DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D131120.4 | SGP/KAZR/C1 - Reprocessing Notice (RID 1117) | sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0 | Does not affect quality |
D140214.3 | SGP/KAZR/C1 - Disk error/failure | sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/18/2011 | 0000 | 03/15/2014 | 2359 | Does not affect quality |
Subject: | SGP/KAZR/C1 - Reprocessing Notice (RID 1117) |
DataStreams: | sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0 |
Description: | See below for details of reprocessing that was completed on 20150626. Original DQR category: 4 (Red, Incorrect). ----------------------------------------------------------- The wrong calibrations were used due to a problem with the software. Mentor discovered the reflectivity between long and short pulse did not agree with each other after analyzing data. Data will be reprocessed. |
Measurements: | sgpkazrspeccmaskgexpolC1.a0:
sgpkazrspeccmaskgecopolC1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/01/2013 | 0500 | 01/14/2014 | 2022 | Missing |
Subject: | SGP/KAZR/C1 - Disk error/failure |
DataStreams: | sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0 |
Description: | Sectors on the KAZR's spectra storage partition are causing some data files to be corrupted. This has caused collections to fail, the disk to fill, and subsequent data loss. |
Measurements: | sgpkazrspeccmaskgexpolC1.a0:
sgpkazrspeccmaskgecopolC1.a0:
|