DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140203.1 | MAO/AERI/M1 - Reprocessing Notice (RID 1102) | maoaerich2M1.b1 | Does not affect quality |
D151203.7 | MAO/AERI/M1 - White light failure | maoaerich2M1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/10/2013 | 1343 | 01/03/2014 | 2015 | Does not affect quality |
Subject: | MAO/AERI/M1 - Reprocessing Notice (RID 1102) |
DataStreams: | maoaerich2M1.b1 |
Description: | See below for details of reprocessing that was completed on 20140708. Original DQR category: 5 (Transparent, Does not affect quality). ---------------------------------------------------------- Site techs were not instructed to update the configuration of the latitude, longitude and altitude on the AERI computer at setup. Mentor has updated configuration file and AERI raw data contains correct coordinates starting 01/03/2014 20:15 UTC. Incorrect values were 42.030445, -70.049472, 48. Correct values should be lat= -3.212972, lon= -60.59806, alt = 49.99. |
Measurements: | maoaerich2M1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/13/2015 | 0000 | 12/02/2015 | 0000 | Missing |
Subject: | MAO/AERI/M1 - White light failure |
DataStreams: | maoaerich2M1.b1 |
Description: | No data during this period. White light had failed. Instrument repairs onsite by Jonathan Gero were successfully completed and instrument was back operational on December 1. End of deployment calibration verification tests passed, instrument was packed and is ready to be installed at the next site. |