D141013.7 NSA/KAZR/C1 - Snow decreasing sensitivity nsakazrspeccmaskgecopolC1.a0 Suspect
D131120.3 NSA/KAZR/C1 - Reprocess: Incorrect calibration coefficients nsakazrspeccmaskgexpolC1.a0 Incorrect
 
DQR ID: D141013.7    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2014-09-02 21:00:00 2014-09-03 18:20:00 Suspect
Subject: NSA/KAZR/C1 - Snow decreasing sensitivity
Affected Datastreams: nsakazrspeccmaskgecopolC1.a0
Description There was a possible build up of snow on the instrument causing decreased sensitivity in the returns.
Measurements
  • signal_to_noise_ratio (Signal-to-noise ratio)
  • reflectivity (Reflectivity)
 
DQR ID: D131120.3    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2011-01-18 00:00:00 2014-02-07 23:59:00 Incorrect
Subject: NSA/KAZR/C1 - Reprocess: Incorrect calibration coefficients
Affected Datastreams: nsakazrspeccmaskgexpolC1.a0
Description The wrong calibrations were used due to a problem with the software. Mentor discovered the reflectivity between long and short pulse did not agree with each other after analyzing data. Data will be reprocessed.
Measurements