DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D151218.14 | OLI/MET/M1 - Intermittent wind data loss | olimetM1.b1 | Does not affect quality |
D160211.1 | OLI/DL/M1 - Instrument not scanning | olidlfptM1.b1 | Incorrect |
D160520.8 | OLI/DL/DLPROF/M1 - Reprocess: Incorrect serial number | olidlfptM1.b1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/13/2015 | 1840 | 05/11/2016 | 2106 | Does not affect quality |
Subject: | OLI/MET/M1 - Intermittent wind data loss |
DataStreams: | olimetM1.b1 |
Description: | Occasionally the wind speed data are missing and wind direction data drop to 0 due to an ongoing sensor circuitry problem. Only invalid data are missing. |
Measurements: | olimetM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/01/2016 | 1945 | 02/11/2016 | 1625 | Incorrect |
Subject: | OLI/DL/M1 - Instrument not scanning |
DataStreams: | olidlfptM1.b1 |
Description: | The elevation motor on the scanner was stuck at 0deg, thus all stares and scans (i.e. PPI) were done with elevation=0. Also the intensity and backscatter fields contain erroneous negative values during this time period. |
Measurements: | olidlfptM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/28/2014 | 0000 | 05/12/2016 | 1800 | Does not affect quality |
Subject: | OLI/DL/DLPROF/M1 - Reprocess: Incorrect serial number |
DataStreams: | olidlfptM1.b1 |
Description: | The instrument serial number that appears in the netcdf files global attributes is incorrect. The correct serial number is 0514-83. |
Measurements: | olidlfptM1.b1:
|