DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160801.4 | ASI/SONDE/S1 - Reprocess: Incorrect lat and lon | asisondewnpnS1.b1 | Incorrect |
D160810.2 | ASI/GNDRAD/M1 - Corrosion of wires and pins | asigndrad60sM1.b1 | Incorrect |
D160822.4 | ASI/CCN/M1 - Incorrect flows | asiaosccn200M1.a1 | Incorrect |
D160902.1 | ASI/MFRSR/M1 - Heater Problem | asimfrsrM1.a0, asimfrsrM1.b1 | Incorrect |
D160907.1 | ASI/CPC/M1 - Collapsed internal tube | asiaoscpcuM1.b1 | Incorrect |
D160907.12 | ASI/CPC/M1 - Instrument testing | asiaoscpcfM1.a1 | Incorrect |
D160907.13 | ASI/CPC/M1 - Instrument maintence | asiaoscpcfM1.a1 | Incorrect |
D160907.14 | ASI/SMPS/M1 - Instrument maintence | asiaossmpsM1.a1 | Incorrect |
D160914.1 | ASI/CEIL/M1 - data communication | asiceilM1.b1 | Suspect |
D161003.1 | ASI/SMPS/M1 - Small leak in the SMPS CPC | asiaossmpsM1.a1 | Suspect |
D161010.1 | ASI/ACSM/M1 - Critical orifice clogged | asiaosacsmM1.a1 | Suspect |
D161018.2 | ASI/SMPS/M1/M1 - Instrument computer failure | asiaossmpsM1.a1 | Missing |
D161021.3 | ASI/SMPS/M1 - Collection | asiaossmpsM1.a1 | Does not affect quality |
D161021.4 | ASI/SMPS/M1 - Different number of particle size bins | asiaossmpsM1.a1 | Does not affect quality |
D161024.2 | ASI/CPC/M1 - Collection | asiaoscpcfM1.a1, asiaoscpcuM1.b1 | Does not affect quality |
D161108.1 | ASI/PSAP/M1 - Transmission less than vendor minimum | asiaospsap3wM1.a1 | Incorrect |
D161108.2 | ASI/PSAP/M1 - Invalid data for transmission less than 0.7 | asiaospsap3wM1.a1 | Incorrect |
D161108.4 | ASI/PSAP/M1 - Periods where transmission < 0.7 for > 1 hour or pump turned off | asiaospsap3wM1.a1 | Incorrect |
D161109.2 | ASI/OZONE/M1 - Not sampling outside air | asiaosozoneM1.a1 | Incorrect |
D161109.4 | ASI/TRACEGAS/M1 - Sampling cabin air. | asiaosso2M1.a1 | Incorrect |
D161109.6 | ASI/IRT/M1 - no communication | asiirtM1.b1, asiirt200msM1.a1 | Missing |
D161116.1 | ASI/CCN/M1 - Concentration mismatch | asiaosccn200M1.a1 | Suspect |
D161116.5 | ASI/CCN/M1 - Instrument troubleshooting | asiaosccn200M1.a1 | Incorrect |
D161117.1 | ASI/ACSM/M1 - Critical orifice clog | asiaosacsmM1.a1 | Missing |
D161128.5 | ASI/CPC/M1 - Instrument | asiaoscpcfM1.a1, asiaoscpcuM1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/01/2016 | 0229 | 08/01/2016 | 0323 | Incorrect |
Subject: | ASI/SONDE/S1 - Reprocess: Incorrect lat and lon |
DataStreams: | asisondewnpnS1.b1 |
Description: | Somehow the initial Lat/Lon settings were corrupted. Data has been reprocessed and needs to be collected and ingested. |
Measurements: | asisondewnpnS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/04/2016 | 1945 | 08/08/2016 | 1640 | Incorrect |
Subject: | ASI/GNDRAD/M1 - Corrosion of wires and pins |
DataStreams: | asigndrad60sM1.b1 |
Description: | Data incorrect during the affected time period from corrosion of wires and pins. |
Measurements: | asigndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/20/2016 | 0200 | 08/20/2016 | 1530 | Incorrect |
Subject: | ASI/CCN/M1 - Incorrect flows |
DataStreams: | asiaosccn200M1.a1 |
Description: | Instrument flows incorrect due to water in the flow lines. |
Measurements: | asiaosccn200M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/21/2016 | 2304 | 08/23/2016 | 0920 | Incorrect |
