DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D030311.1 | TWP/SKYRAD/C2 - Reprocess: Incorrect UVB Calibration Coefficient | twpskyrad60sC2.b1 | Suspect |
D030417.1 | TWP/SKYRAD/C2 - UVB Data questionable/incorrect due to uncontrolled sensor temperature | twpskyrad60sC2.b1 | Suspect |
D030821.5 | TWP/SKYRAD/C2 - Reprocessed and Corrected: IRT incorrect min/delta values | twpskyrad60sC2.b1 | Does not affect quality |
D031007.3 | TWP/SKYRAD/C2 - Incorrect Calibration Coefficients Applied | twpskyrad60sC2.b1 | Suspect |
D040812.3 | TWP/SKYRAD/C2 - UVB Data questionable/incorrect due to uncontrolled sensor temperature | twpskyrad60sC2.b1 | Suspect |
D131220.2 | TWP/SKYRAD/C2 - IRT bias | twpskyrad60sC2.b1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/03/2003 | 0500 | 03/11/2003 | 0700 | Suspect |
Subject: | TWP/SKYRAD/C2 - Reprocess: Incorrect UVB Calibration Coefficient |
DataStreams: | twpskyrad60sC2.b1 |
Description: | The uvb calibration coefficient was entered incorrectly as 0.2151 instead of 0.2332. |
Measurements: | twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/20/1998 | 0000 | 03/03/2003 | 2359 | Suspect |
Subject: | TWP/SKYRAD/C2 - UVB Data questionable/incorrect due to uncontrolled sensor temperature |
DataStreams: | twpskyrad60sC2.b1 |
Description: | The UVB was not temperature controlled to 25C as required to ensure accurate data. UVB data should ONLY be used after checking both the bit_data and twpskyrad20sC1.a1 uvb_temp or inst_down_uvb_temp. UVB data where the UVB temperature was outside the range of 24-26C should not be used. |
Measurements: | twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/01/2001 | 0005 | 03/03/2003 | 2102 | Does not affect quality |
Subject: | TWP/SKYRAD/C2 - Reprocessed and Corrected: IRT incorrect min/delta values |
DataStreams: | twpskyrad60sC2.b1 |
Description: | DATA REPROCESSED and CORRECTED: The min/max/delta values for sky_ir_temp were originally incorrect in the 60s files (xxxskyrad60sFn.b1). TWP was -49.7/40/-9999 and NSA was -49/50/50. The min/max/delta for the TWP and NSA data has now been corrected to -60/40/50. The data were archived in November 2016. Please retrieve the reprocessed data for use in your analysis. You may find these data using the ARM Discovery tool at the following link: http://www.archive.arm.gov/discovery/#v/results/s/finst::skyrad/fsite::nsa/ffac::nsa.C1 http://www.archive.arm.gov/discovery/#v/results/s/finst::skyrad/fsite::nsa/ffac::nsa.C2 http://www.archive.arm.gov/discovery/#v/results/s/finst::skyrad/fsite::twp/ffac::twp.C1 http://www.archive.arm.gov/discovery/#v/results/s/finst::skyrad/fsite::twp/ffac::twp.C2 http://www.archive.arm.gov/discovery/#v/results/s/finst::skyrad/fsite::twp/ffac::twp.C3 ORIGINAL DQR CATEGORY: Transparent (Does not affect quality ) REPROCESSED and CORRECTED DATA ARCHIVED:2016-11 REPROCESS IDENTIFICATION NUMBER: RID 119 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- For field sky_ir_temp, the value of valid_min was incorrectly assigned to -49.7 C and no value of valid_delta was assigned. The actual values should be -60 and 50 C, respectively. |
Measurements: | twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/20/2002 | 2347 | 09/20/2003 | 2333 | Suspect |
Subject: | TWP/SKYRAD/C2 - Incorrect Calibration Coefficients Applied |
DataStreams: | twpskyrad60sC2.b1 |
