DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D120827.3 | SGP/ACSM/C1 - Water in inlet sample line | sgpaosacsmC1.b1 | Incorrect |
D121211.5 | SGP/ACSM/C1 - Pump failure | sgpaosacsmC1.b1 | Incorrect |
D130108.1 | SGP/ACSM/C1 - Pump failure | sgpaosacsmC1.b1 | Incorrect |
D130108.2 | SGP/ACSM/C1 - Sample flow restricted | sgpaosacsmC1.b1 | Suspect |
D130201.2 | SGP/ACSM/C1 - Missing data | sgpaosacsmC1.b1 | Missing |
D140417.1 | SGP/ACSM/C1 - Missing data | sgpaosacsmC1.b1 | Missing |
D140418.1 | SGP/ACSM/C1 - Data not available and unrecoverable | sgpaosacsmC1.b1 | Missing |
D140627.1 | SGP/ACSM/C1 - Windows 7 upgrade and turbo molecular pump failure. | sgpaosacsmC1.b1 | Missing |
D141009.2 | SGP/ACSM/C1 - Reprocessing Notice (RID 1237) | sgpaosacsmC1.b1 | Does not affect quality |
D141107.1 | SGP/ACSM/C1 - computer software problem | sgpaosacsmC1.b1 | Missing |
D141107.5 | SGP/ACSM/C1 - Pump Failure | sgpaosacsmC1.b1 | Missing |
D141205.1 | SGP/ACSM/C1 - Unexpected instrument shutdown | sgpaosacsmC1.b1 | Incorrect |
D150210.1 | SGP/ACSM/C1 - Clogged critical orifice | sgpaosacsmC1.b1 | Incorrect |
D150223.2 | SGP/ACSM/C1 - Critical orifice clogged | sgpaosacsmC1.b1 | Missing |
D150410.1 | SGP/ACSM/C1 - missing data | sgpaosacsmC1.b1 | Missing |
D150410.2 | SGP/ACSM/C1 - missing data | sgpaosacsmC1.b1 | Missing |
D150410.3 | SGP/ACSM/C1 - missing data | sgpaosacsmC1.b1 | Missing |
D150410.4 | SGP/ACSM/C1 - missing data | sgpaosacsmC1.b1 | Missing |
D150429.4 | SGP/ACSM/C1 - Vacuum pump failure | sgpaosacsmC1.b1 | Incorrect |
D150513.1 | SGP/ACSM/C1 - Sampling pump failure | sgpaosacsmC1.b1 | Incorrect |
D150702.7 | SGP/ACSM/C1 - Missing data | sgpaosacsmC1.b1 | Missing |
D150708.1 | SGP/ACSM/C1 - Power outage and calibration | sgpaosacsmC1.b1 | Missing |
D150819.1 | SGP/ACSM/C1 - Instrument inlet pressure low | sgpaosacsmC1.b1 | Incorrect |
D150925.1 | SGP/ACSM/C1 - Instrument averaging time changed | sgpaosacsmC1.b1 | Does not affect quality |
D151120.1 | SGP/ACSM/C1 - Filter switching valve failure | sgpaosacsmC1.b1 | Incorrect |
D160104.7 | SGP Site Power Outage Due to Winter Storm | sgpaosacsmC1.b1 | Does not affect quality |
D160105.1 | SGP/ACSM/C1 - Power outage | sgpaosacsmC1.b1 | Missing |
D160119.9 | SGP/ACSM/C1 - instrument calibration | sgpaosacsmC1.b1 | Missing |
D160406.2 | SGP/AOS/C1 - Wildfire smoke plumes advecting over facility | sgpaosacsmC1.b1 | Does not affect quality |
D160715.4 | SGP/ACSM/NEPHELOMETER/TDMA/C1 - Data Quality Evaluation Ongoing | sgpaosacsmC1.b1 | Does not affect quality |
D160909.12 | SGP/ACSM/C1 - Filter switching valve not operating | sgpaosacsmC1.b1 | Incorrect |
D161005.1 | SGP/ACSM/C1 - Low inlet pressure in sample position | sgpaosacsmC1.b1 | Suspect |
D161005.2 | SGP/ACSM/C1 - Instrument leak and updated for new system | sgpaosacsmC1.b1 | Missing |
D161027.2 | SGP/AOS/C1 - Environmental Smoke | sgpaosacsmC1.b1 | Does not affect quality |
D161216.1 | SGP/ACSM/C1 - Critical orifice clog | sgpaosacsmC1.b1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/14/2012 | 1500 | 07/31/2012 | 2359 | Incorrect |
06/03/2012 | 1800 | 06/12/2012 | 0700 | Incorrect |
Subject: | SGP/ACSM/C1 - Water in inlet sample line |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data became progressively more noisy causing large spikes in data. These spikes started occurring for longer periods of time during the day before the instrument went down completely on 7/12. The afternoon data (usually 15:00-24:00 UTC) was incorrect because of water condensation in the inlet line and aerosol collection efficiency change. The organic mass loading is around 0 when the condensed water was absorbing all the particles. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/13/2012 | 2200 | 12/05/2012 | 2210 | Incorrect |
Subject: | SGP/ACSM/C1 - Pump failure |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The carbon vane vacuum pump that pulls air through the instrument failed 11/13 following a power failure. The pump was replaced and returned to service at 2210 GMT on 12/05/2012. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/18/2012 | 1500 | 12/21/2012 | 2130 | Incorrect |
