DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140618.2 | ENA/CPC/NEPH/PSAP/C1 - Impactor not installed | enaaosuhsasC1.a1 | Does not affect quality |
D140731.2 | ENA/UHSAS/C1 - Reprocess: Upper and lower size limit incorrect | enaaosuhsasC1.a1 | Incorrect |
D140926.3 | ENA/UHSAS/C1 - Computer Failure | enaaosuhsasC1.a1 | Missing |
D141216.3 | ENA/UHSAS/C1 - Problem with time fields | enaaosuhsasC1.a1 | Suspect |
D150506.2 | ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit | enaaosuhsasC1.a1 | Incorrect |
D150909.1 | ENA/UHSAS/C1 - Reprocess: Non-optimal bin sizes | enaaosuhsasC1.a1 | Does not affect quality |
D151214.1 | ENA/UHSAS/C1 - High counts on three lowest channels | enaaosuhsasC1.a1 | Suspect |
D151228.1 | ENA/UHSAS/C1 - spikes in lower size bins | enaaosuhsasC1.a1 | Suspect |
D160107.1 | ENA/UHSAS/C1 - HEPA filter attached to inlet | enaaosuhsasC1.a1 | Incorrect |
D160126.5 | ENA/UHSAS/C1 - Noisy counts / no counts | enaaosuhsasC1.a1 | Incorrect |
D160209.1 | ENA/UHSAS/C1 - High counts in lower bins | enaaosuhsasC1.a1 | Suspect |
D160211.2 | ENA/UHSAS/C1 - High counts in lower bins | enaaosuhsasC1.a1 | Incorrect |
D160223.1 | ENA/UHSAS/C1 - Intermittent spikes in the lowest channels | enaaosuhsasC1.a1 | Suspect |
D160314.1 | ENA/UHSAS/C1 - High counts in the 3 lowest size bins | enaaosuhsasC1.a1 | Suspect |
D160319.1 | ENA/UHSAS/C1 - High counts in the 3 lowest size bins | enaaosuhsasC1.a1 | Suspect |
D160321.1 | ENA/UHSAS/C1 - High counts in 3 lowest size bins | enaaosuhsasC1.a1 | Suspect |
D160411.1 | ENA/UHSAS/C1 - Intermittent spikes in the lower size bins | enaaosuhsasC1.a1 | Suspect |
D160411.2 | ENA/UHSAS/C1 - Instrument not sampling ambient air | enaaosuhsasC1.a1 | Incorrect |
D160610.10 | ENA/UHSAS/C1 - High counts in lower size bins | enaaosuhsasC1.a1 | Suspect |
D160610.9 | ENA/UHSAS/C1 - missing data | enaaosuhsasC1.a1 | Missing |
D161025.6 | ENA/AOS/C1 - Instrument mentors working at the site | enaaosuhsasC1.a1 | Suspect |
D170303.12 | ENA/AOS/C1 - Instrument | enaaosuhsasC1.a1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/01/2013 | 0000 | 02/19/2014 | 1351 | Does not affect quality |
Subject: | ENA/CPC/NEPH/PSAP/C1 - Impactor not installed |
DataStreams: | enaaosuhsasC1.a1 |
Description: | No impactor installed. All data is the same as when impactor at 10 um setting. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/17/2014 | 1455 | 07/16/2014 | 1459 | Incorrect |
Subject: | ENA/UHSAS/C1 - Reprocess: Upper and lower size limit incorrect |
DataStreams: | enaaosuhsasC1.a1 |
Description: | The lower size limit and upper size limit (99 bins) is incorrect because the values are not read from the header records. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/17/2014 | 1500 | 09/23/2014 | 2007 | Missing |
Subject: | ENA/UHSAS/C1 - Computer Failure |
DataStreams: | enaaosuhsasC1.a1 |
Description: | The CPC unit for the UHSAS (and other systems) failed. The UHSAS was down at this time and came back up after the CPC unit was fixed. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/17/2014 | 1455 | 07/25/2016 | 1805 | Suspect |
Subject: | ENA/UHSAS/C1 - Problem with time fields |
DataStreams: | enaaosuhsasC1.a1 |
Description: | During the ENA deployment, the UHSAS instrument clock was not synced with a time server. I used a comparison of the peaks in the UHSAS data and the CPC or the PSAP data to determine how far off the clock was. The UHSAS was always within ~2 minutes of the other instruments and the CPU it was attached to. Starting from 2/18/2014 the UHSAS was ~ 45 seconds. slow. It gradually got faster and around 6/2014 it was within seconds of the CPC. It continued in that direction and gradually became slower than the CPC or PSAP. On 1/22/2015 the UHSAS was ~120 seconds slower than the PSAP. At 16:05:00 1/22/2015 the UHSAS time was manually set to the Unit 2 CPU time. This was done remotely and apparantly only to within about 20 seconds. A Document shows the times of the peaks used to determine the time difference. During 2015 until the Domain time II was installed on 07/25/2016 the UHSAS was synced manually about once a month. Between syncs it generally drifted slow. During the period until 06/24/2015 the monthly correction was less than 20 seconds. There was a gap in the corrections and on 10/21/2015 it was corrected from 1 minute 22 seconds slow. The corrections after that were from about 15 to 35 seconds slow. The first time sync after Domain time II was installed corrected the time from 1 minute 12 seconds slow. