DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160401.9 | AWR/MWR/M1 - Thermal instability | awrmwrlosM1.b1 | Suspect |
D160426.1 | AWR/MWR/M1 - Reprocess: Incorrect retrievals | awrmwrlosM1.b1 | Incorrect |
D160621.2 | AWR/MWR/M1 - Instrument thermally unstable | awrmwrlosM1.b1 | Incorrect |
D160715.6 | AWR/MWR/M1 - Thermal instability | awrmwrlosM1.b1 | Suspect |
D160822.2 | AWR/MWR/M1 -Reprocess: Incorrect calibration | awrmwrlosM1.b1 | Incorrect |
D160922.1 | AWR/MWR/M1 - Missing data | awrmwrlosM1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/20/2016 | 0000 | 03/21/2016 | 2359 | Suspect |
Subject: | AWR/MWR/M1 - Thermal instability |
DataStreams: | awrmwrlosM1.b1 |
Description: | The radiometer had problems maintaining thermal control. The radiometer set point was lowered 10 degrees to 318 K on 3/21/2016 |
Measurements: | awrmwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/01/2016 | 0000 | 01/13/2017 | 2359 | Incorrect |
Subject: | AWR/MWR/M1 - Reprocess: Incorrect retrievals |
DataStreams: | awrmwrlosM1.b1 |
Description: | PWV and LWP retrievals are incorrect. Data will be reprocessed at the end of the deployment using coefficients derived from the radiosondes launched at the site. |
Measurements: | awrmwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/12/2016 | 0300 | 06/14/2016 | 0000 | Incorrect |
Subject: | AWR/MWR/M1 - Instrument thermally unstable |
DataStreams: | awrmwrlosM1.b1 |
Description: | The radiometer was thermally unstable. |
Measurements: | awrmwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/04/2016 | 1420 | 07/05/2016 | 2310 | Suspect |
Subject: | AWR/MWR/M1 - Thermal instability |
DataStreams: | awrmwrlosM1.b1 |
Description: | The radiometer was thermally unstable. |
Measurements: | awrmwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/02/2016 | 0000 | 08/11/2016 | 0400 | Incorrect |
Subject: | AWR/MWR/M1 -Reprocess: Incorrect calibration |
DataStreams: | awrmwrlosM1.b1 |
Description: | After the temperature set point was changed the calibration changed. New calibration coefficients were derived on 8/11. |
Measurements: | awrmwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/02/2016 | 0400 | 08/06/2016 | 0444 | Missing |
Subject: | AWR/MWR/M1 - Missing data |
DataStreams: | awrmwrlosM1.b1 |
Description: | Data are missing |
Measurements: | awrmwrlosM1.b1:
|