Subject: | SGP/ACSM/C1 - Pump failure |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The carbon vane vacuum pump that pulls air through the instrument failed again and a section of tubing between the ACSM and pump was partially collapsed. The pump and tubing were replaced and returned to service at 2223 GMT on 12/18/2012. The ACSM was then tuned before resuming data collection at 2130 GMT on 12/21/2012. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/05/2012 | 2211 | 12/18/2012 | 1459 | Suspect |
Subject: | SGP/ACSM/C1 - Sample flow restricted |
DataStreams: | sgpaosacsmC1.b1 |
Description: | This period is between two failures of the carbon vane pump that pulls air through the instrument. During this time the tubing between the pump and ACSM collapsed and may have affected the sample flow rate. The collapsed tubing was replaced with a different type, one that is much stiffer and less prone to collapsing under vacuum. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/11/2012 | 0900 | 11/13/2012 | 2200 | Missing |
Subject: | SGP/ACSM/C1 - Missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/01/2013 | 0000 | 07/03/2013 | 2359 | Missing |
06/01/2013 | 0000 | 06/13/2013 | 2359 | Missing |
02/27/2013 | 0000 | 02/27/2013 | 2359 | Missing |
11/22/2012 | 0000 | 12/05/2012 | 2359 | Missing |
07/01/2012 | 0000 | 07/30/2012 | 2117 | Missing |
10/26/2013 | 0000 | 10/31/2013 | 1418 | Missing |
05/01/2013 | 0000 | 05/24/2013 | 2359 | Missing |
Subject: | SGP/ACSM/C1 - Missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/08/2014 | 0200 | 02/19/2014 | 1000 | Missing |
Subject: | SGP/ACSM/C1 - Data not available and unrecoverable |
DataStreams: | sgpaosacsmC1.b1 |
Description: | There are no data for this time period. The ionization filament burned out. The mentor switched to a backup filament on 2/17 and the data stream was good after it was adjusted and monitored on 2/19. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/25/2014 | 1100 | 06/24/2014 | 0000 | Missing |
Subject: | SGP/ACSM/C1 - Windows 7 upgrade and turbo molecular pump failure. |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Instrument was not operational during this period because of turbo molecular pump failure. the pump was replaced and the instrument restarted. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/01/2012 | 0000 | 10/09/2014 | 0000 | Does not affect quality |
Subject: | SGP/ACSM/C1 - Reprocessing Notice (RID 1237) |
DataStreams: | sgpaosacsmC1.b1 |
Description: | See below for details of reprocessing that was completed on 20150610. Original DQR category: 4 (Red, Incorrect). ----------------------------------------------------------- The data will be reprocessed with the manufacturer's software to correct errors in the calibration values, collection efficiencies, and instrument response. The data in the current form is incorrect and should not be used. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/14/2014 | 1300 | 10/27/2014 | 1530 | Missing |
Subject: | SGP/ACSM/C1 - computer software problem |
DataStreams: | sgpaosacsmC1.b1 |
Description: | There are no data for the time range specified |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/25/2014 | 1100 | 06/23/2014 | 1900 | Missing |
Subject: | SGP/ACSM/C1 - Pump Failure |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Turbo pump failure. No data are available for the specified times. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/03/2014 | 1625 | 12/04/2014 | 2030 | Incorrect |
Subject: | SGP/ACSM/C1 - Unexpected instrument shutdown |
DataStreams: | sgpaosacsmC1.b1 |
Description: | There was an unexpected shutdown of the ionizer, vaporizer, SEM, and inlet valve, possibly caused by a power outage. This resulted in incorrect data values. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/09/2015 | 1930 | 02/10/2015 | 1740 | Incorrect |
