DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D131024.2 | OLI/GNDRAD/M1 - Instrument not properly grounded | oligndrad60sM1.b1 | Incorrect |
D150123.7 | OLI/GNDRAD/M1 - Environmental | oligndrad60sM1.b1 | Incorrect |
D150327.7 | OLI/GNDRAD/M1 - Environmental | oligndrad60sM1.b1 | Incorrect |
D151102.3 | OLI/GNDRAD/M1 - PIR cable connector failure | oligndrad60sM1.b1 | Incorrect |
D170110.1 | OLI/M1 - Severe Storm | oligndrad60sM1.b1 | Does not affect quality |
D170116.4 | OLI/GNDRAD/M1 - power outage | oligndrad60sM1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/19/2013 | 0000 | 10/18/2013 | 1820 | Incorrect |
Subject: | OLI/GNDRAD/M1 - Instrument not properly grounded |
DataStreams: | oligndrad60sM1.b1 |
Description: | At the start of data collection all OLI GNDRAD measurements were corrupted by a grounding problem. Additional grounding methods were put into place on 10/18/2013 eliminating the problem. All GNDRAD data are suspect during the specified time range. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/12/2014 | 2230 | 01/15/2015 | 0030 | Incorrect |
12/31/2014 | 0300 | 01/09/2015 | 0500 | Incorrect |
Subject: | OLI/GNDRAD/M1 - Environmental |
DataStreams: | oligndrad60sM1.b1 |
Description: | GNDRAD system not operating correctly during affected time periods due to power outage. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/01/2015 | 1916 | 03/08/2015 | 0043 | Incorrect |
Subject: | OLI/GNDRAD/M1 - Environmental |
DataStreams: | oligndrad60sM1.b1 |
Description: | AC power out during affected time range affecting ventilator fans and heaters for up_short_hemisp, and up_long_hemisp radiometers. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/22/2015 | 0000 | 10/25/2015 | 0321 | Incorrect |
Subject: | OLI/GNDRAD/M1 - PIR cable connector failure |
DataStreams: | oligndrad60sM1.b1 |
Description: | During the specified time period upwelling PIR cable/connector problems were being corrected. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/05/2017 | 1000 | 01/08/2017 | 1200 | Does not affect quality |
Subject: | OLI/M1 - Severe Storm |
DataStreams: | oligndrad60sM1.b1 |
Description: | The site lost power during a severe winter storm. Operators on site were not able to get out and restart the generators to bring things back up until 1/7/2017. Most of the instruments were brought up at this time, but some problems remained until 1/8/2017 |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/05/2017 | 1815 | 01/07/2017 | 1630 | Missing |
Subject: | OLI/GNDRAD/M1 - power outage |
DataStreams: | oligndrad60sM1.b1 |
Description: | Blizzard conditions resulted in a power outage. |
Measurements: | oligndrad60sM1.b1:
|