DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D130717.4 | SGP/KAZR/C1 - Transmitter failure | sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0 | Incorrect |
D131120.4 | SGP/KAZR/C1 - Reprocessing Notice (RID 1117) | sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/29/2013 | 1900 | 07/08/2013 | 1900 | Incorrect |
Subject: | SGP/KAZR/C1 - Transmitter failure |
DataStreams: | sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0 |
Description: | KAZR TWTA failure. Power started dropping on 2013.04.30 and by 2013.05.09 it was down to approximately 4 watts. Data are not useable during this period. |
Measurements: | sgpkazrspeccmaskgexpolC1.a0:
sgpkazrspeccmaskgecopolC1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/18/2011 | 0000 | 03/15/2014 | 2359 | Does not affect quality |
Subject: | SGP/KAZR/C1 - Reprocessing Notice (RID 1117) |
DataStreams: | sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0 |
Description: | See below for details of reprocessing that was completed on 20150626. Original DQR category: 4 (Red, Incorrect). ----------------------------------------------------------- The wrong calibrations were used due to a problem with the software. Mentor discovered the reflectivity between long and short pulse did not agree with each other after analyzing data. Data will be reprocessed. |
Measurements: | sgpkazrspeccmaskgexpolC1.a0:
sgpkazrspeccmaskgecopolC1.a0:
|