Subject: | SGP/ACSM/C1 - Clogged critical orifice |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Instrument was shut down overnight to clean critical orifice. No useful data were taken during this time. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/09/2015 | 1930 | 02/10/2015 | 1740 | Missing |
Subject: | SGP/ACSM/C1 - Critical orifice clogged |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Inlet pressure was low because the critical orifice was clogged. Orifice was cleaned and the instrument restored to operation. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/29/2011 | 1307 | 08/02/2011 | 1416 | Missing |
08/08/2011 | 1012 | 08/09/2011 | 2154 | Missing |
12/31/2011 | 2216 | 12/31/2011 | 2320 | Missing |
11/23/2011 | 1839 | 11/25/2011 | 1529 | Missing |
10/18/2011 | 0906 | 10/18/2011 | 1042 | Missing |
12/30/2011 | 2023 | 12/30/2011 | 2126 | Missing |
07/14/2011 | 2250 | 07/18/2011 | 1945 | Missing |
12/31/2011 | 1938 | 12/31/2011 | 2042 | Missing |
12/27/2011 | 1755 | 12/27/2011 | 1859 | Missing |
03/01/2011 | 0000 | 03/01/2011 | 2359 | Missing |
10/09/2011 | 1256 | 10/09/2011 | 1947 | Missing |
Subject: | SGP/ACSM/C1 - missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data for the specified periods are missing because if instrument problems. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/11/2012 | 2201 | 11/15/2012 | 0051 | Missing |
10/03/2012 | 0140 | 10/03/2012 | 1832 | Missing |
03/21/2012 | 1058 | 03/21/2012 | 1233 | Missing |
04/01/2012 | 0102 | 04/03/2012 | 0913 | Missing |
05/23/2012 | 1806 | 05/23/2012 | 1909 | Missing |
07/02/2012 | 1623 | 07/03/2012 | 2105 | Missing |
07/12/2012 | 1038 | 07/30/2012 | 2221 | Missing |
06/01/2012 | 0000 | 07/01/2012 | 0000 | Missing |
11/20/2012 | 1358 | 11/21/2012 | 1633 | Missing |
Subject: | SGP/ACSM/C1 - missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data for the periods specified is missing because of instrument problems. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/27/2013 | 1619 | 05/27/2013 | 1723 | Missing |
07/01/2013 | 2347 | 07/04/2013 | 0223 | Missing |
05/10/2013 | 0232 | 05/24/2013 | 1733 | Missing |
12/09/2013 | 1441 | 12/10/2013 | 1618 | Missing |
06/09/2013 | 0250 | 06/13/2013 | 2223 | Missing |
10/25/2013 | 1903 | 10/31/2013 | 1418 | Missing |
09/21/2013 | 2157 | 09/21/2013 | 2300 | Missing |
02/26/2013 | 0341 | 02/28/2013 | 0512 | Missing |
04/01/2013 | 0102 | 04/03/2013 | 0913 | Missing |
Subject: | SGP/ACSM/C1 - missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data from the specified time periods are missing because of instrument problems. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/05/2014 | 2054 | 10/06/2014 | 1916 | Missing |
05/25/2014 | 0000 | 10/01/2014 | 0000 | Missing |
10/28/2014 | 1324 | 10/28/2014 | 1422 | Missing |
02/08/2014 | 0100 | 02/23/2014 | 0314 | Missing |
05/05/2014 | 1248 | 05/05/2014 | 1352 | Missing |
10/09/2014 | 1300 | 10/09/2014 | 1937 | Missing |
12/03/2014 | 1432 | 12/05/2014 | 0214 | Missing |
10/15/2014 | 1656 | 10/27/2014 | 1617 | Missing |
01/30/2014 | 1537 | 01/30/2014 | 1744 | Missing |
03/05/2014 | 1440 | 03/05/2014 | 1544 | Missing |
Subject: | SGP/ACSM/C1 - missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data are missing for the specified periods because of instrument problems. Data for the period from the end of May through end of September are missing because of the failure of two turbo pumps and the return of the instrument to the manufacturer for service. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/08/2015 | 0530 | 04/27/2015 | 1440 | Incorrect |
Subject: | SGP/ACSM/C1 - Vacuum pump failure |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The vacuum pump used for pulling samples past the inlet of the ACSM failed. The instrument was operational, but not seeing any aerosol. Replacement of the pump solved the problem. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/08/2015 | 0337 | 04/24/2015 | 2150 | Incorrect |
