DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160621.4 | ASI/AOS/M1 - Manifold blower not working | asiaoscpcuM1.b1 | Incorrect |
D160907.1 | ASI/CPC/M1 - Collapsed internal tube | asiaoscpcuM1.b1 | Incorrect |
D161024.2 | ASI/CPC/M1 - Collection | asiaoscpcuM1.b1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2016 | 0042 | 06/16/2016 | 1345 | Incorrect |
Subject: | ASI/AOS/M1 - Manifold blower not working |
DataStreams: | asiaoscpcuM1.b1 |
Description: | Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic. |
Measurements: | asiaoscpcuM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/01/2016 | 1345 | 09/01/2016 | 1422 | Incorrect |
Subject: | ASI/CPC/M1 - Collapsed internal tube |
DataStreams: | asiaoscpcuM1.b1 |
Description: | During this time, operator and mentor were troubleshooting a collapsed internal tube on the 3776 CPC. |
Measurements: | asiaoscpcuM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/06/2016 | 1030 | 09/06/2016 | 1815 | Does not affect quality |
Subject: | ASI/CPC/M1 - Collection |
DataStreams: | asiaoscpcuM1.b1 |
Description: | The CPC associated with the SMPS was sampling in place of the CPC associated with this CPCF measurement. The CPC's are the same model number |
Measurements: | asiaoscpcuM1.b1:
|