DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D131018.1 | ENA/CPC/C1 - Butanol Exhausted | enaaoscpcfC1.a1 | Missing |
D140618.2 | ENA/CPC/NEPH/PSAP/C1 - Impactor not installed | enaaoscpcfC1.a1 | Does not affect quality |
D140926.1 | ENA/CPC/C1 - Computer Failure | enaaoscpcfC1.a1 | Missing |
D150506.2 | ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit | enaaoscpcfC1.a1 | Incorrect |
D150617.1 | ENA/CPC/C1 - Instrument problem | enaaoscpcfC1.a1 | Incorrect |
D150618.1 | ENA/CPC/C1 - Instrument problem | enaaoscpcfC1.a1 | Incorrect |
D150618.2 | ENA/CPC/C1 - Instrument Not Sampling due to repair and calibration | enaaoscpcfC1.a1 | Missing |
D150618.3 | ENA/CPC/C1 - Data reprocessed to account for vacuum port too low | enaaoscpcfC1.a1 | Does not affect quality |
D150701.30 | ENA/CPC/C1 - Reprocess: reprocessing from raw data in preparation for not-yet-released b1 level | enaaoscpcfC1.a1 | Does not affect quality |
D151112.5 | ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor not switching | enaaoscpcfC1.a1 | Does not affect quality |
D151112.6 | ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor size switch not connected | enaaoscpcfC1.a1 | Does not affect quality |
D160427.7 | ENA/C1 - Reprocess: Slightly incorrect location fields | enaaoscpcfC1.a1 | Suspect |
D160701.3 | ENA/AOS/C1 - incorrect data | enaaoscpcfC1.a1 | Incorrect |
D160907.10 | ENA/CPC/C1 - New pump slightly under pressure | enaaoscpcfC1.a1 | Suspect |
D160909.20 | ENA/AOS/C1 - Flow problem affecting concentrations | enaaoscpcfC1.a1 | Incorrect |
D161025.6 | ENA/AOS/C1 - Instrument mentors working at the site | enaaoscpcfC1.a1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/10/2013 | 0600 | 10/18/2013 | 0820 | Missing |
Subject: | ENA/CPC/C1 - Butanol Exhausted |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Butanol was not available for startup. Operators were able get instrument started with small amount of butanol in the HTDMA. Data recovery not possilble. Resumed operation with refill of butanol. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/01/2013 | 0000 | 02/19/2014 | 1351 | Does not affect quality |
Subject: | ENA/CPC/NEPH/PSAP/C1 - Impactor not installed |
DataStreams: | enaaoscpcfC1.a1 |
Description: | No impactor installed. All data is the same as when impactor at 10 um setting. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/21/2014 | 1200 | 09/23/2014 | 1049 | Missing |
Subject: | ENA/CPC/C1 - Computer Failure |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Data were lost due to a failed CPU unit |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/22/2015 | 1700 | 04/27/2015 | 1700 | Incorrect |
Subject: | ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit |
DataStreams: | enaaoscpcfC1.a1 |
Description: | There was visit of 7 AOS instrument mentors to the ENA AOS during this period of instrument repair, calibration and modification. Instruments were occasionally off, or with no sample flow, or receiving calibration samples, etc. There was no productive sampling of ambient air during this period other than for testing. No data should be used for this period as ambient samples. Note: Two datastream are missing. The CAPS Extinction Monitor and the Greenhouse gas instrument. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/25/2014 | 1505 | 11/12/2014 | 2359 | Incorrect |
Subject: | ENA/CPC/C1 - Instrument problem |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Vacuum was lost very briefly to the instrument for an unknown reason. The instrument then showed evidence of flooding by an increase in the frequency of the butanol fill cycles, increase in nozzle pressure, and decrease in orifice pressure. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/06/2015 | 1433 | 06/06/2015 | 1504 | Incorrect |
Subject: | ENA/CPC/C1 - Instrument problem |
DataStreams: | enaaoscpcfC1.a1 |
Description: | It was found that the wrong vacuum source was connected to the vacuum port on the CPC and didn't provide enough pull for the instrument to run correctly. While the problem was being rectified, there was a period where the CPC was sampling from the instrument cabin and not from the ambient. Therefore data during this time is not representative of the ambient concentration of aerosol in the atmosphere and should not be used as such. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/12/2014 | 1625 | 06/05/2015 | 1438 | Missing |