Subject: | SGP/ACSM/C1 - Sampling pump failure |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The filter switching valve and sampling pump failed. Data were collected, but no external air was being drawn into the system so there is no aerosol signal. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/24/2015 | 1804 | 06/26/2015 | 1506 | Missing |
Subject: | SGP/ACSM/C1 - Missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The data acquisition program locked up during the period specified. Program restarted and the instrument began collecting data. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/06/2015 | 0236 | 07/07/2015 | 2243 | Missing |
Subject: | SGP/ACSM/C1 - Power outage and calibration |
DataStreams: | sgpaosacsmC1.b1 |
Description: | There was a power outage on 2015-07-06 that shut down the instrument. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/11/2015 | 1739 | 10/06/2015 | 1927 | Incorrect |
Subject: | SGP/ACSM/C1 - Instrument inlet pressure low |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The pressure at the inlet to the instrument was at least 100 torr below ambient after work was performed on the vacuum line serving the ACSM. The inlet pressure also dropped from 1.3 torr to approximately 1 torr. Examination of the data during reprocessing shows that aerosol loadings during the period specified are unreasonably low and suggest that the instrument was not sampling ambient air. The data for the specified period are unreliable and should not be used. I visited SPG to perform ACSM tuning and calibration on 2015-10-06. While there I measured the flow in the line bringing sample to the ACSM. The flow was > 45 lpm. This high rate was pulling a vacuum on the ACSM sampling line. The vacuum was high when the cyclone was in line upstream of the ACSM and was the cause of the low pressure seen after the changes to the system were made in August. The flow past the ACSM inlet should be 3 lpm. The inlet pressure was ~700 torr when it should be ambient. I disconnected the sampling line from the Gast vacuum pump and reconnected and connected it to the Aerodyne pump which is part of the dryer system. The flow rate is now 3 lpm, the ambient pressure 740 torr, and the inlet pressure 1.3 torr. The instrument was calibrated and is now performing properly. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/26/2015 | 1900 | 09/22/2015 | 1319 | Does not affect quality |
Subject: | SGP/ACSM/C1 - Instrument averaging time changed |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The number of sets collected in each data write was changed when the instrument was restarted on 2015-08-26. This resulted in a change in the averaging time from 30 minutes to 10 minutes. Users should be aware that the data during this period are 10 minute averages and not the standard 30 minute average. The original settings were restored on 2015-09-22 at 13:19. Other than the change in averaging time the data are unaffected. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/03/2015 | 1803 | 11/18/2015 | 1836 | Incorrect |
Subject: | SGP/ACSM/C1 - Filter switching valve failure |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data are incorrect or not available for the period the switching valve was not working. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/26/2015 | 0000 | 01/01/2016 | 2359 | Does not affect quality |
Subject: | SGP Site Power Outage Due to Winter Storm |
DataStreams: | sgpaosacsmC1.b1 |
Description: | On Saturday and Sunday, December 26 and 27th a powerful winter storm (Goliath) brought sleet and freezing rain to much of Oklahoma with wind gusts of 45 MPH at times. The result was very very slick roads and mass power outages. The SGP central facility lost power at 8:55 PM CDT on Sunday, December 27th. The power was restored on 7:56 PM CDT on Wednesday, December 30th Site Emergency procedures were activated and Technicians drained water out of instrument reservoirs and other measures as needed, early Monday morning. On Wednesdays as roads and power started was restored to much of the areas east of I-35 , Site Technicians performed maintenance on several Extended Facilities. During this period, data are potentially missing, incorrect, or suspect. Carefully inspect the data before use. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/28/2015 | 0129 | 01/04/2016 | 1558 | Missing |
