DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160610.2 | ASI/ACSM/M1 - Data averaging period changed | asiaosacsmM1.a1 | Does not affect quality |
D160617.11 | ASI/ACSM/M1 - Noisy data from low inlet pressure | asiaosacsmM1.a1 | Suspect |
D160621.4 | ASI/AOS/M1 - Manifold blower not working | asiaosacsmM1.a1 | Incorrect |
D161010.1 | ASI/ACSM/M1 - Critical orifice clogged | asiaosacsmM1.a1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/25/2016 | 0000 | 06/08/2016 | 1300 | Does not affect quality |
Subject: | ASI/ACSM/M1 - Data averaging period changed |
DataStreams: | asiaosacsmM1.a1 |
Description: | The data averaging interval was changed from 10 minutes to 30 minutes to conform to standard ACSM collection protocols. |
Measurements: | asiaosacsmM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/03/2016 | 1530 | 06/13/2016 | 1221 | Suspect |
Subject: | ASI/ACSM/M1 - Noisy data from low inlet pressure |
DataStreams: | asiaosacsmM1.a1 |
Description: | The inlet pressure is low as the result of a clogged critical orifice. Data are corrected for inlet pressure but the signal is low because of very light aerosol loading and are noisy. |
Measurements: | asiaosacsmM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2016 | 0042 | 06/16/2016 | 1345 | Incorrect |
Subject: | ASI/AOS/M1 - Manifold blower not working |
DataStreams: | asiaosacsmM1.a1 |
Description: | Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic. |
Measurements: | asiaosacsmM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2016 | 1132 | 10/04/2016 | 1623 | Suspect |
Subject: | ASI/ACSM/M1 - Critical orifice clogged |
DataStreams: | asiaosacsmM1.a1 |
Description: | The critical orifice was clogged during the time period specified. The orifice was cleaned and the instrument placed back in service. |
Measurements: | asiaosacsmM1.a1:
|