Subject: | ASI/MFRSR/M1 - Heater Problem |
DataStreams: | asimfrsrM1.a0, asimfrsrM1.b1 |
Description: | The heater board failed on 20160821 and was replaced on 20160823. Do not use data while the heater is not functioning. |
Measurements: | asimfrsrM1.a0:
asimfrsrM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/01/2016 | 1345 | 09/01/2016 | 1422 | Incorrect |
Subject: | ASI/CPC/M1 - Collapsed internal tube |
DataStreams: | asiaoscpcuM1.b1 |
Description: | During this time, operator and mentor were troubleshooting a collapsed internal tube on the 3776 CPC. |
Measurements: | asiaoscpcuM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/03/2016 | 1314 | 09/03/2016 | 1327 | Incorrect |
Subject: | ASI/CPC/M1 - Instrument testing |
DataStreams: | asiaoscpcfM1.a1 |
Description: | During this time tests were being run with a HEPA filter on the inlet. Data is not representative of ambient aerosol concentrations |
Measurements: | asiaoscpcfM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/06/2016 | 1004 | 09/06/2016 | 1029 | Incorrect |
09/06/2016 | 1815 | 09/06/2016 | 1819 | Incorrect |
Subject: | ASI/CPC/M1 - Instrument maintence |
DataStreams: | asiaoscpcfM1.a1 |
Description: | During these intervals, connections were being removed from the CPC to prepare for it to take the place of the SMPS CPC. |
Measurements: | asiaoscpcfM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/06/2016 | 1030 | 09/06/2016 | 1154 | Incorrect |
08/29/2016 | 1527 | 08/29/2016 | 1549 | Incorrect |
09/06/2016 | 1725 | 09/06/2016 | 1744 | Incorrect |
08/29/2016 | 1605 | 08/31/2016 | 1719 | Incorrect |
09/02/2016 | 1305 | 09/02/2016 | 1334 | Incorrect |
09/06/2016 | 1800 | 09/06/2016 | 1824 | Incorrect |
09/01/2016 | 1309 | 09/01/2016 | 1814 | Incorrect |
08/29/2016 | 1255 | 08/29/2016 | 1344 | Incorrect |
Subject: | ASI/SMPS/M1 - Instrument maintence |
DataStreams: | asiaossmpsM1.a1 |
Description: | During these time periods, a HEPA filter was connected to different portions of the system for investigation of leaks. |
Measurements: | asiaossmpsM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/12/2016 | 2000 | 08/14/2016 | 0459 | Suspect |
Subject: | ASI/CEIL/M1 - data communication |
DataStreams: | asiceilM1.b1 |
Description: | The data were intermittent, especially on 7/19/2016, due to poor serial communication. The cable was repaired on 7/21 and the serial device server and instrument were power-cycled on 7/28. |
Measurements: | asiceilM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/29/2016 | 1345 | 08/29/2016 | 1526 | Suspect |
08/29/2016 | 1550 | 08/29/2016 | 1604 | Suspect |
09/03/2016 | 1230 | 09/03/2016 | 1304 | Suspect |
05/19/2016 | 0000 | 08/29/2016 | 1254 | Suspect |
09/02/2016 | 1335 | 09/02/2016 | 1419 | Suspect |
Subject: | ASI/SMPS/M1 - Small leak in the SMPS CPC |
DataStreams: | asiaossmpsM1.a1 |
Description: | Periodic spikes were seen in data due to a small leak in the SMPS CPC. These spikes are identifiable and may be corrected in future data products. |
Measurements: | asiaossmpsM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2016 | 1132 | 10/04/2016 | 1623 | Suspect |
Subject: | ASI/ACSM/M1 - Critical orifice clogged |
DataStreams: | asiaosacsmM1.a1 |