Subject: | SGP/ACSM/C1 - Power outage |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Power was out at the SGP central facility from 2015-12-28 to 2015-12-30. The instrument was restarted on 2016-01-04. There are no data for the time period specified. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/14/2016 | 1348 | 01/14/2016 | 1838 | Missing |
Subject: | SGP/ACSM/C1 - instrument calibration |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The ACSM was offline for tuning and calibration during the period specified. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/18/2016 | 0400 | 03/18/2016 | 1100 | Does not affect quality |
03/25/2016 | 0100 | 03/25/2016 | 0700 | Does not affect quality |
Subject: | SGP/AOS/C1 - Wildfire smoke plumes advecting over facility |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Smoke from nearby wildfires advected through SGP C1 during the listed affected time ranges, resulting in high signals observable in many of the listed AOS datastreams. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/01/2011 | 0012 | 09/30/2014 | 2359 | Does not affect quality |
Subject: | SGP/ACSM/NEPHELOMETER/TDMA/C1 - Data Quality Evaluation Ongoing |
DataStreams: | sgpaosacsmC1.b1 |
Description: | A closure analysis is currently being perform on the ACSM, TDMA, and wet and dry NOAA NEPHELOMETERS at SGP C1. If problems are found during this analysis, they will be promptly documented through the DQR system. The results of this analysis will be made available once completed. Please contact the respective instrument mentors with any questions regarding these data over the specified period. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/12/2016 | 2055 | 08/31/2016 | 1249 | Incorrect |
Subject: | SGP/ACSM/C1 - Filter switching valve not operating |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Filter switching valve was not operating because of a faulty circuit board. Problem was fixed and data have been collected from the end of the specified time period. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/02/2016 | 1430 | 09/14/2016 | 1811 | Suspect |
Subject: | SGP/ACSM/C1 - Low inlet pressure in sample position |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The data are suspect because the new filter switching valve was out of adjustment causing low inlet pressure in the sample position. The problem was corrected by the onsite operator. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/03/2016 | 1700 | 11/17/2016 | 2359 | Missing |
Subject: | SGP/ACSM/C1 - Instrument leak and updated for new system |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The instrument developed a leak and the instrument mentor requested the instrument be removed and sent back for repairs and upgrades for the new AOS. The new configuration of the instrument would not fit into the AOS racks. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/17/2016 | 1200 | 10/18/2016 | 0200 | Does not affect quality |
Subject: | SGP/AOS/C1 - Environmental Smoke |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Large aerosol signal associated with local agricultural or grass fires |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/14/2016 | 2058 | 12/15/2016 | 2155 | Suspect |
Subject: | SGP/ACSM/C1 - Critical orifice clog |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Critical orifice was clogged during the period specified. Data are corrected for reduced sample flow, but may be noisy because of reduced signal. |
Measurements: | sgpaosacsmC1.b1:
|