DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D150306.2 | SGP/CCN/C1 - Instrument under counting at lowest super saturation value | sgpaosccnavgC1.c2 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/11/2013 | 1926 | 04/08/2015 | 1730 | Suspect |
Subject: | SGP/CCN/C1 - Instrument under counting at lowest super saturation value |
DataStreams: | sgpaosccnavgC1.c2 |
Description: | The slow temperature decay of the OPC induced droplet evaporation and caused an under counting of the droplets at the lowest %ss, 0.15. |
Measurements: | sgpaosccnavgC1.c2:
|