Description: | The critical orifice was clogged during the time period specified. The orifice was cleaned and the instrument placed back in service. |
Measurements: | asiaosacsmM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/07/2016 | 2000 | 10/14/2016 | 1054 | Missing |
Subject: | ASI/SMPS/M1/M1 - Instrument computer failure |
DataStreams: | asiaossmpsM1.a1 |
Description: | The computer running the SMPS failed. Some data was lost in the interim during the replacement of the computer and some data may be recoverable from the failed computer. |
Measurements: | asiaossmpsM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/06/2016 | 1030 | 09/06/2016 | 1809 | Does not affect quality |
Subject: | ASI/SMPS/M1 - Collection |
DataStreams: | asiaossmpsM1.a1 |
Description: | The CPC associated with the CPCF measurement was sampling in place of the CPC associated with the SMPS measurement. They are both the same model number. |
Measurements: | asiaossmpsM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/29/2016 | 1345 | 08/29/2016 | 1549 | Does not affect quality |
Subject: | ASI/SMPS/M1 - Different number of particle size bins |
DataStreams: | asiaossmpsM1.a1 |
Description: | The data was exported in a different format. There will be a different number of particle size bins during this period. |
Measurements: | asiaossmpsM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/06/2016 | 1030 | 09/06/2016 | 1815 | Does not affect quality |
Subject: | ASI/CPC/M1 - Collection |
DataStreams: | asiaoscpcfM1.a1, asiaoscpcuM1.b1 |
Description: | The CPC associated with the SMPS was sampling in place of the CPC associated with this CPCF measurement. The CPC's are the same model number |
Measurements: | asiaoscpcfM1.a1:
asiaoscpcuM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/15/2016 | 2256 | 08/16/2016 | 0559 | Incorrect |
08/09/2016 | 2252 | 08/10/2016 | 0925 | Incorrect |
08/30/2016 | 2118 | 08/31/2016 | 0654 | Incorrect |
08/14/2016 | 0827 | 08/14/2016 | 1040 | Incorrect |
08/11/2016 | 2213 | 08/12/2016 | 0917 | Incorrect |
08/08/2016 | 2323 | 08/09/2016 | 0820 | Incorrect |
08/14/2016 | 2147 | 08/15/2016 | 0944 | Incorrect |
08/12/2016 | 1757 | 08/12/2016 | 1905 | Incorrect |
08/19/2016 | 0544 | 08/19/2016 | 0938 | Incorrect |
08/18/2016 | 0500 | 08/18/2016 | 0940 | Incorrect |
08/15/2016 | 1235 | 08/15/2016 | 1358 | Incorrect |
08/28/2016 | 0633 | 08/28/2016 | 1209 | Incorrect |
08/08/2016 | 1403 | 08/08/2016 | 1601 | Incorrect |
08/12/2016 | 2148 | 08/13/2016 | 1052 | Incorrect |
08/08/2016 | 0153 | 08/08/2016 | 0912 | Incorrect |
08/13/2016 | 2046 | 08/14/2016 | 0610 | Incorrect |
08/29/2016 | 2249 | 08/30/2016 | 0753 | Incorrect |
08/14/2016 | 1307 | 08/14/2016 | 1522 | Incorrect |
08/13/2016 | 1307 | 08/13/2016 | 1436 | Incorrect |
08/10/2016 | 2359 | 08/11/2016 | 0915 | Incorrect |
08/07/2016 | 0724 | 08/07/2016 | 1428 | Incorrect |
08/31/2016 | 1446 | 08/31/2016 | 1759 | Incorrect |
08/29/2016 | 0233 | 08/29/2016 | 0905 | Incorrect |
Subject: | ASI/PSAP/M1 - Transmission less than vendor minimum |
DataStreams: | asiaospsap3wM1.a1 |
Description: | Periods where Transmission (Tr) < 0.7 for > 1 h due to filter loading have been isolated. Vendor specifies Tr must be > 0.7 for valid readings. All periods of Tr < 0.7 have been isolated. |
