp (8-48 instrument) and short_direct_normal (NIP instrument) datastreams were adversely affected resulting in incorrect data. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/13/2014 | 2156 | 07/30/2014 | 2130 | Incorrect |
Subject: | OLI/SKYRAD/M1 - Solar tracker not functioning correctly |
DataStreams: | oliskyrad60sM1.b1 |
Description: | The solar tracker was not functioning correctly during the affected time range. Data from instruments is directly reliant on correct tracker operation (short_direct_normal and down_short_diffuse) and are not recoverable. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/09/2014 | 0855 | 08/17/2014 | 1805 | Suspect |
Subject: | OLI/SKYRAD/M1 - Case and dome resistances near zero |
DataStreams: | oliskyrad60sM1.b1 |
Description: | PIR1 dome and case resistances intermittently near zero during affected time period. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/27/2014 | 1924 | 08/29/2014 | 2308 | Missing |
Subject: | OLI/SKYRAD/M1 - Missing data |
DataStreams: | oliskyrad60sM1.b1 |
Description: | PIR2 data missing during affected time period. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/16/2014 | 0700 | 11/17/2014 | 2000 | Incorrect |
09/13/2014 | 2000 | 09/22/2014 | 1900 | Incorrect |
11/02/2014 | 1600 | 11/03/2014 | 0300 | Incorrect |
Subject: | OLI/SKYRAD/M1 - Instrument problem |
DataStreams: | oliskyrad60sM1.b1 |
Description: | PIR1 data incorrect during the affected time period. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/31/2014 | 0300 | 01/09/2015 | 1724 | Missing |
Subject: | OLI/SKYRAD/M1 - Environmental |
DataStreams: | oliskyrad60sM1.b1 |
Description: | SKYRAD system not operating during the affected time period due to site power outage. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/12/2015 | 2230 | 01/15/2015 | 0030 | Missing |
Subject: | OLI/SKYRAD/M1 - Missing data resulting from loss of power |
DataStreams: | oliskyrad60sM1.b1 |
Description: | The site lost power. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/12/2014 | 2230 | 01/15/2015 | 0030 | Incorrect |
12/31/2014 | 0300 | 01/09/2015 | 0500 | Incorrect |
Subject: | OLI/GNDRAD/M1 - Environmental |
DataStreams: | oligndrad60sM1.b1 |
Description: | GNDRAD system not operating correctly during affected time periods due to power outage. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/30/2014 | 1200 | 10/01/2014 | 1500 | Incorrect |
01/22/2015 | 2130 | 01/23/2015 | 1500 | Incorrect |
11/20/2014 | 0330 | 11/22/2014 | 1900 | Incorrect |
01/28/2015 | 1000 | 02/05/2015 | 0230 | Incorrect |
12/29/2014 | 1300 | 12/31/2014 | 0400 | Incorrect |
09/28/2014 | 0400 | 09/28/2014 | 1130 | Incorrect |
12/22/2014 | 0900 | 12/23/2014 | 1130 | Incorrect |
11/10/2014 | 2300 | 11/14/2014 | 1900 | Incorrect |
09/20/2014 | 0300 | 09/20/2014 | 2359 | Incorrect |
10/26/2014 | 0700 | 10/26/2014 | 1700 | Incorrect |
12/13/2014 | 0200 | 12/13/2014 | 1330 | Incorrect |
10/28/2014 | 0000 | 11/04/2014 | 0500 | Incorrect |
10/16/2014 | 2000 | 10/23/2014 | 1700 | Incorrect |
12/25/2014 | 1900 | 12/27/2014 | 1100 | Incorrect |
01/19/2015 | 0400 | 01/20/2015 | 0300 | Incorrect |
Subject: | OLI/SKYRAD/M1 - Instrument |
DataStreams: | oliskyrad60sM1.b1 |
Description: | PIR2 data incorrect during the affected time period. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/06/2015 | 1206 | 03/18/2015 | 0054 | Missing |
Subject: | OLI/SKYRAD/M1 - Loss of power |
DataStreams: | oliskyrad60sM1.b1 |
Description: | Power outage during affected time period, no AC power to logger power supply, tracker, ventilator fans, ventilator heaters. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/01/2015 | 1916 | 03/08/2015 | 0043 | Incorrect |
Subject: | OLI/GNDRAD/M1 - Environmental |
DataStreams: | oligndrad60sM1.b1 |
Description: | AC power out during affected time range affecting ventilator fans and heaters for up_short_hemisp, and up_long_hemisp radiometers. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/18/2015 | 0100 | 05/07/2015 | 2330 | Incorrect |
Subject: | OLI/SKYRAD/M1 - Sun tracker not operational |
DataStreams: | oliskyrad60sM1.b1 |
Description: | During the specified time period, the suntracker was not operating correctly. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/20/2014 | 0800 | 11/22/2014 | 2359 | Suspect |
Subject: | OLI/SKYRAD/M1 - Negative offset |