Description: | Incorrect calibration coefficients were entered for three instruments (PSPG, PIRD and PIRG) at the RESET period. The following calculation method should be applied to correct the data, For the downwelling shortwave(PSPG) (DS) Old Calibration coefficient [W/m2\mV] =119.68 New Calibration Coefficient [W/m2\mV] =125.94 DS(new_value)[W/m2]=( DS(old_value)[W/m2]/ 119.68 [W/m2\mV] )* 125.94 [W/m2/mV] For the longwave irradiance (PIRG): If the .a0 and .b1 data set available the following procedure should be applied for the downwelling LW 1) Obtain the following variables from the 20s datastream for each one minute of the affected time range: inst_down_long_hemisp1_dome_resist (Rd) inst_down_long_ hemisp1_case_resist (Rc) inst_down_long_ hemisp1_tp (tpDIR_mV) 2) Using the 20s tp voltage in mV convert to uV. tpDIR[uV] = inst_down_long_hemisp1_tp [mV]*1000[uV/mV] 3) For each 20s data entry, convert Rd and Rc to temperature Td =1/ (a0 + (a1*LN (Rd*1000) + (a2*LN (Rd*1000)^3))) Tc = 1/ (a0 + (a1*LN (Rc*1000) + (a2*LN (Rc*1000)^3))) Where a0 = 0.0010295 a1 = 0.0002391 a2 = 0.0000001568 4) Apply for each 20s data entry: (correct k1=0.2755) WinDIR(W/m^2)= [0.2755*tpDIR] + Sigma*Tc^4 -4*Sigma*(Td^4 - Tc^4) Where, WinDIR = Corrected down_long_hemisp1 (W/m^2) Sigma = 5.6704E-8 (Stefan-Boltzman Constant) 5) Generate new 1 minute average from the three 20-second measurements For the longwave irradiance (PIRD): If the .a0 and .b1 data set available the following procedure should be applied for the downwelling LW 1) Obtain the following variables from the 20s datastream for each one minute of the affected time range: inst_down_long_hemisp2_dome_resist (Rd) inst_down_long_ hemisp2_case_resist (Rc) inst_down_long_ hemisp2_tp (tpDIR_mV) 2) Using the 20s tp voltage in mV convert to uV. tpDIR[uV] = inst_down_long_hemisp2_tp [mV]*1000[uV/mV] 3) For each 20s data entry, convert Rd and Rc to temperature Td =1/ (a0 + (a1*LN (Rd*1000) + (a2*LN (Rd*1000)^3))) Tc = 1/ (a0 + (a1*LN (Rc*1000) + (a2*LN (Rc*1000)^3))) Where a0 = 0.0010295 a1 = 0.0002391 a2 = 0.0000001568 4) Apply for each 20s data entry: (correct k1=0.2786) WinDIR(W/m^2)= [0.2786*tpDIR] + Sigma*Tc^4 -4*Sigma*(Td^4 - Tc^4) Where, WinDIR = Corrected down_long_hemisp2(W/m^2) Sigma = 5.6704E-8 (Stefan-Boltzman Constant) 5) Generate new 1 minute average from the three 20-second measurements |
Measurements: | twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/04/2003 | 0000 | 07/05/2004 | 2359 | Suspect |
Subject: | TWP/SKYRAD/C2 - UVB Data questionable/incorrect due to uncontrolled sensor temperature |
DataStreams: | twpskyrad60sC2.b1 |
Description: | The UVB was not temperature controlled to 25C as required to ensure accurate data. UVB data should ONLY be used after checking both the bit_data and twpskyrad20sC1.a1 uvb_temp or inst_down_uvb_temp. UVB data where the UVB temperature was outside the range of 24-26C should not be used. |
Measurements: | twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/29/2001 | 1200 | 08/16/2007 | 0001 | Suspect |
Subject: | TWP/SKYRAD/C2 - IRT bias |
DataStreams: | twpskyrad60sC2.b1 |
Description: | Measurements of sky temperature, especially those <250K, were positively biased compared with AERI because there was no compensation for the reflectivity of the gold mirror. The problem was corrected by changing the IRT emissivity from 1.000 to 0.987. Use sky temperature measurements less than 230K with caution. Biases as large as 30K have been observed when the actual temperature, as measured by AERI, is less than 210 K. |
Measurements: | twpskyrad60sC2.b1:
|