DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160617.14 | ASI/AOS/CCN/HTDMA/NEPHELOMETER/UHSAS/M1 - Site installation | asiaosuhsasM1.a1 | Incorrect |
D160621.4 | ASI/AOS/M1 - Manifold blower not working | asiaosuhsasM1.a1 | Incorrect |
D160708.10 | ASI/UHSAS/M1 - Instrument offline | asiaosuhsasM1.a1 | Missing |
D160824.10 | ASI/UHSAS/M1 - Unstable flows | asiaosuhsasM1.a1 | Incorrect |
D170202.7 | ASI/CPC/UHSAS/M1 - HEPA filter testing | asiaosuhsasM1.a1 | Incorrect |
D170203.7 | ASI/AOS/M1 - Mentor visit | asiaosuhsasM1.a1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/22/2016 | 0000 | 05/31/2016 | 2359 | Incorrect |
Subject: | ASI/AOS/CCN/HTDMA/NEPHELOMETER/UHSAS/M1 - Site installation |
DataStreams: | asiaosuhsasM1.a1 |
Description: | Site installation, mentors on site. No valid data. |
Measurements: | asiaosuhsasM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2016 | 0042 | 06/16/2016 | 1345 | Incorrect |
Subject: | ASI/AOS/M1 - Manifold blower not working |
DataStreams: | asiaosuhsasM1.a1 |
Description: | Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic. |
Measurements: | asiaosuhsasM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/08/2016 | 1435 | 07/08/2016 | 1615 | Missing |
Subject: | ASI/UHSAS/M1 - Instrument offline |
DataStreams: | asiaosuhsasM1.a1 |
Description: | Instrument offline for flow adjustment. |
Measurements: | asiaosuhsasM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/22/2016 | 1600 | 08/22/2016 | 1900 | Incorrect |
Subject: | ASI/UHSAS/M1 - Unstable flows |
DataStreams: | asiaosuhsasM1.a1 |
Description: | Cleaning the internal flow lines and testing proper operation after a possible obstruction in the flow lines. No valid data from this period. |
Measurements: | asiaosuhsasM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/21/2017 | 1029 | 01/21/2017 | 1030 | Incorrect |
01/25/2017 | 1015 | 01/25/2017 | 1018 | Incorrect |
Subject: | ASI/CPC/UHSAS/M1 - HEPA filter testing |
DataStreams: | asiaosuhsasM1.a1 |
Description: | HEPA filter on the inlet to all three instruments during this time for zero checks. All instruments went to zero counts |
Measurements: | asiaosuhsasM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/19/2017 | 1000 | 01/27/2017 | 1900 | Suspect |
Subject: | ASI/AOS/M1 - Mentor visit |
DataStreams: | asiaosuhsasM1.a1 |
Description: | Mentors at the site working on different instruments. Data quality occasionally affected. |
Measurements: | asiaosuhsasM1.a1:
|