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/22/2015 | 1700 | 04/27/2015 | 1700 | Incorrect |
Subject: | ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit |
DataStreams: | enaaosuhsasC1.a1 |
Description: | There was visit of 7 AOS instrument mentors to the ENA AOS during this period of instrument repair, calibration and modification. Instruments were occasionally off, or with no sample flow, or receiving calibration samples, etc. There was no productive sampling of ambient air during this period other than for testing. No data should be used for this period as ambient samples. Note: Two datastream are missing. The CAPS Extinction Monitor and the Greenhouse gas instrument. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/20/2014 | 1600 | 09/30/2015 | 2359 | Does not affect quality |
Subject: | ENA/UHSAS/C1 - Reprocess: Non-optimal bin sizes |
DataStreams: | enaaosuhsasC1.a1 |
Description: | The UHSAS was programmed with a bin sizes appropriate to the aerial UHSAS. This is not really a problem. The bin sizes are not consistent with the other console UHSAS. The problem is in blindly comparing the data between the various UHSAS. This is not a problem of mislabeling or incorrect size bins. The data are correct for the stated size bins. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/11/2015 | 1001 | 12/12/2015 | 0058 | Suspect |
12/10/2015 | 1117 | 12/11/2015 | 0636 | Suspect |
12/12/2015 | 2017 | 12/13/2015 | 2100 | Suspect |
12/08/2015 | 1739 | 12/08/2015 | 2346 | Suspect |
12/13/2015 | 2041 | 12/14/2015 | 0903 | Suspect |
Subject: | ENA/UHSAS/C1 - High counts on three lowest channels |
DataStreams: | enaaosuhsasC1.a1 |
Description: | High counts in the UHSAS three lowest channels. Probably a contamination (a fiber on the jet nozzle) inside the optical chamber. Rest of the channels should not be affected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/25/2015 | 1200 | 12/28/2015 | 1540 | Suspect |
Subject: | ENA/UHSAS/C1 - spikes in lower size bins |
DataStreams: | enaaosuhsasC1.a1 |
Description: | Intermittent spikes in the lowest 3 size bins. Probably a contamination that was cleared when a HEPA filter was installed for zero testing. Other bins should not be affected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/30/2015 | 1900 | 01/04/2016 | 1600 | Incorrect |
Subject: | ENA/UHSAS/C1 - HEPA filter attached to inlet |
DataStreams: | enaaosuhsasC1.a1 |
Description: | HEPA filter was attached to the UHSAS inlet for troubleshooting the issues related to DQR D151228.1. No ambient measurements were made during this period. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/25/2016 | 1300 | 01/26/2016 | 1745 | Incorrect |
Subject: | ENA/UHSAS/C1 - Noisy counts / no counts |
DataStreams: | enaaosuhsasC1.a1 |
Description: | UHSAS was showing spikes in the lowest channels. An HEPA filter was attached to the inlet for troubleshooting. No valid data during the period with spikes and when the filter was attached. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/05/2016 | 0453 | 02/05/2016 | 1445 | Suspect |
Subject: | ENA/UHSAS/C1 - High counts in lower bins |
DataStreams: | enaaosuhsasC1.a1 |
Description: | High counts in the 3 lowest bins/channels. Possible contamination. Other bins should not be affected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/10/2016 | 1345 | 02/11/2016 | 1245 | Incorrect |