DataStreams: | oliskyrad60sM1.b1 |
Description: | During the affected time period, a negative offset existed between the derived down short hemispheric and down short hemispheric values, possibly associated with instrument responses at low temperatures and irradiance levels. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/01/2015 | 1916 | 03/18/2015 | 1206 | Incorrect |
Subject: | OLI/SKYRAD/M1 - Loss of power caused tracker to stop |
DataStreams: | oliskyrad60sM1.b1 |
Description: | Loss of power to instrument caused the tracker, blower and heater causing data to be incorrect. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/25/2015 | 2130 | 06/25/2015 | 2230 | Incorrect |
06/16/2015 | 1800 | 06/16/2015 | 1930 | Incorrect |
06/06/2015 | 0615 | 06/06/2015 | 1700 | Incorrect |
02/21/2015 | 1400 | 02/21/2015 | 2359 | Incorrect |
02/24/2015 | 0000 | 02/24/2015 | 0830 | Incorrect |
03/19/2015 | 0000 | 03/19/2015 | 2359 | Incorrect |
06/04/2015 | 0000 | 06/04/2015 | 1930 | Incorrect |
02/19/2015 | 0000 | 02/19/2015 | 2359 | Incorrect |
06/04/2015 | 2200 | 06/04/2015 | 2359 | Incorrect |
04/08/2015 | 0000 | 04/08/2015 | 1600 | Incorrect |
06/16/2015 | 1500 | 06/16/2015 | 1515 | Incorrect |
06/07/2015 | 0630 | 06/07/2015 | 1900 | Incorrect |
02/17/2015 | 2100 | 02/17/2015 | 2300 | Incorrect |
05/16/2015 | 2045 | 05/16/2015 | 2200 | Incorrect |
02/26/2015 | 0000 | 02/26/2015 | 2359 | Incorrect |
02/22/2015 | 0000 | 02/22/2015 | 1730 | Incorrect |
06/15/2015 | 1800 | 06/15/2015 | 1830 | Incorrect |
05/17/2015 | 1500 | 05/17/2015 | 1730 | Incorrect |
04/01/2015 | 0430 | 04/01/2015 | 0700 | Incorrect |
06/26/2015 | 1815 | 06/26/2015 | 2230 | Incorrect |
06/24/2015 | 1500 | 06/24/2015 | 1600 | Incorrect |
05/15/2015 | 0500 | 05/15/2015 | 0530 | Incorrect |
03/01/2015 | 0000 | 03/01/2015 | 2359 | Incorrect |
03/02/2015 | 0000 | 03/02/2015 | 2359 | Incorrect |
05/27/2015 | 0630 | 05/27/2015 | 1745 | Incorrect |
03/03/2015 | 0000 | 03/03/2015 | 0230 | Incorrect |
04/07/2015 | 0000 | 04/07/2015 | 2300 | Incorrect |
04/01/2015 | 1600 | 04/01/2015 | 2330 | Incorrect |
02/25/2015 | 1430 | 02/25/2015 | 2359 | Incorrect |
05/08/2015 | 2130 | 05/08/2015 | 2330 | Incorrect |
04/05/2015 | 1200 | 04/05/2015 | 2359 | Incorrect |
06/17/2015 | 1830 | 06/17/2015 | 2030 | Incorrect |
04/06/2015 | 0000 | 04/06/2015 | 2300 | Incorrect |
02/23/2015 | 2100 | 02/23/2015 | 2359 | Incorrect |
06/03/2015 | 1600 | 06/03/2015 | 2359 | Incorrect |
03/18/2015 | 0500 | 03/18/2015 | 2359 | Incorrect |
02/20/2015 | 0000 | 02/20/2015 | 1900 | Incorrect |
04/04/2015 | 0130 | 04/04/2015 | 0500 | Incorrect |
05/15/2015 | 1700 | 05/15/2015 | 1730 | Incorrect |
05/28/2015 | 0330 | 05/28/2015 | 2359 | Incorrect |
06/22/2015 | 1530 | 06/22/2015 | 1700 | Incorrect |
02/24/2015 | 1830 | 02/24/2015 | 2359 | Incorrect |
04/09/2015 | 1430 | 04/09/2015 | 2359 | Incorrect |
05/26/2015 | 1115 | 05/26/2015 | 1400 | Incorrect |
06/28/2015 | 1750 | 06/28/2015 | 1900 | Incorrect |
02/15/2015 | 1130 | 02/15/2015 | 1500 | Incorrect |
04/10/2015 | 0000 | 04/10/2015 | 2000 | Incorrect |
04/03/2015 | 1930 | 04/03/2015 | 2100 | Incorrect |
02/18/2015 | 2100 | 02/18/2015 | 2359 | Incorrect |
02/16/2015 | 1600 | 02/16/2015 | 2200 | Incorrect |
05/18/2015 | 0330 | 05/18/2015 | 2300 | Incorrect |
05/08/2015 | 1800 | 05/08/2015 | 1900 | Incorrect |
06/05/2015 | 0000 | 06/05/2015 | 2030 | Incorrect |
02/15/2015 | 2130 | 02/15/2015 | 2359 | Incorrect |
06/29/2015 | 1630 | 06/29/2015 | 2045 | Incorrect |
Subject: | OLI/SKYRAD/M1 - Spikes in data |
DataStreams: | oliskyrad60sM1.b1 |
Description: | Intermittent periods of PIR2 spiking during the affected time period. |
Measurements: | oliskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/22/2015 | 0000 | 10/25/2015 | 0321 | Incorrect |
Subject: | OLI/GNDRAD/M1 - PIR cable connector failure |
DataStreams: | oligndrad60sM1.b1 |
Description: | During the specified time period upwelling PIR cable/connector problems were being corrected. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/12/2016 | 2058 | 07/03/2016 | 1900 | Suspect |
Subject: | OLI/SKYRAD/M1 - Intermittent spiking |
DataStreams: | oliskyrad60sM1.b1 |
Description: | PIR1 data suspect during the affected time period due to intermittent spiking. PIR2 data is unaffected. |
Measurements: | oliskyrad60sM1.b1:
|