DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160114.6 | ENA/GHG/C1 - System probationary period | enaaosghgC1.a1 | Incorrect |
D160323.4 | ENA/GHG/C1 - Reprocessed and Corrected: Valve position fields incorrect type | enaaosghgC1.a1 | Does not affect quality |
D161014.8 | ENA/GHG/C1 - Reprocess: addition of height global attribute | enaaosghgC1.a1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/22/2015 | 1028 | 10/16/2015 | 2359 | Incorrect |
Subject: | ENA/GHG/C1 - System probationary period |
DataStreams: | enaaosghgC1.a1 |
Description: | System probationary period, may not be useful for science data due to non-optimal functioning of the measurement system. Additional information: 2015-04-22 10:28:00,2015-07-08 09:59:00 - Chiller failure, pressure controller failure during shipment for field deployment 2015-07-08 10:00:00,2015-10-12 00:00:00 - Chiller repaired, pressure controller failure 2015-10-12 00:01:00,2015-10-16 23:59:59 - Pressure controller replacement and system testing. Related to BCR-02211 |
Measurements: | enaaosghgC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/31/2015 | 2256 | 04/19/2016 | 2359 | Does not affect quality |
Subject: | ENA/GHG/C1 - Reprocessed and Corrected: Valve position fields incorrect type |
DataStreams: | enaaosghgC1.a1 |
Description: | DATA REPROCESSED and CORRECTED: The raw data variable mp_valve_position was being read as an integer instead of a float. The aos_ingest logic was corrected and the data were run through the corrected ingest to fix the issue. The data were reprocessed in July 2016. Please retrieve the reprocessed data for use in your analysis. You may find these data using the ARM Discovery tool at the following link: http://www.archive.arm.gov/discovery/#v/results/s/finst::ghg/fsite::ena/ffac::ena.C1 ORIGINAL DQR CATEGORY: Red( Incorrect) REPROCESSED and CORRECTED DATA ARCHIVED:2016-07 REPROCESS IDENTIFICATION NUMBER: RID 1453 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- Raw data variable MP_VALVE_POSITION needs to be read as a float and not integer. |
Measurements: | enaaosghgC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/22/2015 | 0000 | 09/15/2016 | 2359 | Does not affect quality |
Subject: | ENA/GHG/C1 - Reprocess: addition of height global attribute |
DataStreams: | enaaosghgC1.a1 |
Description: | The mentor has requested the addition of a 'height' global attribute to record the inlet height. The inlet for the ENA AOSGHG has always been at a height of 10 meters, but the datastreams did not have this new global attribute. Both the aosghg.a1 and aosghgaux.a1 datastreams will be updated to add this new global attribute. The mentor has requested that the entire dataset be reprocessed to make this addition to the global attributes. We will be releasing dowstream aosghg.b0 and aosghg.b1 processing that will incorporate this new global attribute from the a1 input datastreams. |
Measurements: | enaaosghgC1.a1:
|