DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140625.3 | ENA/MWRP/C1 - Bad calibration | enamwrpC1.b1 | Incorrect |
D140724.3 | ENA/MWRP/C1 - Reprocess: Bias in brightness temperatures | enamwrpC1.b1 | Suspect |
D140905.2 | ENA/MWRP/C1 - Incorrect IRT readings | enamwrpC1.b1 | Incorrect |
D141212.1 | ENA/MWRP/C1 - Reprocessing Notice (RID 1261) | enamwrpC1.b1 | Does not affect quality |
D150219.2 | ENA/MWRP/C1 - Reprocess: Calibration drift | enamwrpC1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/19/2014 | 0000 | 06/04/2014 | 1459 | Incorrect |
Subject: | ENA/MWRP/C1 - Bad calibration |
DataStreams: | enamwrpC1.b1 |
Description: | The radiometer was poorly calibrated following a power outage at the site. |
Measurements: | enamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/04/2014 | 1400 | 07/24/2014 | 1934 | Suspect |
Subject: | ENA/MWRP/C1 - Reprocess: Bias in brightness temperatures |
DataStreams: | enamwrpC1.b1 |
Description: | The brightness temperatures between 22.234 and 26.234 GHz show a bias of 1.7 to 3.6 degrees K. Data needs to be recalibrated. As a consequence the retrieved precipitable water vapor has a positive bias. |
Measurements: | enamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/20/2014 | 2230 | 08/25/2014 | 1330 | Incorrect |
Subject: | ENA/MWRP/C1 - Incorrect IRT readings |
DataStreams: | enamwrpC1.b1 |
Description: | The infrared temperature readings are higher than those from the IRT and skyrad instruments. |
Measurements: | enamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/29/2014 | 0000 | 11/24/2014 | 1538 | Does not affect quality |
Subject: | ENA/MWRP/C1 - Reprocessing Notice (RID 1261) |
DataStreams: | enamwrpC1.b1 |
Description: | See below for details of reprocessing that was completed on 20150309. Original DQR category: 3 (Yellow, Suspect). ----------------------------------------------------------- The radiometer's calibration appears unstable. The brightness temperatures are warmer than expected. Data need to be recalibrated. |
Measurements: | enamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/26/2014 | 0000 | 10/08/2015 | 1600 | Incorrect |
Subject: | ENA/MWRP/C1 - Reprocess: Calibration drift |
DataStreams: | enamwrpC1.b1 |
Description: | The data are incorrect mostly because of fast drift in the calibration. Brightness temperatures become negative after 9/7/2015. |
Measurements: | enamwrpC1.b1:
|