Subject: | ENA/UHSAS/C1 - High counts in lower bins |
DataStreams: | enaaosuhsasC1.a1 |
Description: | High counts in the 4 lowest bins/channels. Possible contamination. Instrument set up to sample through a filter for troubleshooting. No valid data from the period. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/21/2016 | 2134 | 02/21/2016 | 2238 | Suspect |
02/22/2016 | 0122 | 02/22/2016 | 0300 | Suspect |
02/19/2016 | 2327 | 02/20/2016 | 0249 | Suspect |
02/21/2016 | 0555 | 02/21/2016 | 0636 | Suspect |
Subject: | ENA/UHSAS/C1 - Intermittent spikes in the lowest channels |
DataStreams: | enaaosuhsasC1.a1 |
Description: | Intermittent spikes in the 3 lowest size bins. Rest of the bins should be unaffected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/12/2016 | 0310 | 03/12/2016 | 0737 | Suspect |
03/11/2016 | 1246 | 03/11/2016 | 1542 | Suspect |
03/13/2016 | 1425 | 03/13/2016 | 1708 | Suspect |
03/13/2016 | 1206 | 03/13/2016 | 1240 | Suspect |
Subject: | ENA/UHSAS/C1 - High counts in the 3 lowest size bins |
DataStreams: | enaaosuhsasC1.a1 |
Description: | High counts in the lowest 3 size bins. The rest of the bins should be unaffected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/18/2016 | 0525 | 03/18/2016 | 0735 | Suspect |
Subject: | ENA/UHSAS/C1 - High counts in the 3 lowest size bins |
DataStreams: | enaaosuhsasC1.a1 |
Description: | High counts in the 3 lowest size bins. Other size bins should not be affected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/20/2016 | 0450 | 03/20/2016 | 0822 | Suspect |
Subject: | ENA/UHSAS/C1 - High counts in 3 lowest size bins |
DataStreams: | enaaosuhsasC1.a1 |
Description: | High counts in the 3 lowest size bins. Other size bins should not be affected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/28/2016 | 2200 | 04/01/2016 | 2330 | Suspect |
Subject: | ENA/UHSAS/C1 - Intermittent spikes in the lower size bins |
DataStreams: | enaaosuhsasC1.a1 |
Description: | Intermittent spikes in the 3 lowest size bins. Other size bins should not be affected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/31/2016 | 1600 | 04/01/2016 | 1025 | Incorrect |
03/30/2016 | 1530 | 03/30/2016 | 1900 | Incorrect |
03/30/2016 | 0900 | 03/30/2016 | 1100 | Incorrect |
Subject: | ENA/UHSAS/C1 - Instrument not sampling ambient air |
DataStreams: | enaaosuhsasC1.a1 |
Description: | Instrument was periodically offline or not sampling ambient air for troubleshooting purposes. Data missing or no valid data. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/07/2016 | 1340 | 06/08/2016 | 0200 | Suspect |
Subject: | ENA/UHSAS/C1 - High counts in lower size bins |
DataStreams: | enaaosuhsasC1.a1 |
Description: | High counts in the 3 lowest size bins. Other size bins should not be affected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/06/2016 | 2210 | 06/07/2016 | 1340 | Missing |
Subject: | ENA/UHSAS/C1 - missing data |
DataStreams: | enaaosuhsasC1.a1 |
Description: | Instrument's internal hard drive was full and no new data was being saved except a few samples occasionally. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/26/2016 | 0830 | 10/01/2016 | 0100 | Suspect |
Subject: | ENA/AOS/C1 - Instrument mentors working at the site |
DataStreams: | enaaosuhsasC1.a1 |
Description: | Instrument mentors working at the site. Sample lines and electrical connections being reconfigured. Instruments that had no work done on them may still be affected. |
Measurements: | enaaosuhsasC1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/20/2016 | 0000 | 03/03/2017 | 0000 | Does not affect quality |
Subject: | ENA/AOS/C1 - Instrument |
DataStreams: | enaaosuhsasC1.a1 |
Description: | Pentras dry air supply intermittently not working. Dry air missing for the humidigraph and the HT-DMA. Dry air supply for those instruments partially replaced by regular undried compressed air, lowering the delivery pressure. No direct effect on other instruments seen. |
Measurements: | enaaosuhsasC1.a1:
|