Subject: | ENA/CPC/C1 - Instrument Not Sampling due to repair and calibration |
DataStreams: | enaaoscpcfC1.a1 |
Description: | During this period the CPC was shipped back to the manufacturer to mitigate damage due to flooding. The instrument was found to be flooded. Repair and calibration was performed. Most of the down time was related to delays in shipment due to customs on Terceira Island. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/05/2015 | 1717 | 06/06/2015 | 1432 | Does not affect quality |
Subject: | ENA/CPC/C1 - Data reprocessed to account for vacuum port too low |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Recorded calibrations relating the effect on flow of reduced nozzle and orifice pressure was used to correct the data. When the CPC was put back in operation, the external vacuum port was connected to a vacuum line, which didn't provide enough pull. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2013 | 0000 | 10/01/2015 | 2359 | Does not affect quality |
Subject: | ENA/CPC/C1 - Reprocess: reprocessing from raw data in preparation for not-yet-released b1 |
DataStreams: | enaaoscpcfC1.a1 |
Description: | End to end reprocessing to generate new a1 datastreams is needed so that when the harmonization software is released to production, we will be ready with the a1 datastreams that are input to the subsequent processing software. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/27/2015 | 1052 | 03/30/2015 | 1800 | Does not affect quality |
Subject: | ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor not switching |
DataStreams: | enaaoscpcfC1.a1 |
Description: | AOS tech was working in the AOS and accidentally unplugged the impactor power supply. This left the impactor in the 10 micron state. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/19/2014 | 1351 | 04/26/2015 | 1016 | Does not affect quality |
Subject: | ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor size switch not connected |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Impactor installed and plumbed, HOWEVER, the impactor's 'size read' switch NOT connected. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/29/2013 | 0941 | 04/30/2016 | 2359 | Suspect |
Subject: | ENA/C1 - Reprocess: Slightly incorrect location fields |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Some of the instruments are being processed with the previous "GRW-Azores" lat/lon/alt instead of the new "ENA-Azores" lat/lon/alt. The differences are minor, but will be corrected. The respective GRW and ENA lat, lon, and alt values are as follows: GRW lat = 39.0911 ; ENA lat = 39.0916 ; GRW lon = -28.0297 ; ENA lon = -28.0257 ; GRW alt = 26 ; ENA alt = 30.48 ; |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/20/2016 | 1820 | 06/21/2016 | 0835 | Incorrect |
Subject: | ENA/AOS/C1 - incorrect data |
DataStreams: | enaaoscpcfC1.a1 |
Description: | During a measurement of the inlet flow rate, the cpcf inlet became plugged and the instrument was flooded with butanol. After removing the butanol and allowing the system to dry, the cpc returned to normal operation. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/24/2016 | 0900 | 08/29/2016 | 0831 | Suspect |
Subject: | ENA/CPC/C1 - New pump slightly under pressure |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Due to a pump failure, a different pump was used on the system. The orifice pressure was lower with this additional pump and the system may have been slightly under the pressure required to maintain critical flow on the CPC's critical orifice. Flow rate measurements during this time show very little change in the output flow rate. |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/08/2016 | 0035 | 09/08/2016 | 1501 | Incorrect |
Subject: | ENA/AOS/C1 - Flow problem affecting concentrations |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Due to a flow problem with another instrument on the same sample line, the concentrations reported by the CPC are not representative of the atmospheric concentration of particles |
Measurements: | enaaoscpcfC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/26/2016 | 0830 | 10/01/2016 | 0100 | Suspect |
Subject: | ENA/AOS/C1 - Instrument mentors working at the site |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Instrument mentors working at the site. Sample lines and electrical connections being reconfigured. Instruments that had no work done on them may still be affected. |
Measurements: | enaaoscpcfC1.a1:
|