Measurements: | asiaospsap3wM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/25/2016 | 1311 | 09/25/2016 | 1428 | Incorrect |
09/24/2016 | 0653 | 09/24/2016 | 1119 | Incorrect |
09/01/2016 | 0000 | 09/01/2016 | 0609 | Incorrect |
Subject: | ASI/PSAP/M1 - Invalid data for transmission less than 0.7 |
DataStreams: | asiaospsap3wM1.a1 |
Description: | Data Gaps > 1h due to Tr<0.7, pump turned off, etc. |
Measurements: | asiaospsap3wM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/21/2016 | 1358 | 10/21/2016 | 1917 | Incorrect |
10/25/2016 | 1611 | 10/25/2016 | 1816 | Incorrect |
10/24/2016 | 1504 | 10/24/2016 | 1707 | Incorrect |
10/03/2016 | 0216 | 10/03/2016 | 1016 | Incorrect |
10/14/2016 | 0742 | 10/14/2016 | 1007 | Incorrect |
Subject: | ASI/PSAP/M1 - Periods where transmission < 0.7 for > 1 hour or pump turned off |
DataStreams: | asiaospsap3wM1.a1 |
Description: | Vendor specifies Tr > 0.7 for reliable measurements. |
Measurements: | asiaospsap3wM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/25/2016 | 2046 | 10/26/2016 | 1135 | Incorrect |
Subject: | ASI/OZONE/M1 - Not sampling outside air |
DataStreams: | asiaosozoneM1.a1 |
Description: | Inlet turned off (or the blower turned off). So sampling was not on outside air. |
Measurements: | asiaosozoneM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/25/2016 | 2046 | 10/26/2016 | 1135 | Incorrect |
Subject: | ASI/TRACEGAS/M1 - Sampling cabin air. |
DataStreams: | asiaosso2M1.a1 |
Description: | Inlet blower turned off so instrument sampling cabin air. |
Measurements: | asiaosso2M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/08/2016 | 2021 | 11/09/2016 | 1412 | Missing |
10/25/2016 | 1700 | 10/27/2016 | 1832 | Missing |
Subject: | ASI/IRT/M1 - no communication |
DataStreams: | asiirtM1.b1, asiirt200msM1.a1 |
Description: | There was no serial communication with the IRT. |
Measurements: | asiirt200msM1.a1:
asiirtM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/20/2016 | 1530 | 09/16/2016 | 1500 | Suspect |
Subject: | ASI/CCN/M1 - Concentration mismatch |
DataStreams: | asiaosccn200M1.a1 |
Description: | Concentration mismatch between the CCN columns at the same supersaturation set point after water was removed from one of the sheath air lines. Not clear which column was affected. |
Measurements: | asiaosccn200M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/16/2016 | 1500 | 11/16/2016 | 1700 | Incorrect |
Subject: | ASI/CCN/M1 - Instrument troubleshooting |
DataStreams: | asiaosccn200M1.a1 |
Description: | Troubleshooting instrument flow issues. No valid data. |
Measurements: | asiaosccn200M1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/10/2016 | 2059 | 11/11/2016 | 1452 | Missing |
Subject: | ASI/ACSM/M1 - Critical orifice clog |
DataStreams: | asiaosacsmM1.a1 |
Description: | There are no data for the time period specified due to a critical orifice clog. |
Measurements: | asiaosacsmM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/21/2016 | 1540 | 11/21/2016 | 1705 | Incorrect |
Subject: | ASI/CPC/M1 - Instrument |
DataStreams: | asiaoscpcfM1.a1, asiaoscpcuM1.b1 |
Description: | REQUIRED |
Measurements: | asiaoscpcfM1.a1:
asiaoscpcuM1.b1:
|