DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D080717.2 | SGP/CSPHOT/C1 - Instrument failure | sgpcsphotaotfiltC1.a1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2008-06-22 | 00:00:00 | 2008-07-24 | 23:59:00 | Suspect |
Subject: | SGP/CSPHOT/C1 - Instrument failure |
---|---|
Affected Datastreams: | sgpcsphotaotfiltC1.a1 |
Description | There was a distinct cosine curve in the AOD data, which indicated a problem with the instrument, possibly due to a bad eprom in the control box. The instrument was removed from service on 7/24/08 and replaced on 9/24/08. |